Accurate high-current DC current ratio measurements
暂无分享,去创建一个
A setup has been developed for the accurate calibration of DC current ratios for DC currents up to 600 A. The setup is based on a DC current comparator and suitable for the calibration of DC current ratio measurement devices up to 100 A primary current. For higher primary currents, multiple turns of the primary conductor are made through the device under test. It has been verified that for well-designed devices this adds less than 1 part in 106 to the measurement uncertainty. A preliminary evaluation indicates that the total uncertainty of the setup is better than 5 parts in 106 in current ratio for primary currents up to 600 A.
[1] N. Kusters. The Precise Measurement of Current Ratios , 1964 .
[2] M. Kraft. Measurement Techniques of Low-Value High-Current Single-Range Current Shunts from 15 Amps to 3000 Amps , 2006 .
[3] Gert Rietveld,et al. Automated Low-Ohmic Resistance Measurements at the $\mu \Omega/\Omega$ Level , 2007, IEEE Transactions on Instrumentation and Measurement.
[4] Gert Rietveld,et al. Low-Ohmic Resistance Comparison: Measurement Capabilities and Resistor Traveling Behavior , 2013, IEEE Transactions on Instrumentation and Measurement.