Effect of hinge gap width on the microflow structures in 27-mm bileaflet mechanical heart valves.

BACKGROUND AND AIM OF THE STUDY Most bileaflet mechanical heart valves (BMHVs) incorporate some retrograde flow through their hinge mechanism to prevent flow stasis and inhibit microthrombus formation. This reverse flow is characterized by high velocities and shear stresses, thereby promoting platelet activation and hemolysis inside the hinge region. In the present study, the thromboembolic potential of three 27-mm BMHVs with varying hinge gap widths was assessed via in-vitro characterization of the hinge microflow structures. METHODS Three 27-mm BMHV prototypes with different hinge gap widths (50, 100, and 200 microm) were provided by St. Jude Medical Inc. The valves were mounted in the mitral position of a left heart flow simulator, and two-dimensional laser Doppler velocimetry was used to measure the hinge velocity fields. RESULTS All three valve prototypes revealed Reynolds shear stress (RSS) levels above 2000 dynes/cm2, which exceeded the threshold for platelet activation and hemolysis. The hinge flow fields were characterized by leakage jets during systole, and a strong vortical flow during diastole. The leakage jet size and corresponding RSS levels were found to increase with the hinge gap width. All three gap widths had RSS >4000 dynes/cm2 (range: 5640 to 13,315 dynes/cm2). The hinge with the smallest gap width registered the highest jet velocity magnitude (2.08 m/s) during systole. CONCLUSION The study results showed that the hinge gap width influences washout and RSS levels inside the hinge recess. The 100-microm hinge gap width provided optimum fluid dynamic performance. In contrast, the two valves with large and small hinge gap widths may have higher thromboembolic potential.

[1]  A P Yoganathan,et al.  A microstructural flow analysis within a bileaflet mechanical heart valve hinge. , 1996, The Journal of heart valve disease.

[2]  Joseph D. Bronzino,et al.  The Biomedical Engineering Handbook , 1995 .

[3]  Brandon R. Travis,et al.  An In Vitro Study of the Hinge and Near-Field Forward Flow Dynamics of the St. Jude Medical® Regent™ Bileaflet Mechanical Heart Valve , 2004, Annals of Biomedical Engineering.

[4]  R M Hochmuth,et al.  Shear-induced aggregation and lysis of platelets. , 1976, Transactions - American Society for Artificial Internal Organs.

[5]  A P Yoganathan,et al.  Velocity measurements and flow patterns within the hinge region of a Medtronic Parallel bileaflet mechanical valve with clear housing. , 1996, The Journal of heart valve disease.

[6]  L Zuckerman,et al.  Shear-induced activation of platelets. , 1979, Journal of biomechanics.

[7]  N H Hwang,et al.  Human red blood cell hemolysis in a turbulent shear flow: contribution of Reynolds shear stresses. , 1984, Biorheology.

[8]  Jeffrey Thomas Ellis An in vitro investigation of the leakage and hinge flow fields through bileaflet mechanical heart valves and their relevance to thrombogenesis , 1999 .

[9]  Ajit P Yoganathan,et al.  Microflow fields in the hinge region of the CarboMedics bileaflet mechanical heart valve design. , 2002, The Journal of thoracic and cardiovascular surgery.

[10]  Hwa Liang Leo,et al.  Bileaflet Aortic Valve Prosthesis Pivot Geometry Influences Platelet Secretion and Anionic Phospholipid Exposure , 2001, Annals of Biomedical Engineering.

[11]  P. J. Drury,et al.  Mechanical and other problems of artificial valves. , 1994, Current topics in pathology. Ergebnisse der Pathologie.

[12]  H Reul,et al.  Leakage flow at mechanical heart valve prostheses: improved washout or increased blood damage? , 1999, The Journal of heart valve disease.

[13]  Brandon Travis The effects of bileaflet prosthesis pivot geometry on turbulence and blood damage potential , 2001 .

[14]  A P Yoganathan,et al.  An in vitro investigation of the retrograde flow fields of two bileaflet mechanical heart valves. , 1996, The Journal of heart valve disease.

[15]  Helene A. Simon Influence of the Implant Location on the Hinge and Leakage Flow Fields Through Bileaflet Mechanical Heart Valves , 2004 .

[16]  F. R. Rosendaal,et al.  Thromboembolic and Bleeding Complications in Patients With Mechanical Heart Valve Prostheses , 1994, Circulation.

[17]  Ajit P. Yoganathan Cardiac Valve Prostheses , 1999 .

[18]  A P Yoganathan,et al.  A comparison of the hinge and near-hinge flow fields of the St Jude medical hemodynamic plus and regent bileaflet mechanical heart valves. , 2000, The Journal of thoracic and cardiovascular surgery.

[19]  P. Lu,et al.  A reevaluation and discussion on the threshold limit for hemolysis in a turbulent shear flow. , 2001, Journal of biomechanics.

[20]  H. Leo An in vitro investigation of the flow fields through bileaflet and polymeric prosthetic heart valves. , 2005 .

[21]  Richard A. Lange,et al.  Prosthetic heart valves. , 1996, The New England journal of medicine.