Diagnosis of Subsegmental Pulmonary Emboli: A Multi-Center Study Using a Porcine Model

Abstract: We measured sensitivity, positive predictive value, and free-response receiver operating characteristic (FROC) of 20 radiologists detecting subsegmental-sized pulmonary emboli in a porcine model using either contrast-enhanced computed tomography (CT) or digital subtraction (DS) pulmonary angiography. Colored methacrylate beads (4.2 and 3.8 mm diameter) were injected into 9 anesthetized juvenile pigs. CT and DS pulmonary angiography images were obtained before and after a pulmonary infiltrate was introduced into the lower lobes. Following imaging, the pigs were euthanized, and the pulmonary arterial tree was cast using clear methacrylate allowing direct visualization of emboli. The 20 radiologists used a custom-made computer application to display the images on their personal computer and record their diagnoses. The results were mailed electronically to the coordinating center for comparison with the cast of the pulmonary vasculature. Twenty-three emboli were included in the statistical analysis. Overall sensitivity for spiral CT and angiography, respectively, was: 60 ± 18% and 72 ± 11% (P = 0.06). Positive predictive value for spiral CT and angiography, respectively, was: 49 ± 24% and 58 ± 23% (P = 0.25). There was a large variation in both sensitivity and positive predicted values between Readers. There was no difference in sensitivity or positive predictive value between radiologists from community or academic centers (P > 0.27). FROC analysis showed no significant difference between CT or DS (P = 0.27). In conclusion, in this porcine model, there is no overall diagnostic advantage to using DS pulmonary angiography rather than contrast-enhanced spiral CT for the diagnosis of PE when images are interpreted by radiologists located in either academic or community hospital settings.

[1]  Dev P Chakraborty,et al.  Observer studies involving detection and localization: modeling, analysis, and validation. , 2004, Medical physics.

[2]  U Joseph Schoepf,et al.  CT angiography for diagnosis of pulmonary embolism: state of the art. , 2004, Radiology.

[3]  M. Prins,et al.  Single-Detector Helical Computed Tomography as the Primary Diagnostic Test in Suspected Pulmonary Embolism: A Multicenter Clinical Management Study of 510 Patients , 2003, Annals of Internal Medicine.

[4]  P. Grenier,et al.  Diagnostic strategy for patients with suspected pulmonary embolism: a prospective multicentre outcome study , 2002, The Lancet.

[5]  S. Swensen,et al.  Outcomes after withholding anticoagulation from patients with suspected acute pulmonary embolism and negative computed tomographic findings: a cohort study. , 2002, Mayo Clinic proceedings.

[6]  Alexander Crispin,et al.  Subsegmental pulmonary emboli: improved detection with thin-collimation multi-detector row spiral CT. , 2002, Radiology.

[7]  B Ghaye,et al.  Peripheral pulmonary arteries: how far in the lung does multi-detector row spiral CT allow analysis? , 2001, Radiology.

[8]  H. Coxson,et al.  Pulmonary embolism: comparison of gadolinium-enhanced MR angiography with contrast-enhanced spiral CT in a porcine model. , 2001, Academic radiology.

[9]  A Vieillard-Baron,et al.  Pulmonary embolism detection: prospective evaluation of dual-section helical CT versus selective pulmonary arteriography in 157 patients. , 2000, Radiology.

[10]  G G King,et al.  Spiral computed tomography is comparable to angiography for the diagnosis of pulmonary embolism. , 2000, American journal of respiratory and critical care medicine.

[11]  J. Philbrick,et al.  The role of spiral volumetric computed tomography in the diagnosis of pulmonary embolism. , 2000, Archives of internal medicine.

[12]  G. Raskob,et al.  Sensitivity and Specificity of Helical Computed Tomography in the Diagnosis of Pulmonary Embolism , 2000, Annals of Internal Medicine.

[13]  N L Müller,et al.  Clinically suspected pulmonary embolism: utility of spiral CT. , 1999, Radiology.

[14]  J. Leyendecker,et al.  Effect of anatomic distribution of pulmonary emboli on interobserver agreement in the interpretation of pulmonary angiography. , 1998, AJR. American journal of roentgenology.

[15]  N L Müller,et al.  Pulmonary embolism: prospective comparison of spiral CT with ventilation-perfusion scintigraphy. , 1997, Radiology.

[16]  A. van Rossum,et al.  Pulmonary embolism: validation of spiral CT angiography in 149 patients. , 1996, Radiology.

[17]  X Marchandise,et al.  Diagnosis of pulmonary embolism with spiral CT: comparison with pulmonary angiography and scintigraphy. , 1996, Radiology.

[18]  K. Berbaum,et al.  Receiver operating characteristic rating analysis. Generalization to the population of readers and patients with the jackknife method. , 1992, Investigative radiology.

[19]  A. Alavi,et al.  Complications and Validity of Pulmonary Angiography in Acute Pulmonary Embolism , 1992, Circulation.

[20]  D. Chakraborty,et al.  Free-response methodology: alternate analysis and a new observer-performance experiment. , 1990, Radiology.

[21]  P. Investigators,et al.  Value of the ventilation/perfusion scan in acute pulmonary embolism. Results of the prospective investigation of pulmonary embolism diagnosis (PIOPED). , 1990, JAMA.

[22]  D P Chakraborty,et al.  Maximum likelihood analysis of free-response receiver operating characteristic (FROC) data. , 1989, Medical physics.

[23]  R. G. Fraser,et al.  Digital and conventional chest imaging: a modified ROC study of observer performance using simulated nodules. , 1986, Radiology.

[24]  K. Horsfield,et al.  Morphometry of the Small Pulmonary Arteries in Man , 1978, Circulation research.

[25]  E. P. King Pulmonary embolism. , 2001, South African medical journal = Suid-Afrikaanse tydskrif vir geneeskunde.