Evaluating the Effect of Flow and Interface Type on Pressures Delivered With Bubble CPAP in a Simulated Model

BACKGROUND: Bubble CPAP, used for spontaneously breathing infants to avoid intubation or postextubation support, can be delivered with different interface types. This study compared the effect that interfaces had on CPAP delivery. We hypothesized that there would be no difference between set and measured levels between interface types. METHODS: A validated preterm infant nasal airway model was attached to the ASL 5000 breathing simulator. The simulator was programmed to deliver active breathing of a surfactant-deficient premature infant with breathing frequency at 70 breaths/min inspiratory time of 0.30 s, resistance of 150 cm H2O/L/s, compliance of 0.5 mL/cm H2O, tidal volume of 5 mL, and esophageal pressure of −10 cm H2O. Nasal CPAP prongs, size 4030, newborn and infant RAM cannulas were connected to a nasal airway model and a bubble CPAP system. CPAP levels were set at 4, 5, 6, 7, 8, and 9 cm H2O with flows of 6, 8, and 10 L/min each. Measurements were recorded after 1 min of stabilization. The analysis was performed using SAS 9.4. The Kolmogorov-Smirnov test assessed normality of the data. The Friedman test was used to compare non-normally distributed repeated measures. The Wilcoxon signed-rank test was used to conduct post hoc analysis. All tests were 2-sided, and P values of <.05 were considered as indicating significant differences unless otherwise indicated. RESULTS: At lower set CPAP levels, 4–6 cm H2O, measured CPAP dropped precipitously with the nasal prongs with the highest flow setting. At higher CPAP levels, 7–9 cm H2O measured CPAP concomitantly increased as the flow setting increased. Statistically significant differences in set and measured CPAP occurred for all devices across all CPAP levels, with the measured CPAP less than set for all conditions, P < .001. CONCLUSIONS: Set flow had a profound effect on measured CPAP. The concomitant drop in measured pressure with high and low flows could be attributed to increased resistance to spontaneous breathing or insufficient flow to meet inspiratory demand. Clinicians should be aware of the effect that the interface and flow have on CPAP delivery.

[1]  R. Chatburn,et al.  Evaluation of a Nasal Cannula in Noninvasive Ventilation Using a Lung Simulator , 2015, Respiratory Care.

[2]  C. P. Richardson,et al.  Volume Oscillations Delivered to a Lung Model Using 4 Different Bubble CPAP Systems , 2015, Respiratory Care.

[3]  L. Papile,et al.  Respiratory Support in Preterm Infants at Birth , 2014, Pediatrics.

[4]  Tiffany M. Youngquist,et al.  Effects of Condensate in the Exhalation Limb of Neonatal Circuits on Airway Pressure During Bubble CPAP , 2013, Respiratory Care.

[5]  W. T. Bass,et al.  An integrative review of skin breakdown in the preterm infant associated with nasal continuous positive airway pressure. , 2013, Journal of obstetric, gynecologic, and neonatal nursing : JOGNN.

[6]  R. Chatburn,et al.  High Flow Through a Nasal Cannula and CPAP Effect in a Simulated Infant Model , 2011, Respiratory Care.

[7]  A. de Klerk,et al.  Randomized Trial Comparing 3 Approaches to the Initial Respiratory Management of Preterm Neonates , 2011, Pediatrics.

[8]  Abhik Das,et al.  Early CPAP versus surfactant in extremely preterm infants. , 2010, The New England journal of medicine.

[9]  R. Diblasi,et al.  Nasal continuous positive airway pressure (CPAP) for the respiratory care of the newborn infant. , 2009, Respiratory care.

[10]  K. Barrington Management of respiratory failure in the preterm infant. , 2008, Minerva pediatrica.

[11]  R. Habib,et al.  Work of Breathing During Nasal Continuous Positive Airway Pressure in Preterm Infants: A Comparison of Bubble vs Variable-Flow Devices , 2005, Journal of Perinatology.

[12]  A. Kopelman Airway Obstruction in Two Extremely Low Birthweight Infants Treated with Oxygen Cannulas , 2003, Journal of Perinatology.

[13]  D. Holbert,et al.  Use of Oxygen Cannulas in Extremely Low Birthweight Infants is Associated with Mucosal Trauma and Bleeding, and Possibly with Coagulase-negative Staphylococcal Sepsis , 2003, Journal of Perinatology.

[14]  Kyong-Soon Lee,et al.  A Comparison of Underwater Bubble Continuous Positive Airway Pressure with Ventilator-Derived Continuous Positive Airway Pressure in Premature Neonates Ready for Extubation , 1998, Neonatology.

[15]  C. Morley,et al.  Computerised determination of spontaneous inspiratory and expiratory times in premature neonates during intermittent positive pressure ventilation. II: Results from 20 babies. , 1994, Archives of disease in childhood. Fetal and neonatal edition.

[16]  A. Fanaroff,et al.  A device for administration of continuous positive airway pressure by the nasal route. , 1973, Pediatrics.