Mechanical stresses on closed cusps of porcine bioprosthetic valves: correlation with sites of calcification.

The purpose of this study was to (1) evaluate the magnitude and distribution of mechanical stresses on closed leaflets of porcine bioprosthetic valves (PBVs) by using a nonlinear, finite-element model, and (2) determine if a relationship exists between the magnitude of stresses at various sites on the leaflet and the incidence of calcification at those sites. Mechanical stresses were found to be highest near the commissures and lowest at the base of the leaflet. At a pressure of 100 mm Hg applied to the closed valve, the maximal principal normal stress was 160 kPa near the commissures, 140 kPa in the body of the leaflet, 70 kPa at the free margin, and 60 kPa near the base. Similarly, the maximal shear stress, at an applied pressure of 100 mm Hg, was 80 kPa near the commissures, 70 kPa in the body of the leaflet, 50 kPa at the free margin, and 40 kPa near the base. This distribution of mechanical stresses on the PBV leaflet coincided with the incidence of calcification of the various regions of the leaflet. Calcification was found most frequently near the commissures, less frequently in the body of the leaflet and free margin, and least frequently at the base. These observations suggest a possible causative relation between the magnitude of mechanical stresses and the site of calcification of PBV leaflets.

[1]  N. Broom,et al.  Fatigue-induced damage in glutaraldehyde-preserved heart valve tissue. , 1978, The Journal of thoracic and cardiovascular surgery.

[2]  R E Clark,et al.  Stress analysis of aortic valve leaflets with smoothed geometrical data. , 1977, Journal of biomechanics.

[3]  L. Cohn,et al.  Long-term failure rate and morphologic correlations in porcine bioprosthetic heart valves. , 1983, The American journal of cardiology.

[4]  W. Roberts,et al.  Comparison of late degenerative changes in porcine bioprostheses in the mitral and aortic valve position in the same patient. , 1983, The American journal of cardiology.

[5]  P D Stein,et al.  Relation of calcification to torn leaflets of spontaneously degenerated porcine bioprosthetic valves. , 1985, The Annals of thoracic surgery.

[6]  W. Roberts,et al.  Structural changes in glutaraldehyde-treated porcine heterografts used as substitute cardiac valves. Transmission and scanning electron microscopic observations in 12 patients. , 1978, The American journal of cardiology.

[7]  M Jones,et al.  Calcific deposits in porcine bioprostheses: structure and pathogenesis. , 1980, The American journal of cardiology.

[8]  E. Arbustini,et al.  Calcific degeneration as the main cause of porcine bioprosthetic valve failure. , 1984, The American journal of cardiology.

[9]  W. Roberts,et al.  Comparison of late (62 to 140 months) degenerative changes in simultaneously implanted and explanted mitral valve positions in six patients , 1984 .

[10]  M Jones,et al.  Structure and classification of cuspal tears and perforations in porcine bioprosthetic cardiac valves implanted in patients. , 1981, The American journal of cardiology.

[11]  Paul D. Stein,et al.  Comparison of finite element stress analysis of aortic valve leaflet using either membrane elements or solid elements , 1985 .

[12]  Y. Missirlis,et al.  Stress analysis of the aortic valve during diastole: important parameters. , 1976, Journal of biomechanics.

[13]  M. Thubrikar,et al.  Role of mechanical stress in calcification of aortic bioprosthetic valves. , 1983, The Journal of thoracic and cardiovascular surgery.

[14]  M. S. Hamid,et al.  Estimation of mechanical stresses on closed cusps of porcine bioprosthetic valves: effects of stiffening, focal calcium and focal thinning. , 1985, The American journal of cardiology.