An algorithm for the ambulatory monitoring of fetal heart rate

A real-time algorithm has been developed for the simultaneous measurement of the fetal and maternal heart rates from the abdominal signal, which was sampled with 13-bit resolution at 500 Hz. The algorithm is based on digital filtering, adaptive thresholding, differencing of local maxima and minima and statistical properties in the time domain. The technique has been developed through a combination and modification of earlier techniques. The algorithm is found to be effective in resolving the fetal heart rate (FHR) when the fetal R waves and that of the maternal are overlapped. The algorithm has been tested on recorded ambulatory data from 36 to 41 gestational weeks of pregnancy, resulting in an average accuracy of 82% (min, 68%) for the fetal heart rate detection. A microcontroller based system has been developed and the proposed algorithm implemented in real-time. The above system is capable of continuous ambulatory FHR monitoring.

[1]  R. L. Longini,et al.  Abdominal-Lead Fetal Electrocardiographic R-Wave Enhancement for Heart Rate Determination , 1980, IEEE Transactions on Biomedical Engineering.

[2]  M.A. Mohd Ali,et al.  Development of a fetal and maternal heart rate recorder for ambulatory use , 1993, Proceedings of Computers in Cardiology Conference.

[3]  A. Kam,et al.  Maternal ECG elimination and foetal ECG detection-comparison of several algorithms , 1998, Proceedings of the 20th Annual International Conference of the IEEE Engineering in Medicine and Biology Society. Vol.20 Biomedical Engineering Towards the Year 2000 and Beyond (Cat. No.98CH36286).

[4]  S.L. Horner,et al.  A non-invasive and invasive fetal electrocardiogram real-time monitoring system , 1994, Computers in Cardiology 1994.

[5]  Willis J. Tompkins,et al.  A Real-Time QRS Detection Algorithm , 1985, IEEE Transactions on Biomedical Engineering.

[6]  P. Steer,et al.  The use of the fetal electrocardiogram in labour , 1994, British journal of obstetrics and gynaecology.

[7]  W. Peasgood,et al.  Antepartum fetal electrocardiogram extraction and analysis , 1993, Proceedings of Computers in Cardiology Conference.

[8]  J. Patrick,et al.  The nonstress test: how long is enough? , 1981, American journal of obstetrics and gynecology.

[9]  M J LAMKEE,et al.  Fetal electrocardiography: A study of the normal recording , 1962 .

[10]  Paul Benjamin Crilly,et al.  An effective algorithm for enhancing a non-invasive fetal electrocardiogram , 1996 .

[11]  J. .. Bemmel,et al.  Detection Procedure to Represent the Foetal Heart Rate and Electrocardiogram , 1966 .

[12]  Shimon Abboud,et al.  Real time fetal ECG monitoring based on a personal computer using correlation technique , 1989, [1989] Proceedings. Computers in Cardiology.

[13]  J H van Bemmel,et al.  Detection procedure to represent the foetal heart rate and electrocardiogram. , 1966, IEEE transactions on bio-medical engineering.