Evaluation of the Alphasense optical particle counter (OPC-N2) and the Grimm portable aerosol spectrometer (PAS-1.108)

Abstract We compared the performance of a low-cost (∼$500), compact optical particle counter (OPC, OPC-N2, Alphasense) to another OPC (PAS-1.108, Grimm Technologies) and reference instruments. We measured the detection efficiency of the OPCs by size from 0.5 to 5 µm for monodispersed, polystyrene latex (PSL) spheres. We then compared number and mass concentrations measured with the OPCs to those measured with reference instruments for three aerosols: salt, welding fume, and Arizona road dust. The OPC-N2 detection efficiency was similar to the PAS-1.108 for particles larger than 0.8 µm (minimum of 79% at 1 µm and maximum of 101% at 3 µm). For 0.5-µm particles, the detection efficiency of the OPC-N2 was underestimated at 78%, whereas PAS-1.108 overestimated concentrations by 183%. The mass concentrations from the OPCs were linear (r ≥ 0.97) with those from the reference instruments for all aerosols, although the slope and intercept were different. The mass concentrations were overestimated for dust (OPC-N2, slope = 1.6; PAS-1.108, slope = 2.7) and underestimated for welding fume (OPC-N2, slope = 0.05; PAS-1.108, slope = 0.4). The coefficient of variation (CV, precision) for OPC-N2 for all experiments was between 4.2% and 16%. These findings suggest that, given site-specific calibrations, the OPC-N2 can provide number and mass concentrations similar to the PAS-1.108 for particles larger than 1 µm. Copyright © 2016 American Association for Aerosol Research

[1]  Richard C. Flagan,et al.  Scanning Electrical Mobility Spectrometer , 1989 .

[2]  P. Hewett,et al.  The particle size distribution, density, and specific surface area of welding fumes from SMAW and GMAW mild and stainless steel consumables. , 1995, American Industrial Hygiene Association journal.

[3]  Wladyslaw W. Szymanski,et al.  On the Sizing Accuracy of Laser Optical Particle Counters , 1986 .

[4]  Geb Thomas,et al.  Inter-comparison of low-cost sensors for measuring the mass concentration of occupational aerosols , 2016, Aerosol science and technology : the journal of the American Association for Aerosol Research.

[5]  Joshua P. Schwarz,et al.  A light-weight, high-sensitivity particle spectrometer for PM2.5 aerosol measurements , 2016 .

[6]  R. Oehmig Principles, methods, and application of particle size analysis , 1992 .

[7]  R. Harrison,et al.  What is responsible for the carcinogenicity of PM2.5? , 2004, Occupational and Environmental Medicine.

[8]  K. Teschke,et al.  Wood dust exposure and risk of lung cancer , 2010, Occupational and Environmental Medicine.

[9]  D. Dockery,et al.  An association between air pollution and mortality in six U.S. cities. , 1993, The New England journal of medicine.

[10]  Yu-Hsiang Cheng,et al.  Comparison of the TSI Model 8520 and Grimm Series 1.108 Portable Aerosol Instruments Used to Monitor Particulate Matter in an Iron Foundry , 2008, Journal of occupational and environmental hygiene.

[11]  M. L. Laucks,et al.  Aerosol Technology Properties, Behavior, and Measurement of Airborne Particles , 2000 .

[12]  Jean-Baptiste Renard,et al.  Small-angle light scattering by airborne particulates: Environnement S.A. continuous particulate monitor , 2010 .

[13]  D. Landen,et al.  Coal dust exposure and mortality from ischemic heart disease among a cohort of U.S. coal miners. , 2011, American journal of industrial medicine.

[14]  G. Mulholland,et al.  Structural Property Effect of Nanoparticle Agglomerates on Particle Penetration through Fibrous Filter , 2009 .

[15]  Bean T. Chen,et al.  Size Distribution and Estimated Respiratory Deposition of Total Chromium, Hexavalent Chromium, Manganese, and Nickel in Gas Metal Arc Welding Fume Aerosols , 2014, Aerosol science and technology : the journal of the American Association for Aerosol Research.

[16]  R. Luukkonen,et al.  Inflammatory response to acute exposure to welding fumes during the working day , 2013, International journal of occupational medicine and environmental health.

[17]  Ulrich Franck,et al.  The effect of particle size on cardiovascular disorders--the smaller the worse. , 2011, The Science of the total environment.

[18]  J. L. Parker,et al.  Particle size distribution and composition in a mechanically ventilated school building during air pollution episodes. , 2008, Indoor air.

[19]  J. Schwartz,et al.  Fine particles are more strongly associated than coarse particles with acute respiratory health effects in schoolchildren. , 2000, Epidemiology.

[20]  J. Schwartz,et al.  The Effect of Fine and Coarse Particulate Air Pollution on Mortality: A National Analysis , 2009, Environmental health perspectives.

[21]  James S. Brown,et al.  Ultrafine particle deposition and clearance in the healthy and obstructed lung. , 2002, American journal of respiratory and critical care medicine.

[22]  Alessandra Spinali,et al.  Effects of particulate matter (PM(10), PM(2.5) and PM(1)) on the cardiovascular system. , 2009, Toxicology.

[23]  Giovanni Battista Bartolucci,et al.  Field comparison of a personal cascade impactor sampler, an optical particle counter and CEN-EU standard methods for PM10, PM2.5 and PM1 measurement in urban environment , 2013 .

[24]  Benjamin Y. H. Liu,et al.  Experimental studies of optical particle counters , 1974 .

[25]  Daniel Krewski,et al.  Health Effects Associated With Exposure to Ambient Air Pollution , 2007, Journal of toxicology and environmental health. Part A.

[26]  F. Belosi,et al.  Performance Evaluation of Four Commercial Optical Particle Counters , 2013 .

[27]  J. Schwartz,et al.  Is Daily Mortality Associated Specifically with Fine Particles? , 1996, Journal of the Air & Waste Management Association.

[28]  C. Rott,et al.  The Annals of Occupational Hygiene , 2010 .

[29]  D. Leith,et al.  Concentration measurement and counting efficiency for the aerodynamic particle sizer 3320 , 2002 .

[30]  M. Neuberger,et al.  Characterizing the performance of two optical particle counters (Grimm OPC1.108 and OPC1.109) under urban aerosol conditions , 2010, Journal of aerosol science.

[31]  Darrin K. Ott,et al.  Comparison of the Grimm 1.108 and 1.109 portable aerosol spectrometer to the TSI 3321 aerodynamic particle sizer for dry particles. , 2006, The Annals of occupational hygiene.

[32]  Hans Grimm,et al.  Aerosol Measurement: The Use of Optical Light Scattering for the Determination of Particulate Size Distribution, and Particulate Mass, Including the Semi-Volatile Fraction , 2009, Journal of the Air & Waste Management Association.

[33]  R. Burnett,et al.  Lung cancer, cardiopulmonary mortality, and long-term exposure to fine particulate air pollution. , 2002, JAMA.

[34]  J. Antonini,et al.  Health Effects of Welding , 2003, Critical reviews in toxicology.

[35]  K. T. Whitby,et al.  Aerosol classification by electric mobility: apparatus, theory, and applications , 1975 .

[36]  L. Bernstein,et al.  Associations of Mortality with Long-Term Exposures to Fine and Ultrafine Particles, Species and Sources: Results from the California Teachers Study Cohort , 2015, Environmental health perspectives.

[37]  V. Grassian,et al.  A laboratory investigation of light scattering from representative components of mineral dust aerosol at a wavelength of 550 nm , 2008 .

[38]  R. Burnett,et al.  Effects of particulate and gaseous air pollution on cardiorespiratory hospitalizations. , 1999, Archives of environmental health.

[39]  Phillip L. Williams,et al.  A comparison of two direct-reading aerosol monitors with the federal reference method for PM2.5 in indoor air , 2002 .