Volume estimation of multi-density nodules with thoracic CT

The purpose of this work was to quantify the effect of surrounding density on the volumetric assessment of lung nodules in a phantom CT study. Eight synthetic multidensity nodules were manufactured by enclosing spherical cores in larger spheres of double the diameter and with a different uniform density. Different combinations of outer/inner diameters (20/10mm, 10/5mm) and densities (100HU/-630HU, 10HU/- 630HU, -630HU/100HU, -630HU/-10HU) were created. The nodules were placed within an anthropomorphic phantom and scanned with a 16-detector row CT scanner. Ten repeat scans were acquired using exposures of 20, 100, and 200mAs, slice collimations of 16x0.75mm and 16x1.5mm, and pitch of 1.2, and were reconstructed with varying slice thicknesses (three for each collimation) using two reconstruction filters (medium and standard). The volumes of the inner nodule cores were estimated from the reconstructed CT data using a matched-filter approach with templates modeling the characteristics of the multi-density objects. Volume estimation of the inner nodule was assessed using percent bias (PB) and the standard deviation of percent error (SPE). The true volumes of the inner nodules were measured using micro CT imaging. Results show PB values ranging from -12.4 to 2.3% and SPE values ranging from 1.8 to 12.8%. This study indicates that the volume of multi-density nodules can be measured with relatively small percent bias (on the order of ±12% or less) when accounting for the properties of surrounding densities. These findings can provide valuable information for understanding bias and variability in clinical measurements of nodules that also include local biological changes such as inflammation and necrosis.

[1]  Wendy J. Post,et al.  Volumetric measurement of pulmonary nodules at low-dose chest CT: effect of reconstruction setting on measurement variability , 2009, European Radiology.

[2]  L. Tanoue,et al.  Variability of Lung Tumor Measurements on Repeat Computed Tomography Scans Taken Within 15 Minutes , 2012 .

[3]  Ehsan Samei,et al.  Quantitative CT: technique dependence of volume estimation on pulmonary nodules , 2012, Physics in medicine and biology.

[4]  Kyle J Myers,et al.  Noncalcified lung nodules: volumetric assessment with thoracic CT. , 2009, Radiology.

[5]  Kazuo Awai,et al.  Computer-aided volumetry of pulmonary nodules exhibiting ground-glass opacity at MDCT. , 2010, AJR. American journal of roentgenology.

[6]  Kyle J Myers,et al.  Benefit of overlapping reconstruction for improving the quantitative assessment of CT lung nodule volume. , 2013, Academic radiology.

[7]  Kyle J. Myers,et al.  A template-based approach for the analysis of lung nodules in a volumetric CT phantom study , 2009, Medical Imaging.

[8]  Kyle J Myers,et al.  Minimum detectable change in lung nodule volume in a phantom CT study. , 2013, Academic radiology.

[9]  William Pao,et al.  A Pilot Study of Volume Measurement as a Method of Tumor Response Evaluation to Aid Biomarker Development , 2010, Clinical Cancer Research.

[10]  Kyle J. Myers,et al.  Information-Theoretic Approach for Analyzing Bias and Variance in Lung Nodule Size Estimation With CT: A Phantom Study , 2010, IEEE Transactions on Medical Imaging.

[11]  Marios A. Gavrielides,et al.  Evaluation of 1D, 2D and 3D nodule size estimation by radiologists for spherical and non-spherical nodules through CT thoracic phantom imaging , 2011, Medical Imaging.