Clinical practice and evaluation of electronic portal imaging device for VMAT quality assurance.

Volumetric-modulated arc therapy (VMAT) is a novel extension of the intensity-modulated radiation therapy (IMRT) technique, which has brought challenges to dose verification. To perform VMAT pretreatment quality assurance, an electronic portal imaging device (EPID) can be applied. This study's aim was to evaluate EPID performance for VMAT dose verification. First, dosimetric characteristics of EPID were investigated. Then 10 selected VMAT dose plans were measured by EPID with the rotational method. The overall variation of EPID dosimetric characteristics was within 1.4% for VMAT. The film system serving as a conventional tool for verification showed good agreement both with EPID measurements ([94.1 ± 1.5]% with 3 mm/3% criteria) and treatment planning system (TPS) calculations ([97.4 ± 2.8]% with 3 mm/3% criteria). In addition, EPID measurements for VMAT presented good agreement with TPS calculations ([99.1 ± 0.6]% with 3 mm/3% criteria). The EPID system performed the robustness of potential error findings in TPS calculations and the delivery system. This study demonstrated that an EPID system can be used as a reliable and efficient quality assurance tool for VMAT dose verification.

[1]  C. Ling,et al.  Commissioning and quality assurance of RapidArc radiotherapy delivery system. , 2008, International journal of radiation oncology, biology, physics.

[2]  Ron S Sloboda,et al.  Quality assurance measurements of a-Si EPID performance. , 2004, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[3]  Suresh Senan,et al.  Volumetric intensity-modulated arc therapy vs. conventional IMRT in head-and-neck cancer: a comparative planning and dosimetric study. , 2009, International journal of radiation oncology, biology, physics.

[4]  Testing the GlaaS algorithm for dose measurements on low- and high-energy photon beams using an amorphous silicon portal imager. , 2008, Medical physics.

[5]  A Fogliata,et al.  Quality assurance of RapidArc in clinical practice using portal dosimetry. , 2011, The British journal of radiology.

[6]  Peter B Greer,et al.  Off-axis dose response characteristics of an amorphous silicon electronic portal imaging device. , 2007, Medical physics.

[7]  L. Cozzi,et al.  Testing the portal imager GLAaS algorithm for machine quality assurance , 2008, Radiation oncology.

[8]  Lei Dong,et al.  Dosimetric accuracy of Kodak EDR2 film for IMRT verifications. , 2005, Medical physics.

[9]  Carsten Brink,et al.  Single Arc Volumetric Modulated Arc Therapy of head and neck cancer. , 2010, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[10]  Lei Dong,et al.  Dosimetry tools and techniques for IMRT. , 2011, Medical physics.

[11]  A. Nahum,et al.  IMAT-SIM: a new method for the clinical dosimetry of intensity-modulated arc therapy (IMAT). , 2007, Medical physics.

[12]  Stine Korreman,et al.  The GLAaS algorithm for portal dosimetry and quality assurance of RapidArc, an intensity modulated rotational therapy , 2008, Radiation oncology.

[13]  Luca Cozzi,et al.  A treatment planning study comparing volumetric arc modulation with RapidArc and fixed field IMRT for cervix uteri radiotherapy. , 2008, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[14]  Lalith Kumaraswamy,et al.  An effective correction algorithm for off-axis portal dosimetry errors. , 2009, Medical physics.

[15]  Luca Cozzi,et al.  Simultaneous integrated boost radiotherapy for bilateral breast: a treatment planning and dosimetric comparison for volumetric modulated arc and fixed field intensity modulated therapy , 2009, Radiation oncology.

[16]  D. Low,et al.  A technique for the quantitative evaluation of dose distributions. , 1998, Medical physics.

[17]  D. Georg,et al.  Dose-response characteristics of an amorphous silicon EPID. , 2005, Medical physics.

[18]  M van Herk,et al.  Comparison of ghosting effects for three commercial a-Si EPIDs. , 2006, Medical physics.

[19]  B. McCurdy,et al.  Dosimetric investigation and portal dose image prediction using an amorphous silicon electronic portal imaging device. , 2001, Medical physics.

[20]  Luca Cozzi,et al.  GLAaS: an absolute dose calibration algorithm for an amorphous silicon portal imager. Applications to IMRT verifications. , 2006, Medical physics.

[21]  Ke Sheng,et al.  Comparison of Elekta VMAT with helical tomotherapy and fixed field IMRT: plan quality, delivery efficiency and accuracy. , 2010, Medical physics.

[22]  Pejman Rowshanfarzad,et al.  Measurement and modeling of the effect of support arm backscatter on dosimetry with a varian EPID. , 2010, Medical physics.

[23]  Tom Depuydt,et al.  The use of an aSi-based EPID for routine absolute dosimetric pre-treatment verification of dynamic IMRT fields. , 2004, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[24]  Sean L Berry,et al.  A field size specific backscatter correction algorithm for accurate EPID dosimetry. , 2010, Medical physics.

[25]  Luca Cozzi,et al.  Intensity modulation with photons for benign intracranial tumours: a planning comparison of volumetric single arc, helical arc and fixed gantry techniques. , 2008, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[26]  Peter Munro,et al.  Dosimetric verification of IMAT delivery with a conventional EPID system and a commercial portal dose image prediction tool. , 2009, Medical physics.