Tomographic motion detection and correction directly in sinogram space.

Patient motion, especially respiratory motion, results in various artefacts such as blurring and streaks in tomographic images. The interplay of the movement of the beam aperture and variations of organ anatomy during delivery can create 'hot' and 'cold' spots throughout the field in intensity-modulated radiation therapy (IMRT). Detection and correction of patient motion is extremely important in tomographic imaging and IMRT. Tomographic projection data (sinogram) encode not only the patient anatomy information, but also the intra-scanning motion information. In this paper, we developed an algorithm to detect and correct the in-plane respiratory motion directly in sinogram space. The respiratory motion is modelled as time-varying scaling along the x and y directions. Its effects on the sinogram are discussed. Based on the traces of some nodal points in the sinogram, the intra-scanning motion is determined. The motion correction is also implemented in sinogram space. The motion-corrected sinogram is used for reconstruction by the filtered back-projection (FBP) method. Computer simulations validate the motion detection and correction algorithm. The reconstructed images from the motion-corrected sinogram eliminate the majority of the artefacts. The method could be applied to projection data used in CT and ECT, as well as in tomotherapy delivery modification and dose reconstruction.

[1]  R J Jaszczak,et al.  A filtered-backprojection algorithm for fan-beam SPECT which corrects for patient motion. , 1995, Physics in medicine and biology.

[2]  C. Langer,et al.  Concurrent chemoradiation using paclitaxel and carboplatin in locally advanced non-small cell lung cancer. , 1999, Seminars in radiation oncology.

[3]  J. Cooper,et al.  Effect of patient motion on tomographic myocardial perfusion imaging. , 1992, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[4]  R M Henkelman,et al.  The double-fissure sign: a motion artifact on thin-section CT scans. , 1987, Radiology.

[5]  R J Jaszczak,et al.  A filtered backprojection algorithm for axial head motion correction in fan-beam SPECT. , 1995, Physics in medicine and biology.

[6]  S. Deans The Radon Transform and Some of Its Applications , 1983 .

[7]  T. Mackie,et al.  Delivery Modification as an Alternative to Patient Repositioning in Tomotherapy , 2000 .

[8]  L. Axel,et al.  Correction of motion artifacts in linogram and projection reconstruction MRI using geometry and consistency constraints. , 1996, Medical physics.

[9]  D J Conces,et al.  Motion artifacts on CT simulate bronchiectasis. , 1988, AJR. American journal of roentgenology.

[10]  C. J. Ritchie,et al.  Predictive respiratory gating: a new method to reduce motion artifacts on CT scans. , 1994, Radiology.

[11]  Gabor T. Herman,et al.  Image Reconstruction From Projections , 1975, Real Time Imaging.

[12]  E. Botvinick,et al.  A quantitative assessment of patient motion and its effect on myocardial perfusion SPECT images. , 1993, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[13]  K L Lam,et al.  Automated localization of the prostate at the time of treatment using implanted radiopaque markers: technical feasibility. , 1995, International journal of radiation oncology, biology, physics.

[14]  D E Dupuy,et al.  Spiral CT of the pancreas. , 1992, Radiology.

[15]  Paul J. Reckwerdt,et al.  Tomotherapy: Optimized planning and delivery of radiation therapy , 1995, Int. J. Imaging Syst. Technol..

[16]  J O Deasy,et al.  Tomotherapy: a new concept for the delivery of dynamic conformal radiotherapy. , 1993, Medical physics.

[17]  P. R. Bevington,et al.  Data Reduction and Error Analysis for the Physical Sciences , 1969 .

[18]  C. Pelizzari,et al.  Unfolding patient motion with biplane radiographs. , 1994, Medical physics.

[19]  D A Jaffray,et al.  The effects of intra-fraction organ motion on the delivery of dynamic intensity modulation. , 1998, Physics in medicine and biology.

[20]  K. Ruchala,et al.  Image/patient registration from (partial) projection data by the Fourier phase matching method. , 1999, Physics in medicine and biology.

[21]  J O Deasy,et al.  An investigation of tomotherapy beam delivery. , 1997, Medical physics.

[22]  J H Stanley,et al.  Dose minimization in computed tomography overscanning. , 1981, Medical physics.

[23]  Peyman Milanfar A model of the effect of image motion in the Radon transform domain , 1999, IEEE Trans. Image Process..

[24]  B. Paliwal,et al.  Analysis and convergence of the iterative convolution/superposition dose reconstruction technique for multiple treatment beams and tomotherapy. , 1997, Medical physics.

[25]  C R Meyer,et al.  Motion correction in fMRI via registration of individual slices into an anatomical volume , 1999, Magnetic resonance in medicine.

[26]  Cameron J. Ritchie,et al.  Respiratory compensation in projection imaging using a magnification and displacement model , 1996, IEEE Trans. Medical Imaging.

[27]  J. Crook,et al.  Prostate motion during standard radiotherapy as assessed by fiducial markers. , 1995, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[28]  J. Michael Fitzpatrick,et al.  The existence of geometrical density-image transformations corresponding to object motion , 1988, Comput. Vis. Graph. Image Process..

[29]  Yongmin Kim,et al.  Correction of computed tomography motion artifacts using pixel-specific back-projection , 1996, IEEE Trans. Medical Imaging.