Validity of inspiratory and expiratory methods of measuring gas exchange with a computerized system.

The accuracy of a computerized metabolic system, using inspiratory and expiratory methods of measuring ventilation, was assessed in eight male subjects. Gas exchange was measured at rest and during five stages on a cycle ergometer. Pneumotachometers were placed on the inspired and expired side to measure inspired (VI) and expired ventilation (VE). The devices were connected to two systems sampling expired O(2) and CO(2) from a single mixing chamber. Simultaneously, the criterion (Douglas bag, or DB) method assessed VE and fractions of O(2) and CO(2) in expired gas (FE(O(2)) and FE(CO(2))) for subsequent calculation of O(2) uptake (VO(2)), CO(2) production (VCO(2)), and respiratory exchange ratio. Both systems accurately measured metabolic variables over a wide range of intensities. Though differences were found between the DB and computerized systems for FE(O(2)) (both inspired and expired systems), FE(CO(2)) (expired system only), and VO(2) (inspired system only), the differences were extremely small (FE(O(2)) = 0.0004, FE(CO(2)) = -0.0003, VO(2) = -0.018 l/min). Thus a computerized system, using inspiratory or expiratory configurations, permits extremely precise measurements to be made in a less time-consuming manner than the DB technique.

[1]  J H Wilmore,et al.  Semiautomated systems approach to the assessment of oxygen uptake during exercise. , 1974, Journal of applied physiology.

[2]  J M Bland,et al.  Statistical methods for assessing agreement between two methods of clinical measurement , 1986 .

[3]  R. E. Johnson,et al.  Production of gaseous nitrogen in human steady-state conditions. , 1972, Journal of applied physiology.

[4]  J. Wilmore,et al.  Adequacy of the Haldane transformation in the computation of exercise V O2 in man. , 1973, Journal of applied physiology.

[5]  P G Versteeg,et al.  Automated Systems for Measurement of Oxygen Uptake During Exercise Testing , 1989, International journal of sports medicine.

[6]  J H Wilmore,et al.  An automated system for assessing metabolic and respiratory function during exercise. , 1976, Journal of applied physiology.

[7]  V B Unnithan,et al.  Validation of the sensormedics (S2900Z) metabolic cart for pediatric exercise testing. , 1994, Canadian journal of applied physiology = Revue canadienne de physiologie appliquee.

[8]  M. Kleiber Physiological Measurements of Metabolic Functions in Man. C. Frank Consolazio, Robert E. Johnson, and Louis J. Pecora. McGraw-Hill, New York, 1963. x + 505 pp. Illus. $14.50 , 1963 .

[9]  R M Gardner,et al.  Turbine flowmeter vs. Fleisch pneumotachometer: a comparative study for exercise testing. , 1987, Journal of applied physiology.

[10]  Barbara E. Ainsworth,et al.  VALIDATION OF AEROSPORT KB1-C PORTABLE METABOLIC SYSTEM , 1999 .

[11]  G. Gaesser,et al.  Assessment of the aerosport TEEM 100 portable metabolic measurement system. , 1996, Medicine and science in sports and exercise.

[12]  P. F. Scholander Analyzer for accurate estimation of respiratory gases in one-half cubic centimeter samples. , 1947, The Journal of biological chemistry.

[13]  F. Dolgener,et al.  Expired air temperature during steady-state running. , 1995, Medicine and science in sports and exercise.

[14]  Lawrence E. Armstrong,et al.  Variability of Respiration and Metabolism: Responses to Submaximal Cycling and Running , 1985 .

[15]  K. Wasserman,et al.  Evaluation of a symmetrically disposed Pitot tube flowmeter for measuring gas flow during exercise. , 1994, Journal of applied physiology.

[16]  R. Gardner,et al.  Computerized determination of pneumotachometer characteristics using a calibrated syringe. , 1982, Journal of applied physiology: respiratory, environmental and exercise physiology.

[17]  C Peel,et al.  Oxygen consumption using the K2 telemetry system and a metabolic cart. , 1993, Medicine and science in sports and exercise.

[18]  Edward T. Howley,et al.  Exercise Physiology: Theory and Application to Fitness and Performance , 1995 .