Stapedectomy — A Long-Term Report
暂无分享,去创建一个
The long-term results with large fenestra stapedectomy with vein graft and Teflon piston are compared with results with the small fenestra stapedectomy with Teflon piston directly into the vestibule. There were 1,943 operations in the former group and 2,155 in the latter when compared in 1970. One hundred consecutive patients from the beginning of each group with follow-up to present were compared. Results were generally the same with no great change in 15 and 20 years as compared to those at 5 years. The complication of perilymph fistula was caused by creating an opening in the footplate much larger than the prosthesis and was eliminated by interposing a living oval window seal if the opening was much larger than the prosthesis and a flap of lining membrane from the promontory when it was not. Other factors that influence a good result are discussed, including the type and the diameter of the piston used, the type of living oval window seal and the method of attachment to the incus. The small fenestra operation was found to be superior to the large, not only for the hearing gain achieved, but the ease of performance and the freedom from complications due to migration of the prosthesis and/or the oval window seal. At present we have done about all that can be done for the conductive components. What remains is the sensorineural component which our studies indicate may be due to an autoimmune response.
[1] U. Fisch,et al. Argon Laser Stapedotomy , 1982, The Annals of otology, rhinology, and laryngology.
[2] T. Mcgee. Comparison of Small Fenestra and Total Stapedectomy , 1981, The Annals of otology, rhinology, and laryngology.
[3] J. Shea. Eversion of the lining membrane of the vestibule , 1974, The Laryngoscope.
[4] J. J. Shea. The teflon piston operation for otosclerosis , 1963, The Laryngoscope.
[5] J. J. Shea,et al. Teflon Piston Operation for Otosclerosis , 1962 .