Reproducibility, interchangeability of measures, time to measure stabilization, and reference values of two tissue oximeters in healthy volunteers

Abstract. This study aimed to compare two tissue oximeters, the INVOS 5100c and the Equanox 7600, in terms of their reproducibility and the interchangeability of their measures. In a randomized order, three measurements were taken at six different sites on both sides of the body in 53 healthy volunteers. Intraclass correlation coefficients (ICC) and within-subject standard deviation (Sw) were calculated for each device. The ICCs were compared using Fisher r-to-z transformation and the Sw were compared using paired-sample t-tests. We found no difference between the reproducibility of the INVOS {ICC=0.92 [95% confidence interval (CI) 0.90 to 0.93]} and Equanox [ICC=0.90 (95% CI 0.88 to 0.93)] in terms of ICCs (p=0.06). However, the Equanox [Sw=1.96 (95% CI 1.91 to 2.02)] showed a better Sw than the INVOS [Sw=2.11 (95% CI 2.05 to 2.17)] (p=0.019). Also, when compared directly to stable condition, the readings produced by the two oximeters varied considerably [ICC 0.43 (95% CI 0.36 to 0.49)]. When taken individually, both tissue oximeters displayed good reproducibility, the Equanox being slightly better than the INVOS in terms of absolute reproducibility. However, when compared, the oximeters showed poor interdevices agreement. Reference values were also described.

[1]  R. Nelson,et al.  Cerebral near infrared spectroscopy: emitter-detector separation must be increased. , 1999, British journal of anaesthesia.

[2]  J. Moyle,et al.  Uses and abuses of pulse oximetry. , 1996, Archives of disease in childhood.

[3]  Ji-Hyun Lee,et al.  Comparison of two devices using near-infrared spectroscopy for the measurement of tissue oxygenation during a vascular occlusion test in healthy volunteers (INVOS® vs. InSpectra™) , 2015, Journal of Clinical Monitoring and Computing.

[4]  J. Hanouz,et al.  Dynamic evaluation of near-infrared peripheral oximetry in healthy volunteers: a comparison between INVOS and EQUANOX. , 2013, Journal of critical care.

[5]  M. Weiss,et al.  Reproducibility of cerebral oxygenation measurement in neonates and infants in the clinical setting using the NIRO 300 oximeter* , 2005, Pediatric critical care medicine : a journal of the Society of Critical Care Medicine and the World Federation of Pediatric Intensive and Critical Care Societies.

[6]  Menelaos Karanikolas,et al.  Baseline cerebral oximetry values in cardiac and vascular surgery patients: a prospective observational study , 2010, Journal of cardiothoracic surgery.

[7]  Gorm Greisen,et al.  Cerebral oxygenation after birth – a comparison of INVOS® and FORE‐SIGHT™ near‐infrared spectroscopy oximeters , 2014, Acta paediatrica.

[8]  N. Galdieri,et al.  Direct comparison between cerebral oximetry by INVOS(TM) and EQUANOX(TM) during cardiac surgery: a pilot study. , 2014 .

[9]  C. Sprung,et al.  Surviving Sepsis Campaign: International Guidelines for Management of Severe Sepsis and Septic Shock 2012 , 2013, Critical care medicine.

[10]  M. Rollins,et al.  Factors Affecting the Performance of 5 Cerebral Oximeters During Hypoxia in Healthy Volunteers , 2013, Anesthesia and analgesia.

[11]  Gorm Greisen,et al.  Precision of measurement of cerebral tissue oxygenation index using near-infrared spectroscopy in preterm neonates. , 2006, Journal of biomedical optics.

[12]  J. Holcomb,et al.  Continuous noninvasive tissue oximetry in the early evaluation of the combat casualty: a prospective study. , 2010, The Journal of trauma.

[13]  Simon Hyttel-Sorensen,et al.  Peripheral tissue oximetry: comparing three commercial near-infrared spectroscopy oximeters on the forearm , 2013, Journal of Clinical Monitoring and Computing.

[14]  M Smith,et al.  A comparison of cerebral oxygenation as measured by the NIRO 300 and the INVOS 5100 Near‐Infrared Spectrophotometers , 2002, Anaesthesia.

[15]  D. Altman,et al.  Measurement error. , 1996, BMJ.

[16]  Gerhard Pichler,et al.  Regional tissue oxygen saturation: comparability and reproducibility of different devices. , 2011, Journal of biomedical optics.

[17]  H. Grocott,et al.  Impact of Extracranial Contamination on Regional Cerebral Oxygen Saturation: A Comparison of Three Cerebral Oximetry Technologies , 2012, Anesthesiology.

[18]  Frank van Bel,et al.  Comparing near-infrared spectroscopy devices and their sensors for monitoring regional cerebral oxygen saturation in the neonate , 2013, Pediatric Research.

[19]  S. Cossette,et al.  Near-infrared Spectroscopy Monitoring During Cardiac Arrest: A Systematic Review and Meta-analysis. , 2016, Academic emergency medicine : official journal of the Society for Academic Emergency Medicine.

[20]  M. Schultz,et al.  Association between static and dynamic thenar near-infrared spectroscopy and mortality in patients with sepsis: A systematic review and meta-analysis , 2014, The journal of trauma and acute care surgery.

[21]  Jorge Quintero,et al.  Prehospital dynamic tissue oxygen saturation response predicts in-hospital lifesaving interventions in trauma patients , 2012, The journal of trauma and acute care surgery.

[22]  Damijan Miklavčič,et al.  Reproducibility of Parameters of Postocclusive Reactive Hyperemia Measured by Near Infrared Spectroscopy and Transcutaneous Oximetry , 2000, Annals of Biomedical Engineering.

[23]  F. Jöbsis Noninvasive, infrared monitoring of cerebral and myocardial oxygen sufficiency and circulatory parameters. , 1977, Science.

[24]  T. Fitzpatrick The validity and practicality of sun-reactive skin types I through VI. , 1988, Archives of dermatology.

[25]  Simon Hyttel-Sorensen,et al.  A comparison between two NIRS oximeters (INVOS, OxyPrem) using measurement on the arm of adults and head of infants after caesarean section. , 2014, Biomedical optics express.

[26]  L. Dibbelt,et al.  Preoperative Cerebral Oxygen Saturation and Clinical Outcomes in Cardiac Surgery , 2011, Anesthesiology.

[27]  F. Bel,et al.  Cerebral Oxygenation During the First Days of Life in Preterm and Term Neonates: Differences Between Different Brain Regions , 2011, Pediatric Research.

[28]  J. Vincent,et al.  Monitoring the microcirculation in the critically ill patient: current methods and future approaches , 2010, Intensive Care Medicine.

[29]  Simon Hyttel-Sorensen,et al.  Tissue oximetry: a comparison of mean values of regional tissue saturation, reproducibility and dynamic range of four NIRS-instruments on the human forearm , 2011, Biomedical optics express.

[30]  J. Chvojka,et al.  [International guidelines for management of severe sepsis and septic shock 2012 - comment]. , 2014, Vnitrni lekarstvi.

[31]  K. Hari Kumar,et al.  A Prospective, Observational Study of Osteoporosis in Men , 2018, Indian journal of endocrinology and metabolism.

[32]  G. Strangman,et al.  Depth Sensitivity and Source-Detector Separations for Near Infrared Spectroscopy Based on the Colin27 Brain Template , 2013, PLoS ONE.

[33]  C. Zaouter,et al.  Influence of ambient light on cerebral oximeters. , 2010, British journal of anaesthesia.

[34]  C. Cooper,et al.  Absolute quantification of deoxyhaemoglobin concentration in tissue near infrared spectroscopy. , 1994, Physics in medicine and biology.

[35]  E. Gratton,et al.  Frequency-domain multichannel optical detector for noninvasive tissue spectroscopy and oximetry , 1995 .

[36]  D. Delpy,et al.  Performance comparison of several published tissue near-infrared spectroscopy algorithms. , 1995, Analytical biochemistry.