Use of Brain MRI after Deep Brain Stimulation Hardware Implantation

ABSTRACT The objective of this study was to examine the experience with and safety of brain 1.5 Tesla (T) magnetic resonance imaging (MRI) in deep brain stimulation (DBS) patients. This was a retrospective review of brain MRI scanning performed on DBS patients at the University of Kansas Medical Center between January 1995 and December 2007. A total of 249 DBS patients underwent 445 brain 1.5 T MRI scan sessions encompassing 1,092 individual scans using a transmit-receive head coil, representing the cumulative scanning of 1,649 DBS leads. Patients with complete implanted DBS systems as well as those with externalized leads underwent brain imaging. For the majority of scans, specific absorption rates localized to the head (SARH) were estimated and in all cases SARH were higher than that specified in the present product labeling. There were no clinical or hardware related adverse events secondary to brain MRI scanning. Our data should not be extrapolated to encourage MRI scanning beyond the present labeling. Rather, our data may contribute to further defining safe MRI scanning parameters that might ultimately be adopted in future product labeling as more centers report in detail their experiences.

[1]  Rajesh Pahwa,et al.  Surgical and hardware complications of subthalamic stimulation , 2004, Neurology.

[2]  Jean A. Tkach,et al.  Variability in RF‐induced heating of a deep brain stimulation implant across MR systems , 2006, Journal of magnetic resonance imaging : JMRI.

[3]  R. Frysinger,et al.  Cortical and subcortical brain shift during stereotactic procedures. , 2007, Journal of neurosurgery.

[4]  Michael S. Okun,et al.  Safety of MRI in patients with implanted deep brain stimulation devices , 2009, NeuroImage.

[5]  A L Benabid,et al.  Neuropsychological changes between “off” and “on” STN or GPi stimulation in Parkinson’s disease , 2000, Neurology.

[6]  Giovanni Calcagnini,et al.  In vitro investigation of pacemaker lead heating induced by magnetic resonance imaging: Role of implant geometry , 2008, Journal of magnetic resonance imaging : JMRI.

[7]  Gordon H Baltuch,et al.  Deep brain stimulation for movement disorders: morbidity and mortality in 109 patients. , 2003, Journal of neurosurgery.

[8]  G. Westby,et al.  Targeting the subthalamic nucleus for deep brain stimulation: technical approach and fusion of pre- and postoperative MR images to define accuracy of lead placement , 2005, Journal of Neurology, Neurosurgery & Psychiatry.

[9]  Calorimetric calibration of head coil SAR estimates displayed on a clinical MR scanner. , 2008, Physics in medicine and biology.

[10]  Wolfgang Reith,et al.  Transient dystonia following magnetic resonance imaging in a patient with deep brain stimulation electrodes for the treatment of Parkinson disease. Case report. , 2003, Journal of neurosurgery.

[11]  Terry O Woods,et al.  Standards for medical devices in MRI: Present and future , 2007, Journal of magnetic resonance imaging : JMRI.

[12]  W. Kainz,et al.  Complexity of MRI induced heating on metallic leads: Experimental measurements of 374 configurations , 2008, Biomedical engineering online.

[13]  Alastair J. Martin,et al.  Magnetic Resonance Imaging of Implanted Deep Brain Stimulators: Experience in a Large Series , 2007, Stereotactic and Functional Neurosurgery.

[14]  J. Volkmann,et al.  Deep brain stimulation for idiopathic or secondary movement disorders. , 2005, Acta neurochirurgica. Supplement.

[15]  W. Hamel,et al.  Documentation of electrode localization , 2002, Movement disorders : official journal of the Movement Disorder Society.

[16]  Wolfgang Kainz,et al.  MR heating tests of MR critical implants , 2007, Journal of magnetic resonance imaging : JMRI.

[17]  Aviva Abosch,et al.  Long-term Hardware-related Complications of Deep Brain Stimulation , 2002, Neurosurgery.

[18]  W. Ondo,et al.  The North American Survey of Placement and Adjustment Strategies for Deep Brain Stimulation , 2005, Stereotactic and Functional Neurosurgery.

[19]  Shuji Usui,et al.  Dependence of RF heating on SAR and implant position in a 1.5T MR system. , 2007, Magnetic resonance in medical sciences : MRMS : an official journal of Japan Society of Magnetic Resonance in Medicine.

[20]  S L Pullman,et al.  Operative techniques and morbidity with subthalamic nucleus deep brain stimulation in 100 consecutive patients with advanced Parkinson’s disease , 2005, Journal of Neurology, Neurosurgery & Psychiatry.

[21]  S. Hemm,et al.  Deep brain stimulation in movement disorders: stereotactic coregistration of two-dimensional electrical field modeling and magnetic resonance imaging. , 2005, Journal of neurosurgery.

[22]  Jean A. Tkach,et al.  Evaluation of specific absorption rate as a dosimeter of MRI‐related implant heating , 2004, Journal of magnetic resonance imaging : JMRI.

[23]  A. Benabid,et al.  Long term effects of bilateral subthalamic nucleus stimulation on cognitive function, mood, and behaviour in Parkinson’s disease , 2004, Journal of Neurology, Neurosurgery & Psychiatry.

[24]  J. Thiran,et al.  Localization of electrodes in the subthalamic nucleus on magnetic resonance imaging. , 2007, Journal of neurosurgery.

[25]  F. Duffner,et al.  Relevance of Image Fusion for Target Point Determination in Functional Neurosurgery , 2002, Acta Neurochirurgica.

[26]  Whole Grain Label Statements Guidance for Industry and FDA Staff , 2006 .

[27]  Abdelhamid Benazzouz,et al.  Imaging of subthalamic nucleus and ventralis intermedius of the thalamus , 2002, Movement disorders : official journal of the Movement Disorder Society.