Automated Condition-Based Suppression of the CPR Artifact in ECG Data to Make a Reliable Shock Decision for AEDs during CPR

Cardiopulmonary resuscitation (CPR) corrupts the morphology of the electrocardiogram (ECG) signal, resulting in an inaccurate automated external defibrillator (AED) rhythm analysis. Consequently, most current AEDs prohibit CPR during the rhythm analysis period, thereby decreasing the survival rate. To overcome this limitation, we designed a condition-based filtering algorithm that consists of three stop-band filters which are turned either ‘on’ or ‘off’ depending on the ECG’s spectral characteristics. Typically, removing the artifact’s higher frequency peaks in addition to the highest frequency peak eliminates most of the ECG’s morphological disturbance on the non-shockable rhythms. However, the shockable rhythms usually have dynamics in the frequency range of (3–6) Hz, which in certain cases coincide with CPR compression’s harmonic frequencies, hence, removing them may lead to destruction of the shockable signal’s dynamics. The proposed algorithm achieves CPR artifact removal without compromising the integrity of the shockable rhythm by considering three different spectral factors. The dataset from the PhysioNet archive was used to develop this condition-based approach. To quantify the performance of the approach on a separate dataset, three performance metrics were computed: the correlation coefficient, signal-to-noise ratio (SNR), and accuracy of Defibtech’s shock decision algorithm. This dataset, containing 14 s ECG segments of different types of rhythms from 458 subjects, belongs to Defibtech commercial AED’s validation set. The CPR artifact data from 52 different resuscitators were added to artifact-free ECG data to create 23,816 CPR-contaminated data segments. From this, 82% of the filtered shockable and 70% of the filtered non-shockable ECG data were highly correlated (>0.7) with the artifact-free ECG; this value was only 13 and 12% for CPR-contaminated shockable and non-shockable, respectively, without our filtering approach. The SNR improvement was 4.5 ± 2.5 dB, averaging over the entire dataset. Defibtech’s rhythm analysis algorithm was applied to the filtered data. We found a sensitivity improvement from 67.7 to 91.3% and 62.7 to 78% for VF and rapid VT, respectively, and specificity improved from 96.2 to 96.5% and 91.5 to 92.7% for normal sinus rhythm (NSR) and other non-shockables, respectively.

[1]  Mi He,et al.  Removal of Cardiopulmonary Resuscitation Artifacts with an Enhanced Adaptive Filtering Method: An Experimental Trial , 2014, BioMed research international.

[2]  Mohsen Ahmadi,et al.  Automated Method for Discrimination of Arrhythmias Using Time, Frequency, and Nonlinear Features of Electrocardiogram Signals , 2018, Sensors.

[3]  F. M. Nolle,et al.  Evaluation of a frequency-domain algorithm to detect ventricular fibrillation in the surface electrocardiogram , 1988, Proceedings. Computers in Cardiology 1988.

[4]  R. Merchant,et al.  Part 1: Executive Summary: 2020 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. , 2020, Circulation.

[5]  Jesus Ruiz,et al.  Rhythm Analysis during Cardiopulmonary Resuscitation: Past, Present, and Future , 2014, BioMed research international.

[6]  O. Kwon The changes in cardiopulmonary resuscitation guidelines: from 2000 to the present , 2019, Journal of exercise rehabilitation.

[7]  C. Li,et al.  A new method without reference channels used for ventricular fibrillation detection during cardiopulmonary resuscitation , 2016, Australasian Physical & Engineering Sciences in Medicine.

[8]  Tobias Werther,et al.  Reduction of CPR artifacts in the ventricular fibrillation ECG by coherent line removal , 2010, Biomedical engineering online.

[9]  Peng Gao,et al.  An Enhanced Adaptive Filtering Method for Suppressing Cardiopulmonary Resuscitation Artifact , 2017, IEEE Transactions on Biomedical Engineering.

[10]  Bihua Chen,et al.  A review of the performance of artifact filtering algorithms for cardiopulmonary resuscitation. , 2013, Journal of healthcare engineering.

[11]  M. Stapff [Artefacts]. , 2019, Bildgebung = Imaging.

[12]  Trygve Eftestøl,et al.  Rhythm Analysis during Cardiopulmonary Resuscitation Using Convolutional Neural Networks , 2020, Entropy.

[13]  P. Steen,et al.  Reducing CPR artefacts in ventricular fibrillation in vitro. , 2001, Resuscitation.

[14]  H U Strohmenger,et al.  Analysis of the ventricular fibrillation ECG signal amplitude and frequency parameters as predictors of countershock success in humans. , 1997, Chest.

[15]  Jesus Ruiz,et al.  Detection of ventricular fibrillation in the presence of cardiopulmonary resuscitation artefacts. , 2007, Resuscitation.

[16]  Jesus Ruiz,et al.  Cardiopulmonary resuscitation artefact suppression using a Kalman filter and the frequency of chest compressions as the reference signal. , 2010, Resuscitation.

[17]  K. Chon,et al.  Deep Neural Network Approach for Continuous ECG‐Based Automated External Defibrillator Shock Advisory System During Cardiopulmonary Resuscitation , 2021, Journal of the American Heart Association.

[18]  R. Koster,et al.  Automated external defibrillator and operator performance in out-of-hospital cardiac arrest. , 2017, Resuscitation.

[19]  Irena Jekova,et al.  Optimization of End-to-End Convolutional Neural Networks for Analysis of Out-of-Hospital Cardiac Arrest Rhythms during Cardiopulmonary Resuscitation , 2021, Sensors.

[20]  G. Ristagno,et al.  See through ECG technology during cardiopulmonary resuscitation to analyze rhythm and predict defibrillation outcome , 2016, Current opinion in critical care.

[21]  Jeffrey M. Hausdorff,et al.  Physionet: Components of a New Research Resource for Complex Physiologic Signals". Circu-lation Vol , 2000 .

[22]  U. Irusta,et al.  A Reliable Method for Rhythm Analysis during Cardiopulmonary Resuscitation , 2014, BioMed research international.

[23]  Huadong Zhu,et al.  The performance of a new shock advisory algorithm to reduce interruptions during CPR. , 2019, Resuscitation.

[24]  Jesus Ruiz,et al.  A method to remove CPR artefacts from human ECG using only the recorded ECG. , 2008, Resuscitation.

[25]  Bystander automated external defibrillator application in non-shockable out-of-hospital cardiac arrest. , 2019, Resuscitation.

[26]  Yongqin Li,et al.  An Algorithm Used for Ventricular Fibrillation Detection Without Interrupting Chest Compression , 2012, IEEE Transactions on Biomedical Engineering.

[27]  Andrew J. Thomas,et al.  Survival in out-of-hospital cardiac arrests with initial asystole or pulseless electrical activity and subsequent shockable rhythms. , 2013, Resuscitation.

[28]  Judy Powell,et al.  Perishock Pause: An Independent Predictor of Survival From Out-of-Hospital Shockable Cardiac Arrest , 2011, Circulation.

[29]  U. Irusta,et al.  An alternative to derive the instantaneous frequency of the chest compressions to suppress the CPR Artifact , 2010, 2010 Computing in Cardiology.

[30]  F. Kette,et al.  What is Ventricular Tachycardia for an Automated External Defibrillator , 2014 .

[31]  J P Ornato,et al.  Automatic external defibrillators for public access defibrillation: recommendations for specifying and reporting arrhythmia analysis algorithm performance, incorporating new waveforms, and enhancing safety. A statement for health professionals from the American Heart Association Task Force on Automa , 1997, Circulation.

[32]  Zhang Guang,et al.  Detection of Shockable Rhythm during Chest Compression based on Machine Learning , 2019, 2019 IEEE 8th Joint International Information Technology and Artificial Intelligence Conference (ITAIC).

[33]  Klaus Rheinberger,et al.  Removal of CPR Artifacts From the Ventricular Fibrillation ECG by Adaptive Regression on Lagged Reference Signals , 2008, IEEE Transactions on Biomedical Engineering.

[34]  Ali Bahrami Rad,et al.  ECG Rhythm Analysis During Manual Chest Compressions Using an Artefact Removal Filter and Random Forest Classifiers , 2018, 2018 Computing in Cardiology Conference (CinC).