Heating of the Eye by a Retinal Prosthesis: Modeling, Cadaver and In Vivo Study

In order to develop retinal implants with a large number of electrodes, it is necessary to ensure that they do not cause damage to the neural tissue by the heat that the electrical circuits generate. Knowledge of the amount of power that induces thermal damage will assist in development of power budgets for implants, which has a significant effect upon the design of the prostheses circuitry. In this study, temperatures were measured at multiple locations on the retina while the retina was heated in cadaver and in vivo preparations using a variety of prosthesis implantation sites. A finite element thermal model of the cat eye was also created and validated by the cadaver and in vivo tests, allowing for a much larger spectrum of thermal influences to be evaluated without additional animal experimentation. To ensure that retinal tissue temperatures are not increased by more than 2 °C, a 5 mm × 5 mm, suprachoroidally implanted heating element must not dissipate more than 135 mW (5.4 mW/mm2).

[1]  E. Y. K. Ng,et al.  FEM simulation of the eye structure with bioheat analysis , 2006, Comput. Methods Programs Biomed..

[2]  James D. Weiland,et al.  Thermal elevation in the human eye and head due to the operation of a retinal prosthesis , 2004, IEEE Transactions on Biomedical Engineering.

[3]  J. Weiland,et al.  Visual performance using a retinal prosthesis in three subjects with retinitis pigmentosa. , 2007, American journal of ophthalmology.

[4]  G. L. Ruskell,et al.  Anatomy and histology of the eye and orbit in domestic animals. , 1960 .

[5]  Gislin Dagnelie,et al.  Facial recognition using simulated prosthetic pixelized vision. , 2003, Investigative ophthalmology & visual science.

[6]  R. Normann,et al.  Thermal Impact of an Active 3-D Microelectrode Array Implanted in the Brain , 2007, IEEE Transactions on Neural Systems and Rehabilitation Engineering.

[7]  E. Y. K. Ng,et al.  Ocular surface temperature: A 3D FEM prediction using bioheat equation , 2007, Comput. Biol. Medicine.

[8]  Akimasa Hirata Improved Heat Transfer Modeling of the Eye for Electromagnetic Wave Exposures , 2007, IEEE Transactions on Biomedical Engineering.

[9]  D. Irvine,et al.  Cochlear implant use following neonatal deafness influences the cochleotopic organization of the primary auditory cortex in cats , 2009, The Journal of comparative neurology.

[10]  J. A. Scott The computation of temperature rises in the human eye induced by infrared radiation. , 1988, Physics in medicine and biology.

[11]  R. P. Benedict,et al.  Fundamentals of temperature, pressure, and flow measurements , 1977 .

[12]  Eberhart Zrenner,et al.  Investigation of thermal effects of infrared lasers on the rabbit retina: a study in the course of development of an active subretinal prosthesis , 2007, Graefe's Archive for Clinical and Experimental Ophthalmology.

[13]  Mark S Humayun,et al.  Heat effects on the retina. , 2003, Ophthalmic surgery, lasers & imaging : the official journal of the International Society for Imaging in the Eye.

[14]  B W Raaymakers,et al.  Modelling the impact of blood flow on the temperature distribution in the human eye and the orbit: fixed heat transfer coefficients versus the Pennes bioheat model versus discrete blood vessels , 2006, Physics in medicine and biology.

[15]  Atul K. Jain,et al.  Effect of pulse duration on size and character of the lesion in retinal photocoagulation. , 2008, Archives of ophthalmology.

[16]  Nicholas Lachlan Opie,et al.  Thermal safety of a retinal prosthesis , 2011 .

[17]  David B. Grayden,et al.  Thermal heating of a retinal prosthesis: Thermal model and in-vitro study , 2010, 2010 Annual International Conference of the IEEE Engineering in Medicine and Biology.

[18]  S. Resnikoff,et al.  Global data on visual impairment in the year 2002. , 2004, Bulletin of the World Health Organization.

[19]  Chris E. Williams,et al.  Evaluation of stimulus parameters and electrode geometry for an effective suprachoroidal retinal prosthesis , 2010, Journal of neural engineering.

[20]  G. Lazzi,et al.  Thermal effects of bioimplants , 2005, IEEE Engineering in Medicine and Biology Magazine.