Characterizing WR-8 waveguide-to-CPW probes using two methods implemented within the NIST Uncertainty Framework

We individually characterize and provide uncertainties for a pair of WR-8 rectangular waveguide-to-coplanar waveguide (CPW) probes over a frequency range of 90 to 140 GHz utilizing two methods implemented within the NIST Microwave Uncertainty Framework. The first method consists of a two-tier approach, where a vector network analyzer (VNA) calibration is first performed in rectangular waveguide, and then a second-tier calibration is performed on-wafer. The second method determines the scattering parameters of the probes from two one-tier calibrations. We show that the two methods yield nearly equivalent results and uncertainties.

[1]  Dylan F. Williams,et al.  > Replace This Line with Your Paper Identification Number (double-click Here to Edit) < , 2022 .

[2]  Howard C. Reader,et al.  Wideband Frequency-Domain Characterization of High-Impedance Probes , 2001, 58th ARFTG Conference Digest.

[3]  J. Martens Common adapter/fixture extraction techniques: sensitivities to calibration anomalies , 2009, 2009 74th ARFTG Microwave Measurement Conference.

[4]  P. D. Hale,et al.  Traceability of high-speed electrical waveforms at NIST, NPL, and PTB , 2012, 2012 Conference on Precision electromagnetic Measurements.

[5]  R. Collin Foundations for microwave engineering , 1966 .

[6]  U. Arz,et al.  Coplanar microwave probe characterization: Caveats and pitfalls , 2006, 2006 67th ARFTG Conference.

[7]  Dylan F. Williams,et al.  Covariance-Based Vector-Network-Analyzer Uncertainty Analysis for Time- and Frequency-Domain Measurements , 2010, IEEE Transactions on Microwave Theory and Techniques.

[8]  Jack C. M. Wang,et al.  An optimal vector-network-analyzer calibration algorithm , 2003 .

[9]  Paul Penfield,et al.  De-Embedding and Unterminating , 1974 .

[10]  J. Randa,et al.  Comparison of adapter characterization methods , 1999 .

[11]  T. Tischler,et al.  Modeling dispersion and radiation characteristics of conductor-backed CPW with finite ground width , 2000, IMS 2000.

[12]  P. D. Hale,et al.  Establishing traceability of an electronic calibration unit using the NIST Microwave Uncertainty Framework , 2012, 79th ARFTG Microwave Measurement Conference.