A quasi-dynamic nonlinear finite element model to investigate prosthetic interface stresses during walking for trans-tibial amputees.

BACKGROUND To predict the interface pressure between residual limb and prosthetic socket for trans-tibial amputees during walking. METHODS A quasi-dynamic finite element model was built based on the actual geometry of residual limb, internal bones and socket liner. To simulate the friction/slip boundary conditions between the skin and liner, automated surface-to-surface contact was used. Besides variable external loads and material inertia, the coupling between the large rigid displacement of knee joint and small elastic deformation of residual limb and prosthetic components were also considered. RESULTS Interface pressure distribution was found to have the same profile during walking. The high pressures fall over popliteal depression, middle patella tendon, lateral tibia and medial tibia regions. Interface pressure predicted by static or quasi-dynamic analysis had the similar double-peaked waveform shape in stance phase. INTERPRETATION The consideration of inertial effects and motion of knee joint cause 210% average variation of the area between the pressure curve and the horizontal line of pressure threshold between two cases, even though there is only a small change in the peak pressure. The findings in this paper show that the coupling dynamic effects of inertial loads and knee flexion must be considered to study interface pressure between residual limb and prosthetic socket during walking.

[1]  Chih-Han Chang,et al.  Effects of liner stiffness for trans-tibial prosthesis: a finite element contact model. , 2004, Medical engineering & physics.

[2]  V. C. Roberts,et al.  Clinical investigation of the pressure and shear stress on the trans-tibial stump with a prosthesis. , 1998, Medical engineering & physics.

[3]  H B Skinner,et al.  Analysis of a below-knee patellar tendon-bearing prosthesis: a finite element study. , 1991, Journal of rehabilitation research and development.

[4]  Dr D. P. Reynolds,et al.  Interface load analysis for computer-aided design of below-knee prosthetic sockets , 1992, Medical and Biological Engineering and Computing.

[5]  V. C. Roberts,et al.  Development of a non-linear finite element modelling of the below-knee prosthetic socket interface. , 1995, Medical engineering & physics.

[6]  D. Childress,et al.  Normal and shear stresses on a residual limb in a prosthetic socket during ambulation: comparison of finite element results with experimental measurements. , 1993, Journal of rehabilitation research and development.

[7]  W. Lee,et al.  Load transfer mechanics between trans-tibial prosthetic socket and residual limb--dynamic effects. , 2004, Journal of biomechanics.

[8]  Xiaohong Jia,et al.  Finite element modeling of the contact interface between trans-tibial residual limb and prosthetic socket. , 2004, Medical engineering & physics.

[9]  S G Zachariah,et al.  Finite element estimates of interface stress in the trans-tibial prosthesis using gap elements are different from those using automated contact. , 2000, Journal of biomechanics.

[10]  M Zhang,et al.  Finite element modelling of a residual lower-limb in a prosthetic socket: a survey of the development in the first decade. , 1998, Medical engineering & physics.

[11]  D S Childress,et al.  Generic, geometric finite element analysis of the transtibial residual limb and prosthetic socket. , 1997, Journal of rehabilitation research and development.

[12]  Tracy L Beil,et al.  Interface pressures during ambulation using suction and vacuum-assisted prosthetic sockets. , 2002, Journal of rehabilitation research and development.