Continuous bed motion acquisition for an LSO PET/CT scanner

Standard whole-body PET/CT protocols collect PET emission data for a sequence of discrete, overlapping bed positions. An alternative acquisition mode in which the patient bed moves continuously through the scanner has some significant advantages, including uniform axial signal-to-noise, elimination of resolution artifacts by sampling continuously in the axial direction, and a reduction in both noise from detector normalization and artifacts from small patient movements. To validate this approach, continuous bed motion acquisition has been implemented on a high resolution, 16-slice LSO PET/CT scanner (CPS Innovations, Knoxville, TN). The emission data are acquired in list mode with the bed moving at a constant velocity in the range 0.3-0.7 mm/s for a total scan time in patients of around 20 minutes covering an axial extent of 36-84 cm. The absolute position of the bed is read and inserted directly into the list mode data stream. Following acquisition, the emission data are rebinned into a fully 3D data set and reconstructed using a 3D OSEM algorithm. Continuous movement of the patient bed yields uniform signal-to-noise throughout the axial imaging field. For short imaging times, the bed movement in the standard acquisition becomes a significant fraction of the total scan time and continuous bed motion acquisition offers a more efficient alternative. Results are presented for some representative patient studies with both step-and-shoot and continuous bed motion acquisition.

[1]  J. Young,et al.  Implementation of true continuous 2D/3D whole body PET scanning , 2000, 2000 IEEE Nuclear Science Symposium. Conference Record (Cat. No.00CH37149).

[2]  T. Gremillion,et al.  LSO PET/CT Pico performance improvements with ultra Hi-Rez option , 2004, IEEE Symposium Conference Record Nuclear Science 2004..

[3]  Richard E. Carson,et al.  Whole body PET using overlapped 3D acquisition and weighted image summation , 1996, 1996 IEEE Nuclear Science Symposium. Conference Record.

[4]  Arion F. Chatziioannou,et al.  Resolution characterization of continuous axial sampling in whole body PET , 1995, 1995 IEEE Nuclear Science Symposium and Medical Imaging Conference Record.

[5]  J. H. Reed,et al.  Continuous bed motion acquisition on a whole body combined PET/CT system , 2002, 2002 IEEE Nuclear Science Symposium Conference Record.

[6]  Magnus Dahlbom,et al.  Characterization of sampling schemes for whole body PET imaging , 1993 .

[7]  Andrew J. Reader,et al.  Spiral whole-body PET - Implementation and initial results. , 1999 .

[8]  Paul Kinahan,et al.  Attenuation correction for a combined 3D PET/CT scanner. , 1998, Medical physics.

[9]  Keishi Kitamura,et al.  Implementation of continuous 3D whole-body PET scanning using on-the-fly Fourier rebinning. , 2002, Physics in medicine and biology.

[10]  E. Hoffman,et al.  Methods for improving image quality in whole body PET scanning , 1991, Conference Record of the 1991 IEEE Nuclear Science Symposium and Medical Imaging Conference.

[11]  Arion F. Chatziioannou,et al.  Study of the effects of whole body PET spatial sampling schemes on data SNR , 1996, 1996 IEEE Nuclear Science Symposium. Conference Record.