Mitigate B1+ inhomogeneity by nonlinear gradients and RF shimming

High-field MRI has the challenge of inhomogeneous B1+ and consequently an inhomogeneous flip angle distribution. This causes spatially dependent contrast and makes clinical diagnosis difficult. Under the small flip angle approximation and using nonlinear spatial encoding magnetic fields (SEMs), we propose a method to remap the B1+ map into a lower dimension coordinate system. Combining with RF shimming method, a simple pulse sequence design using nonlinear SEMs can achieve a homogenous flip angle distribution efficiently. Using simulations, we demonstrate that combining RF shimming and spatially selective RF excitation using generalized SEMs (SAGS) using linear and quadratic SEMs in a multi-spoke k-space trajectory can mitigate the B1+ inhomogeneity at 7T efficiently without using parallel RF transmission.

[1]  Michael B. Smith,et al.  Exploring the limits of RF shimming for high‐field MRI of the human head , 2006, Magnetic resonance in medicine.

[2]  Douglas C Noll,et al.  Fast‐kz three‐dimensional tailored radiofrequency pulse for reduced B1 inhomogeneity , 2006, Magnetic resonance in medicine.

[3]  P. Börnert,et al.  Transmit SENSE , 2003, Magnetic resonance in medicine.

[4]  R. Goebel,et al.  7T vs. 4T: RF power, homogeneity, and signal‐to‐noise comparison in head images , 2001, Magnetic resonance in medicine.

[5]  T. Ibrahim,et al.  Effect of RF coil excitation on field inhomogeneity at ultra high fields: a field optimized TEM resonator. , 2001, Magnetic resonance imaging.

[6]  Johnpauly A k-Space Analysis of Small-Tip-Angle Excitation , 2012 .

[7]  Peter Andersen,et al.  9.4T human MRI: Preliminary results , 2006, Magnetic resonance in medicine.

[8]  Michael B. Smith,et al.  Combination of optimized transmit arrays and some receive array reconstruction methods can yield homogeneous images at very high frequencies , 2005, Magnetic resonance in medicine.

[9]  J W Carlson,et al.  Electromagnetic fields of surface coil in vivo NMR at high frequencies , 1991, Magnetic resonance in medicine.

[10]  Douglas C Noll,et al.  Spatial domain method for the design of RF pulses in multicoil parallel excitation , 2006, Magnetic resonance in medicine.

[11]  Michael B. Smith,et al.  Central brightening due to constructive interference with, without, and despite dielectric resonance , 2005, Journal of magnetic resonance imaging : JMRI.

[12]  Mark E Ladd,et al.  RF excitation using time interleaved acquisition of modes (TIAMO) to address B1 inhomogeneity in high‐field MRI , 2010, Magnetic resonance in medicine.

[13]  D. Sodickson,et al.  Electrodynamic constraints on homogeneity and radiofrequency power deposition in multiple coil excitations , 2009, Magnetic resonance in medicine.

[14]  G. Metzger,et al.  Local B1+ shimming for prostate imaging with transceiver arrays at 7T based on subject‐dependent transmit phase measurements , 2008, Magnetic resonance in medicine.

[15]  E. Atalar,et al.  Specific absorption rate reduction using nonlinear gradient fields , 2013, Magnetic resonance in medicine.

[16]  Steen Moeller,et al.  A geometrically adjustable 16‐channel transmit/receive transmission line array for improved RF efficiency and parallel imaging performance at 7 Tesla , 2008, Magnetic resonance in medicine.

[17]  Oliver Speck,et al.  Parallel imaging in non-bijective, curvilinear magnetic field gradients: a concept study , 2008, Magnetic Resonance Materials in Physics, Biology and Medicine.

[18]  Dan Xu,et al.  Reduced field‐of‐view excitation using second‐order gradients and spatial‐spectral radiofrequency pulses , 2013, Magnetic resonance in medicine.

[19]  P A Bottomley,et al.  RF magnetic field penetration, phase shift and power dissipation in biological tissue: implications for NMR imaging. , 1978, Physics in medicine and biology.

[20]  Gigi Galiana,et al.  O‐space imaging: Highly efficient parallel imaging using second‐order nonlinear fields as encoding gradients with no phase encoding , 2010, Magnetic resonance in medicine.

[21]  Steen Moeller,et al.  B1 destructive interferences and spatial phase patterns at 7 T with a head transceiver array coil , 2005, Magnetic resonance in medicine.

[22]  D. Alsop,et al.  A spiral volume coil for improved RF field homogeneity at high static magnetic field strength , 1998, Magnetic resonance in medicine.

[23]  K. Uğurbil,et al.  Efficient high‐frequency body coil for high‐field MRI , 2004, Magnetic resonance in medicine.

[24]  J. Duyn,et al.  Tailored excitation using nonlinear B0‐shims , 2012, Magnetic resonance in medicine.