Preliminary evaluation of the tomographic performance of the mediSPECT small animal imaging system

Abstract We report on the tests of a prototype (MediSPECT) system developed at University & INFN Napoli, for Single Photon Emission Computed Tomography (SPECT) imaging on small animals with a small Field of View (FoV) and high spatial resolution. MediSPECT is a SPECT imaging system based on a 1-mm-thick CdTe pixel detector, bump-bonded to the Medipix2 CMOS readout circuit operating in single-photon counting. The CdTe detector has 256×256 square array of pixels arranged with a 55 μm pitch, for a sensitive area of 14×14 mm 2 . In its present version, this system implements a single detector head, mounted on a rotating gantry. For preliminary testing and calibration of the acquisition equipment and image reconstruction algorithms, 90 projections of a γ-ray point source ( 109 Cd) through a single pinhole (diameter 0.4 mm; radius of rotation about 2.5 cm; focal length about 4.5 cm) were acquired for 20 min each in a step-and-shoot mode. Capillaries, 800 μm in diameter, were arranged in a Y-shape to form a more complex phantom ( 125 I, 1 mm pinhole diameter, 45 projections, each acquired for 25 min). Images were reconstructed with a custom algorithm implementing standard OS-EM with center of rotation correction and spatial resolution of 0.2 mm over a FoV of 2 mm was obtained.

[1]  H. B. Barber,et al.  Compact CT/SPECT small-animal imaging system , 2002, IEEE Transactions on Nuclear Science.

[2]  H. Malcolm Hudson,et al.  Accelerated image reconstruction using ordered subsets of projection data , 1994, IEEE Trans. Medical Imaging.

[3]  F. Beekman,et al.  Design and simulation of a high-resolution stationary SPECT system for small animals. , 2004, Physics in medicine and biology.

[4]  M. C. Montesi,et al.  Experimental study on in vivo optical and radionuclide imaging in small animals , 2003, IEEE Transactions on Nuclear Science.

[5]  R. Jaszczak,et al.  Pinhole collimation for ultra-high-resolution, small-field-of-view SPECT. , 1994, Physics in medicine and biology.

[6]  Giovanni Mettivier,et al.  Preliminary test of Medisoft 4: control software for the Medipix2 read-out chip , 2002 .

[7]  Eric C. Frey,et al.  Pinhole SPECT of mice using the LumaGEM gamma camera , 2001 .

[8]  Simon R. Cherry,et al.  Towards in vivo nuclear microscopy: iodine-125 imaging in mice using micro-pinholes , 2002, European Journal of Nuclear Medicine and Molecular Imaging.

[9]  Donald W. Wilson,et al.  SemiSPECT: a small-animal single-photon emission computed tomography (SPECT) imager based on eight cadmium zinc telluride (CZT) detector arrays. , 2006, Medical physics.

[10]  Horst Halling,et al.  Compact high resolution detector for small animal SPECT , 1999, 1999 IEEE Nuclear Science Symposium. Conference Record. 1999 Nuclear Science Symposium and Medical Imaging Conference (Cat. No.99CH37019).

[11]  Alessandra Boschi,et al.  YAP-(S)PET small animal scanner: Quantitative results , 2003 .