Digitization and calibration of the expendable bathythermograph
暂无分享,去创建一个
Abstract A study was undertaken of signal digitization and temperature calibration in expendable bathythermographs (XBT's) to learn how to minimize temperature errors in that system. An XBT digitizer was built into a PC-type microcomputer and used to calibrate 24 XBT probes at 5 temperature points, and later, to calibrate 72 probes at a single temperature. Twenty of the first set of probes were fastened rigidly in pairs and dropped in the ocean as a field test of the calibrations. Calibration of individual probes reduced the standard deviation of temperature calibration errors from around 0.05°C to