A Micropower Miniature Piezoelectric Actuator for Implantable Middle Ear Hearing Device

This paper describes the design and development of a small actuator using a miniature piezoelectric stack and a flextensional mechanical amplification structure for an implantable middle ear hearing device (IMEHD). A finite-element method was used in the actuator design. Actuator vibration displacement was measured using a laser vibrometer. Preliminary evaluation of the actuator for an IMEHD was conducted using a temporal bone model. Initial results from one temporal bone study indicated that the actuator was small enough to be implanted within the middle ear cavity, and sufficient stapes displacement can be generated for patients with mild to moderate hearing losses, especially at higher frequency range, by the actuator suspended onto the stapes. There was an insignificant mass-loading effect on normal sound transmission (<;3 dB) when the actuator was attached to the stapes and switched off. Improved vibration performance is predicted by more firm attachment. The actuator power consumption and its generated equivalent sound pressure level are also discussed. In conclusion, the actuator has advantages of small size, lightweight, and micropower consumption for potential use as IMHEDs.

[1]  George B. Wanna,et al.  Middle Ear Implantable Hearing Devices: An Overview , 2009, Trends in amplification.

[2]  K Gyo,et al.  Surgical rehabilitation of deafness with partially implantable hearing aid using piezoelectric ceramic bimorpli ossicular vibrator. , 1997, Auris, nasus, larynx.

[3]  Kuang-Hua Chang,et al.  Three-dimensional Modeling of Middle Ear Biomechanics and Its Applications , 2002, Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology.

[4]  D. Backous,et al.  Implantable middle ear hearing devices: current state of technology and market challenges , 2006, Current opinion in otolaryngology & head and neck surgery.

[5]  Christopher Niezrecki,et al.  Piezoelectric actuation: State of the art , 2001 .

[6]  W H Ko,et al.  Engineering principles of mechanical stimulation of the middle ear. , 1995, Otolaryngologic clinics of North America.

[7]  E. Abel,et al.  Preliminary Assessment of Remote Photoelectric Excitation of an Actuator for a Hearing Implant , 2005, 2005 IEEE Engineering in Medicine and Biology 27th Annual Conference.

[8]  C. Toumazou,et al.  A 126-/spl mu/W cochlear chip for a totally implantable system , 2005, IEEE Journal of Solid-State Circuits.

[9]  Jin-Ho Cho,et al.  Evaluation of an implantable piezoelectric floating mass transducer for sensorineural hearing loss , 2009 .

[10]  S. Merchant,et al.  Testing a Method for Quantifying the Output of Implantable Middle Ear Hearing Devices , 2007, Audiology and Neurotology.

[11]  H P Zenner,et al.  [An implantable piezoelectric hearing aid transducer for inner ear deafness. II: Clinical implant]. , 1997, HNO.

[12]  Hans Leysieffer,et al.  Total Implantation of the Active Hearing Implant Tica for Middle Ear Disease: A Temporal Bone Study , 2001, The Annals of otology, rhinology, and laryngology.

[13]  H. Pau,et al.  Vibroplasty involving direct coupling of the floating mass transducer to the oval window niche , 2010, The Journal of Laryngology & Otology.

[14]  Mark W Wood,et al.  Middle Ear Electromagnetic Semi-Implantable Hearing Device: Results of the Phase II SOUNDTEC Direct System Clinical Trial , 2002, Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology.

[15]  A J Maniglia,et al.  The history and development of the implantable hearing aid. , 1995, Otolaryngologic clinics of North America.

[16]  N Yanagihara,et al.  Long-term results using a piezoelectric semi-implantable middle ear hearing device: the Rion Device E-type. , 2001, Otolaryngologic clinics of North America.

[17]  Xuelin Wang,et al.  A totally implantable hearing system – Design and function characterization in 3D computational model and temporal bones , 2010, Hearing Research.

[18]  Y. Liu,et al.  Assessment of multi-layer piezoelectric actuator technology for middle-ear implants , 2002 .

[19]  N. Stasche,et al.  Middle ear transmission disorders--tympanic membrane vibration analysis by laser-Doppler-vibrometry. , 1994, Acta oto-laryngologica.

[20]  H P Zenner,et al.  Total implantation of the Implex TICA hearing amplifier implant for high frequency sensorineural hearing loss: the Tübingen University experience. , 2001, Otolaryngologic clinics of North America.

[21]  M. Vlaming,et al.  Studies on the mechanics of the normal human middle ear. , 1986, Clinical otolaryngology and allied sciences.

[22]  Don Nakmali,et al.  Magnetic Resonance Imaging Compatibility and Safety of the SOUNDTEC Direct System , 2006, The Laryngoscope.

[23]  Ilhan A. Aksay,et al.  Scaling Analysis for the Axial Displacement and Pressure of Flextensional Transducers , 2005 .

[24]  Rong Z Gan,et al.  Human Middle Ear Transfer Function Measured by Double Laser Interferometry System , 2004, Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology.

[25]  J. W. Baumann,et al.  Ein implantierbarer piezoelektrischer Hörgerätewandler für Innenohrschwerhörige Teil II: Klinisches Implantat , 1997, HNO.

[26]  Eric Abel,et al.  Methodology for laser vibrometer studies of stapes footplate displacement , 2000 .

[27]  E W Abel,et al.  Effects of draining cochlear fluids on stapes displacement in human middle-ear models. , 2001, The Journal of the Acoustical Society of America.

[28]  Geoffrey R. Ball Implantable electromagnetic hearing transducer , 1997 .