Comparison of the Goldmann-Weekers Dark Adaptometer™ and LKC Technologies Scotopic Sensitivity Tester-1™

The standard for dark adaptation has long been the Goldmann-Weekers Dark Adaptometer™ (Haag-Streit). More recently, portable, relatively inexpensive LED-based dark adaptometers have become commercially available. These devices have potential use in areas with limited resources to screen for night-blindness, commonly caused worldwide by vitamin A deficiency. In order to determine the sensitivity to detecting changes in night vision, this study compared one such device, LKC Technologies Scotopic Sensitivity Tester-1™ (SST- 1) to the Goldmann-Weekers in patients with hereditary retinal degeneration and loss of rod function. Dark-adapted final thresholds and rod full-field ERG responses were obtained from 87 patients and 24 normal subjects. Linear regression analysis, discrepancy analysis, and receiver operator characteristic curves for both devices show that the SST-1 quantifies psychophysical rod function nearly as well as the Goldmann-Weekers, within some limitations. We conclude, therefore, that the SST-1 is a viable alternative to the Goldmann-Weekers for the screening of night-blinding retinal disorders.

[1]  W. Swanson,et al.  Extracting thresholds from noisy psychophysical data , 1992, Perception & psychophysics.

[2]  D. Hood,et al.  Rod phototransduction in retinitis pigmentosa: estimation and interpretation of parameters derived from the rod a-wave. , 1994, Investigative ophthalmology & visual science.

[3]  M. Marmor,et al.  Standard for clinical electroretinography (1999 update) , 1998, Documenta Ophthalmologica.

[4]  M. A. Bouman,et al.  Spatial and temporal summation in the human dark-adapted retina. , 1981, Journal of the Optical Society of America.

[5]  Eberhart Zrenner,et al.  Standard for clinical electroretinography (1994 update) , 2005, Documenta Ophthalmologica.

[6]  L. L. Sloan,et al.  Pigmentary degeneration of the retina: early diagnosis and natural history. , 1967, Investigative ophthalmology.

[7]  J. Hanley,et al.  The meaning and use of the area under a receiver operating characteristic (ROC) curve. , 1982, Radiology.

[8]  J M Bland,et al.  Statistical methods for assessing agreement between two methods of clinical measurement , 1986 .

[9]  James E. Birren,et al.  Age Changes in Rate and Level of Visual Dark Adaptation , 1950 .

[10]  N. Congdon,et al.  Pupillary and visual thresholds in young children as an index of population vitamin A status. , 1995, The American journal of clinical nutrition.

[11]  A. Sommer,et al.  Vitamin A Deficiency and Its Consequences: A Field Guide to Detection and Control , 1995 .

[12]  E. Birch,et al.  The relationship between rod perimetric thresholds and full-field rod ERGs in retinitis pigmentosa. , 1987, Investigative ophthalmology & visual science.