Accuracy in tangential breast treatment set-up: a portal imaging study.

To test the accuracy and reproducibility of the tangential breast treatment set-up used in The Netherlands Cancer Institute, a portal imaging study was performed in 12 patients treated for early stage breast cancer. With an on-line electronic portal imaging device (EPID) images were obtained of each patient in several fractions and compared with simulator films and with each other. In five patients, multiple images (on the average 7) per fraction were obtained to evaluate set-up variations due to respiratory movement. The central lung distance (CLD) and other set-up parameters varied within one fraction about 1 mm (1 SD). The average variation of these parameters between various fractions was about 2 mm (1 SD). The differences between simulator and treatment set-up over all patients and all fractions was on the average 2-3 mm for the central beam edge to skin distance and the central lung distance. It can be concluded that the tangential breast treatment set-up is very stable and reproducible and that respiration does not have a significant influence on treatment volume. The EPID appears to be an adequate tool for studies of treatment set-up accuracy like this.

[1]  H. Hamers,et al.  Entrance and exit TL-dosimetry in the conservative treatment of breast cancer: a pilot study for the EORTC-Radiotherapy Cooperative Group. , 1991, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[2]  M van Herk,et al.  Radiation field edge detection in portal images. , 1991, Physics in medicine and biology.

[3]  R Mohan,et al.  The role of uncertainty analysis in treatment planning. , 1991, International journal of radiation oncology, biology, physics.

[4]  S Webb,et al.  Clinical dosimetry for radiotherapy to the breast based on imaging with the prototype Royal Marsden Hospital CT simulator. , 1987, Physics in medicine and biology.

[5]  J van de Geijn,et al.  A technique for field matching in primary breast irradiation. , 1983, International journal of radiation oncology, biology, physics.

[6]  J. W. Epstein,et al.  On-line radiotherapy imaging with an array of fiber-optic image reducers. , 1990, International journal of radiation oncology, biology, physics.

[7]  O. Nascimben,et al.  Three-field isocentric technique for breast irradiation using individualized shielding blocks. , 1988, International journal of radiation oncology, biology, physics.

[8]  J. Lebesque,et al.  Field matching in breast irradiation: An exact solution to a geometrical problem. , 1986, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[9]  P. Wootton,et al.  Dose Specification for Reporting External Beam Therapy with Photons and Electrons , 1979 .

[10]  M van Herk,et al.  Fast evaluation of patient set-up during radiotherapy by aligning features in portal and simulator images. , 1991, Physics in medicine and biology.

[11]  R. Siddon,et al.  Three-dimensional photon dose distributions with and without lung corrections for tangential breast intact treatments. , 1989, International journal of radiation oncology, biology, physics.

[12]  B J Mijnheer,et al.  In vivo dosimetry during tangential breast treatment. , 1991, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[13]  J Bijhold,et al.  A method for the measurement of field placement errors in digital portal images. , 1990, Physics in medicine and biology.

[14]  R L Siddon,et al.  Solution to treatment planning problems using coordinate transformations. , 1981, Medical physics.

[15]  M. Sontag,et al.  Three-dimensional photon treatment planning of the intact breast. , 1991, International journal of radiation oncology, biology, physics.

[16]  D. McShan,et al.  The influence of lung density corrections on treatment planning for primary breast cancer. , 1988, International journal of radiation oncology, biology, physics.

[17]  K. Kiel,et al.  Technique for breast irradiation using custom blocks conforming to the chest wall contour. , 1990, International Journal of Radiation Oncology, Biology, Physics.

[18]  M van Herk,et al.  First clinical experience with a newly developed electronic portal imaging device. , 1990, International journal of radiation oncology, biology, physics.

[19]  U F Rosenow,et al.  A technique for treating local breast cancer using a single set-up point and asymmetric collimation. , 1990, International journal of radiation oncology, biology, physics.

[20]  G K Svensson,et al.  Three-field technique for breast treatment using a rotatable half-beam block. , 1981, International journal of radiation oncology, biology, physics.

[21]  B Kihlén,et al.  Reproducibility of field alignment in radiation therapy. A large-scale clinical experience. , 1989, Acta oncologica.

[22]  E B Podgorsak,et al.  A simple isocentric technique for irradiation of the breast, chest wall and peripheral lymphatics. , 1984, The British journal of radiology.

[23]  R. Siddon,et al.  Can simulation measurements be used to predict the irradiated lung volume in the tangential fields in patients treated for breast cancer? , 1990, International journal of radiation oncology, biology, physics.

[24]  A G Visser,et al.  Performance of a prototype fluoroscopic radiotherapy imaging system. , 1990, International journal of radiation oncology, biology, physics.

[25]  G E Hanks,et al.  A nondivergent three field matching technique for breast irradiation. , 1990, International journal of radiation oncology, biology, physics.

[26]  G K Svensson,et al.  Three-field technique for breast irradiation using tangential field corner blocks. , 1983, International journal of radiation oncology, biology, physics.

[27]  M van Herk,et al.  A matrix ionisation chamber imaging device for on-line patient setup verification during radiotherapy. , 1988, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[28]  T Knöös,et al.  Comparison of measured and calculated absorbed doses from tangential irradiation of the breast. , 1986, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.