Change in the angle of the anterior genu is associated with occlusion rate after Pipeline flow diversion.

OBJECTIVE: The Pipeline embolization device (PED) is a self-expanding mesh stent that diverts blood flow away from an aneurysm; it has been successfully used to treat aneurysms of the proximal internal carotid artery (ICA). PEDs have a remarkable ability to alter regional blood flow along the tortuous segments of the ICA and were incidentally found to alter the angle of the anterior genu after treatment. The authors quantified these changes and explored their implications as they relate to treatment effect. METHODS: The authors retrospectively reviewed cases of aneurysms treated with a PED between the ophthalmic and posterior communicating arteries from 2012 through 2015. The angles of the anterior genu were measured on the lateral projections of cerebral angiograms obtained before and after treatment with a PED. The angles of the anterior genu of patients without aneurysms were used as normal controls. RESULTS: Thirty-eight patients were identified who had been treated with a PED; 34 (89.5%) had complete obliteration and 4 (10.5%) had persistence of their aneurysm at last follow-up (mean 11.3 months). After treatment, 32 patients had an increase, 3 had a decrease, and 3 had no change in the angle of the anterior genu. The average measured angle of the anterior genu was 36.7° before treatment and 44.3° after treatment (p < 0.0001). The average angle of the anterior genu of control patients was 43.32° (vs 36.7° for the preoperative angle in the patients with aneurysms, p < 0.057). The average change in the angle of patients with postoperative Raymond scores of 1 was 9.10°, as compared with 1.25° in patients with postoperative Raymond scores > 1 (p < 0.001). CONCLUSIONS: Treatment with a PED significantly changes the angle of the anterior genu. An average change of 9.1° was associated with complete obliteration of treated aneurysms. These findings have important implications for the treatment and management of cerebral aneurysm.

[1]  S. Miyamoto,et al.  Contribution of the straightening effect of the parent artery to decreased recanalization in stent-assisted coiling of large aneurysms. , 2017, Journal of neurosurgery.

[2]  D. Kallmes,et al.  Histopathological findings following pipeline embolization in a human cerebral aneurysm at the basilar tip , 2016, Interventional neuroradiology : journal of peritherapeutic neuroradiology, surgical procedures and related neurosciences.

[3]  R. Tamargo,et al.  Classification of cavernous internal carotid artery tortuosity: a predictor of procedural complexity in Pipeline embolization , 2014, Journal of NeuroInterventional Surgery.

[4]  Adnan H Siddiqui,et al.  Pipeline for uncoilable or failed aneurysms: results from a multicenter clinical trial. , 2013, Radiology.

[5]  M. Valença,et al.  Carotid siphon geometry and variants of the circle of Willis in the origin of carotid aneurysms. , 2012, Arquivos de neuro-psiquiatria.

[6]  Sebastian Fischer,et al.  Pipeline embolization device (PED) for neurovascular reconstruction: initial experience in the treatment of 101 intracranial aneurysms and dissections , 2011, Neuroradiology.

[7]  J T Keller,et al.  Segments of the internal carotid artery: a new classification. , 1996, Neurosurgery.

[8]  H. Woo,et al.  Long-Term Clinical and Angiographic Outcomes Following Pipeline Embolization Device Treatment of Complex Internal Carotid Artery Aneurysms: Five-Year Results of the Pipeline for Uncoilable or Failed Aneurysms Trial , 2017, Neurosurgery.

[9]  Makoto Yamamoto,et al.  Relationships between geometrical parameters and mechanical properties for a helical braided flow diverter stent. , 2017, Technology and health care : official journal of the European Society for Engineering and Medicine.

[10]  Lu Wang,et al.  Mechanical characteristics of novel polyester/NiTi wires braided composite stent for the medical application , 2016 .