A modular, extendible and field-tolerant multichannel vector magnetometer based on current sensor SQUIDs

We present the prototype module of our extendible and robust multichannel SQUID magnetometer system. A large multi-module arrangement can be implemented by using up to 7 modules. The system is intended for high-precision measurements of biomagnetism and spin precession. Further demanding applications are magnetorelaxometry and ultra-low-field nuclear magnetic resonance (ULF NMR), where pulsed magnetic fields of up to 100 mT are typically applied. The system is operated inside the Berlin magnetically shielded room (BMSR-2) and equipped with 18 magnetometers consisting of niobium (Nb) wire-wound pick-up coils. A total of 16 small pick-up coils with 17.1 mm diameter form a regular grid with individual channels arranged to ensure system sensitivity covers all three orthogonal spatial directions. Two large hexagonal pick-up coils with an equivalent diameter of 74.5 mm sensitive in z-direction surround the grid at two different heights and are suitable for the detection of deep sources. Each pick-up coil is connected to the input of a thin-film Nb SQUID current sensor via a detachable superconducting contact. The SQUIDs are equipped with integrated input current limiters. Feedback into the pick-up coils is employed to minimise crosstalk between channels. The current sensor chip package includes a superconducting shield of Nb. The field distortion of the prototype and a multi-module arrangement was analysed by numerical simulation. The measured noise of the small magnetometers was between 0.6 and 1.5 fT , and well below 1 fT for the large ones. Using a software gradiometer, we achieved a minimum noise level of 0.54 fT . We performed ULF NMR experiments, verifying the system's robustness against pulsed fields, and magnetoencephalographgy (MEG) on somatosensory evoked neuronal activity. The low noise performance of our 18-channel prototype enabled the detection of high-frequency components at around 1 kHz by MEG.

[1]  G. Curio,et al.  Ultra-low-noise EEG/MEG systems enable bimodal non-invasive detection of spike-like human somatosensory evoked responses at 1 kHz , 2015, Physiological measurement.

[2]  V. Foglietti,et al.  Multichannel Instrumentation for Biomagnetism , 1992 .

[3]  John Clarke,et al.  SQUID-detected magnetic resonance imaging in microtesla fields. , 2007, Annual review of biomedical engineering.

[4]  A. Walker Electroencephalography, Basic Principles, Clinical Applications and Related Fields , 1982 .

[5]  Y. Haruta,et al.  Contribution of primary somatosensory area 3b to somatic cognition: a neuromagnetic study , 2002, Neuroreport.

[6]  J. R. Claycomb,et al.  Superconducting magnetic shields for SQUID applications , 1999 .

[7]  R. Ilmoniemi,et al.  Sampling theory for neuromagnetic detector arrays , 1993, IEEE Transactions on Biomedical Engineering.

[8]  L. Trahms,et al.  Nuclear magnetic relaxation in water revisited. , 2011, The Journal of chemical physics.

[9]  Ki Woong Kim Ultra-Low Field Nuclear Magnetic Resonance , 2018 .

[10]  Michelle A. Espy,et al.  Ultra-Low Field Nuclear Magnetic Resonance: A New MRI Regime , 2014 .

[11]  D. Drung,et al.  Highly Sensitive and Easy-to-Use SQUID Sensors , 2007, IEEE Transactions on Applied Superconductivity.

[12]  Alex I. Braginski,et al.  Fundamentals and technology of SQUIDs and SQUID systems , 2004 .

[13]  M. Burghoff,et al.  Type-I superconductor pick-up coil in superconducting quantum interference device-based ultra-low field nuclear magnetic resonance , 2014 .

[14]  Martin Burghoff,et al.  An advanced phantom study assessing the feasibility of neuronal current imaging by ultra-low-field NMR. , 2013, Journal of magnetic resonance.

[15]  Jukka Nenonen,et al.  Thermal Noise in Biomagnetic Measurements , 1993 .

[16]  H.J.M. ter Brake,et al.  Elimination of flux-transformer crosstalk in multichannel SQUID magnetometers , 1986 .

[17]  Leif Grönberg,et al.  All-planar SQUIDs and pickup coils for combined MEG and MRI , 2011 .

[18]  Jari Penttilä,et al.  Hybrid ultra‐low‐field MRI and magnetoencephalography system based on a commercial whole‐head neuromagnetometer , 2013, Magnetic resonance in medicine.

[19]  A. Matlashov,et al.  SQUID-based systems for co-registration of ultra-low field nuclear magnetic resonance images and magnetoencephalography , 2012 .

[20]  A. N. Matlashov,et al.  SQUID-based instrumentation for ultralow-field MRI , 2007 .

[21]  T. Fedele High-frequency electroencephalography (hf-EEG): non-invasive detection of spike-related brain activity , 2014 .

[22]  Martin Burghoff,et al.  Magnetic resonance imaging at frequencies below 1 kHz. , 2013, Magnetic resonance imaging.