Tissue response to the failed Gerard double-cup. Histologic analysis of 40 uncemented hip arthroplasties.

Femoral head remnants, joint capsules and acetabular membranes were retrieved at revision of 40 failed uncemented, Gerard double-cup hip arthroplasties. All femoral heads were covered with a fibrous membrane containing polyethylene and metal particles, macrophages, and giant cells. The same histopathologic features were found in the joint capsules and acetabular membranes. There was a relationship between the amount of polyethylene particles in the fibrous membrane of the femoral head and the presence of osteonecrosis. Resorption of necrotic bone contributed to failure of the arthroplasty, aided by granulomatous reactions to the wear particles that appeared to induce osteolysis. The overall findings indicate that the severity of the foreign body reaction in periprosthetic tissues is related to the amount of wear particles. Future designs of resurfacing hip arthroplasties should be focused on wear-resistant material combinations.

[1]  D. Howie,et al.  Peripheral wear of Wagner resurfacing hip arthroplasty acetabular components. , 1991, The Journal of arthroplasty.

[2]  D. Howie,et al.  Pseudo-abscess of the psoas bursa in failed double-cup arthroplasty of the hip. , 1991, The Journal of bone and joint surgery. British volume.

[3]  D. Murray,et al.  Macrophages stimulate bone resorption when they phagocytose particles. , 1990, The Journal of bone and joint surgery. British volume.

[4]  H. Bertram,et al.  Osteolysis in alloarthroplasty of the hip. The role of ultra-high molecular weight polyethylene wear particles. , 1990, Clinical orthopaedics and related research.

[5]  D. Howie,et al.  Resurfacing hip arthroplasty. Classification of loosening and the role of prosthesis wear particles. , 1990, Clinical orthopaedics and related research.

[6]  B. Morrey,et al.  Effect of femoral head size on wear of the polyethylene acetabular component. , 1990, The Journal of bone and joint surgery. American volume.

[7]  W. Willems,et al.  Histopathologic evaluation in failed Gerard double cup arthroplasty. , 1988, Clinical orthopaedics and related research.

[8]  J Schatzker,et al.  A study of implant failure in the Wagner resurfacing arthroplasty. , 1985, The Journal of bone and joint surgery. American volume.

[9]  J. Kabo,et al.  Frictional torque in surface and conventional hip replacement. , 1983, The Journal of bone and joint surgery. American volume.