Examining Root Causes of Medical Device Alarm System Failures Using Reports From Food and Drug Administration’s Medical Product Safety Network (MedSun)

This article analyzes medical device reports related to medical device alarm system performance issues by categorizing the underlying failure causes and examining alarm system misses and false alarms. When a miss or false alarm occurs, the typical causes of the failure were either device or operator related. The most typical device-related failure was an unexpected component failure, whereas the typical operator-related failure resulted from a human factor performance issue with the device. These results suggest that, by improving the quality and usability of medical devices, the number of problematic false alarms and misses may be reduced.

[1]  M. Makary,et al.  Medical error—the third leading cause of death in the US , 2016, British Medical Journal.

[2]  M. Cvach Monitor alarm fatigue: an integrative review. , 2012, Biomedical instrumentation & technology.

[3]  Frank E Block,et al.  Decreasing False Alarms by Obtaining the Best Signal and Minimizing Artifact from Physiological Sensors. , 2015, Biomedical instrumentation & technology.

[4]  Judy Edworthy,et al.  Medical audible alarms: a review , 2013, J. Am. Medical Informatics Assoc..

[5]  Maria M Cvach,et al.  Daily Electrode Change and Effect on Cardiac Monitor Alarms: An Evidence-Based Practice Approach , 2013, Journal of nursing care quality.

[6]  M. Görges,et al.  Medical device alarms , 2011, Biomedizinische Technik. Biomedical engineering.

[7]  Marjorie Funk,et al.  Attitudes and practices related to clinical alarms. , 2014, American journal of critical care : an official publication, American Association of Critical-Care Nurses.

[9]  K. Jenpanich,et al.  [Drug administration]. , 1976, Thai journal of nursing.

[10]  T. Clark Impact of Clinical Alarms on Patient Safety: A Report From the American College of Clinical Engineering Healthcare Technology Foundation , 2007 .

[11]  B. Drew,et al.  Insights into the Problem of Alarm Fatigue with Physiologic Monitor Devices: A Comprehensive Observational Study of Consecutive Intensive Care Unit Patients , 2014, PloS one.

[12]  James Welch,et al.  An evidence-based approach to reduce nuisance alarms and alarm fatigue. , 2011, Biomedical instrumentation & technology.

[13]  K. Ruskin,et al.  Alarm fatigue: impacts on patient safety , 2015, Current opinion in anaesthesiology.

[14]  C. Goodman Association for the Advancement of Medical Instrumentation , 1988 .

[15]  Frank E Block Understanding Caregiver Workflows: A Proposal for Alarm Condition Priorities. , 2016, Biomedical instrumentation & technology.