Comparison of a lock-in amplifier and a digital sampling voltmeter as null detectors for differential measurements between sinusoidal and stepwise approximated ac signals
暂无分享,去创建一个
This paper presents the results of using two different types of a null detector, a lock-in amplifier and a digital sampling voltmeter, to measure the difference between a sinusoidal and a stepwise approximated AC voltage generated by means of a programmable Josephson voltage standard. Preliminary results of ac-dc difference measurement made on a thermal converter using the two null detectors agree within 0.5 µV/V at 10 Hz and 20 Hz.
[1] Harald Moser,et al. Evaluation of the synchronous generation and sampling technique , 2003, IEEE Trans. Instrum. Meas..
[2] Alain Rüfenacht,et al. Systematic Error Analysis of Stepwise-Approximated AC Waveforms Generated by Programmable Josephson Voltage Standards , 2009, IEEE Transactions on Instrumentation and Measurement.