Effect of a magnetic core in the receiver coil of a biomedical inductive power system

The design of an inductive power system is investigated in order to determine the relative benefits provided by the inclusion of a magnetic core in the receiver coil. Both transmitter and receiver coils are chosen to have dimensions similar to those used in many biomedical devices. It is found that while a magnetic core provides enhanced coupling, the operating frequency is limited by losses in the magnetic material. Consequently, the performance of air-core coils is shown to match that of coils with magnetic cores when operated at a higher frequency. Work is ongoing to consider the effects of other limitations imposed by biomedical environments.

[1]  Reid R. Harrison,et al.  Designing Efficient Inductive Power Links for Implantable Devices , 2007, 2007 IEEE International Symposium on Circuits and Systems.

[2]  J.H. Schulman,et al.  Battery powered BION FES network , 2004, The 26th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[3]  C.M. Zierhofer,et al.  Geometric approach for coupling enhancement of magnetically coupled coils , 1996, IEEE Transactions on Biomedical Engineering.

[4]  M. Samet,et al.  Design of Wireless Power and Data Transmission Circuits for Implantable Biomicrosystem , 2007 .

[5]  Wentai Liu,et al.  An efficient inductive power link design for retinal prosthesis , 2004, 2004 IEEE International Symposium on Circuits and Systems (IEEE Cat. No.04CH37512).

[6]  Sérgio Francisco Pichorim,et al.  Design of coils for millimeter- and submillimeter-sized biotelemetry , 2004, IEEE Transactions on Biomedical Engineering.