A novel approach for realization of primary vibration calibration standard by homodyne laser interferometer in frequency range of 0.1 Hz to 20 kHz

Abstract The paper discusses the primary vibration calibration standard of NPL, India capable of calibrating the reference accelerometers in frequency range from 0.1 Hz to 20 kHz as per ISO 16063-11. The excitation subsystem produces constant vibration at a specified amplitude and frequency, while the measurement system uses NI interface for measuring the quadrature output. The acceleration level and voltage level at the calibration frequency f is determined by applying a Discrete Fourier Transform to the voltage and displacement signals, and then examining the spectral component at frequency f . A PC-based data acquisition system acquires the accelerometer voltage signal and analog quadrature interferometer photodetector signal pair as well as a digital quadrature pair whereby the software processes the demodulated photodetector signals to reconstruct the armature displacement. The validation of the calibration results for standard reference accelerometers with manufacturer results and uncertainty in calibration in entire frequency range 0.1 Hz to 20 kHz is reported in the present work.

[1]  Andrew Lewis,et al.  Advice from the CCL on the use of unstabilized lasers as standards of wavelength: the helium–neon laser at 633 nm , 2009 .

[2]  C S Veldman A novel implementation of an ISO standard method for primary vibration calibration by laser interferometry , 2003 .

[3]  Takashi Usuda,et al.  Correction and evaluation of the effect due to parasitic motion on primary accelerometer calibration , 2010 .

[4]  Gwo-Sheng Peng,et al.  Correction of nonlinearity in one-frequency optical interferometry , 1996 .

[5]  Yong-Bong Lee,et al.  Determination of the sensitivity phase of an accelerometer based on an analysis of the harmonic components of the interference signal , 2008 .

[6]  Hans-Jürgen von Martens Evaluation of uncertainty in measurements—problems and tools , 2002 .

[7]  P L Heydemann,et al.  Determination and correction of quadrature fringe measurement errors in interferometers. , 1981, Applied optics.

[8]  Richard W. Bono,et al.  The Effect of High Transverse Inputs on Accelerometer Calibration , 2010 .

[9]  Richard W. Bono,et al.  IMPROVED LOW FREQUENCY ACCELEROMETER CALIBRATION , 2009 .

[10]  S. R. Olsen,et al.  An in situ rapid heat–quench cell for small-angle neutron scattering , 2008 .

[11]  Qiao Sun,et al.  Modifications of the sine-approximation method for primary vibration calibration by heterodyne interferometry , 2009 .

[12]  Robert D. Sill,et al.  A New Solution for Shock and Vibration Calibration of Accelerometers , 2007 .