Remote monitoring of left ventricular assist device parameters after HeartAssist-5 implantation.

Although several left ventricular assist devices (LVADs) have been used widely, remote monitoring of LVAD parameters has been available only recently. We present our remote monitoring experience with an axial-flow LVAD (HeartAssist-5, MicroMed Cardiovascular, Inc., Houston, TX, USA). Five consecutive patients who were implanted a HeartAssist-5 LVAD because of end-stage heart failure due to ischemic (n=4) or idiopathic (n=1) cardiomyopathy, and discharged from hospital between December 2011 and January 2013 were analyzed. The data (pump speed, pump flow, power consumption) obtained from clinical visits and remote monitoring were studied. During a median follow-up of 253 (range: 80-394) days, fine tuning of LVADs was performed at clinical visits. All patients are doing well and are in New York Heart Association Class-I/II. A total of 39 alarms were received from three patients. One patient was hospitalized for suspected thrombosis and was subjected to physical examinations as well as laboratory and echocardiographic evaluations; however, no evidence of thrombus washout or pump thrombus was found. The patient was treated conservatively. Remaining alarms were due to insufficient water intake and were resolved by increased water consumption at night and summer times, and fine tuning of pump speed. No alarms were received from the remaining two patients. We believe that remote monitoring is a useful technology for early detection and treatment of serious problems occurring out of hospital thereby improving patient care. Future developments may ease troubleshooting, provide more data from the patient and the pump, and eventually increase physician and patient satisfaction. Despite all potential clinical benefits, remote monitoring should be taken as a supplement to rather than a substitute for routine clinical visits for patient follow-up.

[1]  Seong Wook Choi,et al.  Personal digital assistant-based, internet-enabled remote communication system for a wearable pneumatic biventricular assist device. , 2007, Artificial organs.

[2]  G P Noon,et al.  Clinical experience with the MicroMed DeBakey ventricular assist device. , 2001, The Annals of thoracic surgery.

[3]  Christian Sticherling,et al.  Remote monitoring of cardiovascular implantable electronic devices: prerequisite or luxury? , 2009, Swiss medical weekly.

[4]  T Mussivand,et al.  Progress with the HeartSaver Ventricular Assist Device. , 1999, The Annals of thoracic surgery.

[5]  K S Holmes,et al.  The HeartSaver left ventricular assist device: an update. , 2001, The Annals of thoracic surgery.

[6]  M. Linnaluoto,et al.  Remote monitoring of implantable cardioverter defibrillator patients: a safe, time-saving, and cost-effective means for follow-up , 2008, Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology.

[7]  Renato Pietro Ricci,et al.  Telemedicine and cardiac implants: what is the benefit? , 2012, European heart journal.

[8]  Wojciech Zareba,et al.  ISHNE/EHRA expert consensus on remote monitoring of cardiovascular implantable electronic devices (CIEDs). , 2012, Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology.