Static and dynamic error of a biplanar videoradiography system using marker-based and markerless tracking techniques.

The use of biplanar videoradiography technology has become increasingly popular for evaluating joint function in vivo. Two fundamentally different methods are currently employed to reconstruct 3D bone motions captured using this technology. Marker-based tracking requires at least three radio-opaque markers to be implanted in the bone of interest. Markerless tracking makes use of algorithms designed to match 3D bone shapes to biplanar videoradiography data. In order to reliably quantify in vivo bone motion, the systematic error of these tracking techniques should be evaluated. Herein, we present new markerless tracking software that makes use of modern GPU technology, describe a versatile method for quantifying the systematic error of a biplanar videoradiography motion capture system using independent gold standard instrumentation, and evaluate the systematic error of the W.M. Keck XROMM Facility's biplanar videoradiography system using both marker-based and markerless tracking algorithms under static and dynamic motion conditions. A polycarbonate flag embedded with 12 radio-opaque markers was used to evaluate the systematic error of the marker-based tracking algorithm. Three human cadaveric bones (distal femur, distal radius, and distal ulna) were used to evaluate the systematic error of the markerless tracking algorithm. The systematic error was evaluated by comparing motions to independent gold standard instrumentation. Static motions were compared to high accuracy linear and rotary stages while dynamic motions were compared to a high accuracy angular displacement transducer. Marker-based tracking was shown to effectively track motion to within 0.1 mm and 0.1 deg under static and dynamic conditions. Furthermore, the presented results indicate that markerless tracking can be used to effectively track rapid bone motions to within 0.15 deg for the distal aspects of the femur, radius, and ulna. Both marker-based and markerless tracking techniques were in excellent agreement with the gold standard instrumentation for both static and dynamic testing protocols. Future research will employ these techniques to quantify in vivo joint motion for high-speed upper and lower extremity impacts such as jumping, landing, and hammering.

[1]  T. Hewett,et al.  Biomechanical Measures of Neuromuscular Control and Valgus Loading of the Knee Predict Anterior Cruciate Ligament Injury Risk in Female Athletes: A Prospective Study , 2005, The American journal of sports medicine.

[2]  Stephanie K. Kline,et al.  Measuring dynamic in-vivo glenohumeral joint kinematics: technique and preliminary results. , 2008, Journal of biomechanics.

[3]  Michael J Bey,et al.  In vivo measurement of subacromial space width during shoulder elevation: technique and preliminary results in patients following unilateral rotator cuff repair. , 2007, Clinical biomechanics.

[4]  Arie E. Kaufman Volume visualization , 1996, CSUR.

[5]  Scott Tashman,et al.  A technique to measure three-dimensional in vivo rotation of fused and adjacent lumbar vertebrae. , 2008, The spine journal : official journal of the North American Spine Society.

[6]  T. Hewett,et al.  A comparison of dynamic coronal plane excursion between matched male and female athletes when performing single leg landings. , 2006, Clinical biomechanics.

[7]  J Erik Giphart,et al.  Knee kinematic profiles during drop landings: a biplane fluoroscopy study. , 2011, Medicine and science in sports and exercise.

[8]  John A. Nelder,et al.  A Simplex Method for Function Minimization , 1965, Comput. J..

[9]  P. Slusallek,et al.  High-speed volume ray casting with CUDA , 2008, 2008 IEEE Symposium on Interactive Ray Tracing.

[10]  Scott Tashman,et al.  Validation of three-dimensional model-based tibio-femoral tracking during running. , 2009, Medical engineering & physics.

[11]  Mindy F Levin,et al.  Differences in multi-joint kinematic patterns of repetitive hammering in healthy, fatigued and shoulder-injured individuals. , 2005, Clinical biomechanics.

[12]  David B Baier,et al.  Scientific rotoscoping: a morphology-based method of 3-D motion analysis and visualization. , 2010, Journal of experimental zoology. Part A, Ecological genetics and physiology.

[13]  Scott Tashman,et al.  Model-based tracking of the hip: implications for novel analyses of hip pathology. , 2011, The Journal of arthroplasty.

[14]  Scott Tashman,et al.  Kinematics of the ACL‐deficient canine knee during gait: Serial changes over two years , 2004, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[15]  Scott W Wolfe,et al.  Carpal and forearm kinematics during a simulated hammering task. , 2010, The Journal of hand surgery.

[16]  Frank Seehaus,et al.  Accuracy of Model-based RSA Contour Reduction in a Typical Clinical Application , 2008, Clinical orthopaedics and related research.

[17]  Jessica M. Deneweth,et al.  Tibiofemoral Joint Kinematics of the Anterior Cruciate Ligament-Reconstructed Knee During a Single-Legged Hop Landing , 2010, The American journal of sports medicine.

[18]  Tyson L Hedrick,et al.  Software techniques for two- and three-dimensional kinematic measurements of biological and biomimetic systems , 2008, Bioinspiration & biomimetics.

[19]  Mubarak Shah,et al.  MinGPU: a minimum GPU library for computer vision , 2008, Journal of Real-Time Image Processing.

[20]  D. B. Baier,et al.  X-ray reconstruction of moving morphology (XROMM): precision, accuracy and applications in comparative biomechanics research. , 2010, Journal of experimental zoology. Part A, Ecological genetics and physiology.

[21]  W E Garrett,et al.  Measurement of in vivo anterior cruciate ligament strain during dynamic jump landing. , 2011, Journal of biomechanics.

[22]  John C. Russ,et al.  The Image Processing Handbook , 2016, Microscopy and Microanalysis.

[23]  Scott Tashman,et al.  Journal of Orthopaedic Surgery and Research Accuracy of Biplane X-ray Imaging Combined with Model-based Tracking for Measuring In-vivo Patellofemoral Joint Motion , 2022 .

[24]  Scott Tashman,et al.  Abnormal Rotational Knee Motion during Running after Anterior Cruciate Ligament Reconstruction , 2004, The American journal of sports medicine.

[25]  S. Tashman,et al.  Dynamic Function of the ACL-reconstructed Knee during Running , 2007, Clinical orthopaedics and related research.

[26]  Scott Tashman,et al.  In-vivo measurement of dynamic joint motion using high speed biplane radiography and CT: application to canine ACL deficiency. , 2003, Journal of biomechanical engineering.

[27]  Frank Seehaus,et al.  Experimental analysis of Model-Based Roentgen Stereophotogrammetric Analysis (MBRSA) on four typical prosthesis components. , 2009, Journal of biomechanical engineering.

[28]  Kevin R Ford,et al.  Gender differences in the kinematics of unanticipated cutting in young athletes. , 2005, Medicine and science in sports and exercise.

[29]  T. Gill,et al.  In Vivo Elongation of the Anterior Cruciate Ligament and Posterior Cruciate Ligament during Knee Flexion , 2004, The American journal of sports medicine.

[30]  Milan Sonka,et al.  Image Processing, Analysis and Machine Vision , 1993, Springer US.

[31]  L. Chen,et al.  An investigation on the accuracy of three-dimensional space reconstruction using the direct linear transformation technique. , 1994, Journal of biomechanics.

[32]  T. Gill,et al.  In vivo kinematics of the ACL during weight‐bearing knee flexion , 2005, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[33]  Scott Tashman,et al.  Comments on "validation of a non-invasive fluoroscopic imaging technique for the measurement of dynamic knee joint motion". , 2008, Journal of biomechanics.

[34]  Scott Tashman,et al.  Validation of a new model-based tracking technique for measuring three-dimensional, in vivo glenohumeral joint kinematics. , 2006, Journal of biomechanical engineering.

[35]  A. Weiss,et al.  In vivo radiocarpal kinematics and the dart thrower's motion. , 2005, The Journal of bone and joint surgery. American volume.

[36]  S. Woo,et al.  Relationship of anterior knee laxity to knee translations during drop landings: a bi-plane fluoroscopy study , 2011, Knee Surgery, Sports Traumatology, Arthroscopy.

[37]  Scott Tashman,et al.  In vivo measurement of 3-D skeletal kinematics from sequences of biplane radiographs: Application to knee kinematics , 2001, IEEE Transactions on Medical Imaging.

[38]  Joseph J Crisco,et al.  Automatic determination of anatomical coordinate systems for three-dimensional bone models of the isolated human knee. , 2010, Journal of biomechanics.