MR-compatible polyetheretherketone-based guide wire assisting MR-guided stenting of iliac and supraaortic arteries in swine: Feasibility study

The purpose of this study was to demonstrate first magnetic resonance (MR)-guided stenting of iliac and supraaortic arteries using a polyetheretherketone-based (PEEK) MR-compatible guide wire. In vitro and animal experiments were performed in a short magnet wide-bore scanner (1.5 Tesla, Espree®, Siemens Healthcare, Erlangen, Germany). For all experiments, a 0.035″ MR-compatible guide wire prototoype was used. This wire had a compound core of PEEK with reinforcing fibres, a soft and atraumatic tip and a hydrophilic coating. For its passive visualization, paramagnetic markings were attached. All experiments were performed through a vascular introducer sheath under MR-guidance. In vitro repetitive selective over the wire catheterizations of either the right carotid artery and the left subclavian artery were performed. In vivo, selective catheterization and over-the-wire stenting of the brachiocephalic trunc and the left subclavian artery were performed. The common iliac arteries were catheterized retrogradely (left) and cross-over (right). Angioplasty and stenting were performed over-the-wire. All procedures were successful. Visibility of the PEEK-based guide-wire was rated good in vitro and acceptable in vivo. Guide wire pushability and endovascular device support were good. The PEEK-based MR-compatible guide wire is well visible and usable under MR-guidance. It supports over-the-wire treatment of iliac arteries and supraaortic arteries.

[1]  D. Bilecen,et al.  MR angiography with blood pool contrast agents , 2007, European Radiology.

[2]  J. Meaney,et al.  Recent advances in contrast-enhanced magnetic resonance angiography. , 2007, European radiology.

[3]  Simon Wildermuth,et al.  MR-guided percutaneous angioplasty: Assessment of tracking safety, catheter handling and functionality , 2009, CardioVascular and Interventional Radiology.

[4]  Thomas Kahn,et al.  Interventional and intraoperative MR: review and update of techniques and clinical experience , 2004, European Radiology.

[5]  Ogan Ocali,et al.  Intravascular magnetic resonance imaging using a loopless catheter antenna , 1997, Magnetic resonance in medicine.

[6]  M. Bock,et al.  [An algorithm for passive marker localization in interventional MRI]. , 2007, Zeitschrift fur medizinische Physik.

[7]  W. Nitz,et al.  On the heating of linear conductive structures as guide wires and catheters in interventional MRI , 2001, Journal of magnetic resonance imaging : JMRI.

[8]  M. Bock,et al.  MR‐guided intravascular interventions: Techniques and applications , 2008, Journal of magnetic resonance imaging : JMRI.

[9]  Michael O Zenge,et al.  Interventional magnetic resonance angiography with no strings attached: Wireless active catheter visualization , 2005, Magnetic resonance in medicine.

[10]  R W Günther,et al.  Real-time MR Guidance for inferior vena cava filter placement in an animal model. , 2001, Journal of vascular and interventional radiology : JVIR.

[11]  Mark E Ladd,et al.  Interventional MRA using actively visualized catheters, TrueFISP, and real‐time image fusion , 2003, Magnetic resonance in medicine.

[12]  R. Günther,et al.  Magnetic Resonance-Guided Angioplasty With Delivery of Contrast-Media Doped Solutions to the Vessel Wall: An Experimental Study in Swine , 2008, Investigative radiology.

[13]  Deniz Bilecen,et al.  MR-guided endovascular interventions: a comprehensive review on techniques and applications , 2008, European Radiology.

[14]  R Frayne,et al.  MR-guided angioplasty of renal artery stenosis in a pig model: a feasibility study. , 2000, Journal of vascular and interventional radiology : JVIR.

[15]  T. Paunesku,et al.  Feasibility of blood oxygenation level-dependent MR imaging to monitor hepatic transcatheter arterial embolization in rabbits. , 2005, Journal of vascular and interventional radiology : JVIR.

[16]  M. Bock,et al.  Ein Algorithmus zur Lokalisation von passiven Markersystemen in der interventionellen Magnetresonanztomographie , 2007 .

[17]  Klaus Scheffler,et al.  A polymer‐based MR‐compatible guidewire: A study to explore new prospects for interventional peripheral magnetic resonance angiography (ipMRA) , 2006, Journal of magnetic resonance imaging : JMRI.

[18]  Raimund Erbel,et al.  Feasibility of real-time magnetic resonance-guided stent-graft placement in a swine model of descending aortic dissection. , 2006, European heart journal.

[19]  H. Quick,et al.  Anesthesia for cardiovascular interventions and magnetic resonance imaging in pigs. , 2007, Journal of the American Association for Laboratory Animal Science : JAALAS.

[20]  J. Lewin,et al.  MR-guided endovascular interventions: device visualization, tracking, navigation, clinical applications, and safety aspects. , 2005, Magnetic resonance imaging clinics of North America.

[21]  James A. Gehl,et al.  MR imaging- versus conventional X-ray fluoroscopy-guided renal angioplasty in swine: prospective randomized comparison. , 2006, Radiology.

[22]  Frank G. Shellock,et al.  Reference Manual for Magnetic Resonance Safety, Implants, and Devices , 2009 .

[23]  Cengizhan Ozturk,et al.  Magnetic Resonance Imaging-guided Vascular Interventions , 2005, Topics in magnetic resonance imaging : TMRI.

[24]  W. Nitz,et al.  [Stent angioplasty of pelvic artery stenosis with MRI control: initial clinical results]. , 2000, RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin.

[25]  Harald H. Quick,et al.  First Magnetic Resonance Imaging-Guided Aortic Stenting and Cava Filter Placement Using a Polyetheretherketone-Based Magnetic Resonance Imaging-Compatible Guidewire in Swine: Proof of Concept , 2009, CardioVascular and Interventional Radiology.

[26]  W. Rogers,et al.  Real-time magnetic resonance-guided placement of retrievable inferior vena cava filters: comparison with fluoroscopic guidance with use of in vitro and animal models. , 2006, Journal of vascular and interventional radiology : JVIR.

[27]  C. Dumoulin,et al.  Magnetic resonance angiography. , 1986, Radiology.

[28]  Tobias Schaeffter,et al.  Simultaneous real‐time visualization of the catheter tip and vascular anatomy for MR‐guided PTA of iliac arteries in an animal model , 2002, Journal of magnetic resonance imaging : JMRI.

[29]  Charles L Dumoulin,et al.  Feasibility of stent placement in carotid arteries with real-time MR imaging guidance in pigs. , 2005, Radiology.

[30]  Andreas Melzer,et al.  Improvement of the MR imaging behavior of vascular implants , 2006, Minimally invasive therapy & allied technologies : MITAT : official journal of the Society for Minimally Invasive Therapy.

[31]  E. Spuentrup,et al.  An MR guidewire based on micropultruded fiber‐reinforced material , 2008, Magnetic resonance in medicine.

[32]  Gregor Schaefers,et al.  Testing methods for MR safety and compatibility of medical devices , 2006, Minimally invasive therapy & allied technologies : MITAT : official journal of the Society for Minimally Invasive Therapy.