Nanometer scale marker for fluorescent microscopy

To establish a calibration method of optical performance in fluorescence microscopy, we fabricated a fluorescent nanometer-scale marker by combining a dry dye method for polymer film and fine lithography. The marker has a 50 nm line-and-space fluorescent pattern, finer than the optical diffraction limit. A spin-coated poly(methyl methacrylate) thin film on a silicon wafer was densely doped with Rhodamine 6G using a simple vacuum process, named the vapor-transportation method, and then the pattern was formed on the film using electron-beam lithography. The figure accuracy of the fabricated marker was calibrated by electron microscopes. Using this marker, one can quantitatively evaluate the optical properties; i.e., the contrast-transfer function, the point-spread function, magnification, and so on. To show practical use of the marker, we demonstrated the evaluation of a fluorescent microscope system.

[1]  C. Sheppard,et al.  Theory and practice of scanning optical microscopy , 1984 .

[2]  Tetsuo Moriya,et al.  A Novel Formation Method of Thin Polymer Film with Densely Dispersed Organic Dye by using Vacuum Technique , 2000 .

[3]  Nobutaka Tanigaki,et al.  Doping of functional materials into poly (p-phenylene vinylene) by the vapor transportation method , 2004 .

[4]  Reusable biological film: a new application of blue membrane , 2004 .

[5]  Takashi Hiraga,et al.  Selective Doping of Photochromic Dye into Nanostructures of Diblock Copolymer Films by Vaporization in a Vacuum , 2004 .

[6]  David J. Smith High Resolution Electron Microscopy - Principles and Practice , 2003, Microscopy and Microanalysis.

[7]  M G Somekh,et al.  Quantitative optical microscope with enhanced resolution using a pixelated liquid crystal spatial light modulator , 2004, Journal of microscopy.

[8]  V. Nampoori,et al.  Loss characterization in rhodamine 6G doped polymer film waveguide by side illumination fluorescence , 2004 .

[9]  Måns Ehrenberg,et al.  Rotational brownian motion and fluorescence intensify fluctuations , 1974 .

[10]  D. Nyyssonen,et al.  Submicrometer Linewidth Metrology In the Optical Microscope , 1987, Journal of Research of the National Bureau of Standards.

[11]  T. David Binnie,et al.  EXPERIMENTAL APPROACH FOR MEASURING RESOLUTION OF COMPLEMENTARY METAL OXIDE SEMICONDUCTOR IMAGING SYSTEMS , 1998 .

[12]  Neelam Gupta,et al.  Telecentric confocal optics for aberration correction of acousto-optic tunable filters. , 2004, Applied optics.

[13]  E. Abbe Beiträge zur Theorie des Mikroskops und der mikroskopischen Wahrnehmung , 1873 .

[14]  M. Higuchi,et al.  Stepwise radial complexation of imine groups in phenylazomethine dendrimers , 2002, Nature.

[15]  Stefan W. Hell,et al.  Laser-diode-stimulated emission depletion microscopy , 2003 .

[16]  Takashige Omatsu,et al.  Two-point-separation in super-resolution fluorescence microscope based on up-conversion fluorescence depletion technique. , 2003, Optics express.

[17]  Y. Iketaki,et al.  Two-color far-field super-resolution microscope using a doughnut beam , 2003 .

[18]  E. Stelzer Light microscopy: Beyond the diffraction limit? , 2002, Nature.

[19]  Yasushi Miyashita,et al.  Dendritic spine geometry is critical for AMPA receptor expression in hippocampal CA1 pyramidal neurons , 2001, Nature Neuroscience.

[20]  P. Davidovits,et al.  Scanning Laser Microscope , 1969, Nature.