Inertance estimation in a lumped-parameter hydraulic simulator of human circulation.

Pulsatile mock loop systems are largely used to investigate the cardiovascular system in vitro. They consist of a pump, which replicates the heart, coupled with a lumped-parameter hydraulic afterload, which simulates vasculature. An accurate dimensioning of components is required for a reliable mimicking of the physiopathological behavior of the system. However, it is not possible to create a component for the afterload inertance, and inertance contributions are present in the entire circuit. Hence, in the literature, inertance is neglected or qualitatively evaluated. In this paper, we propose two quantitative methods (Maximum-likelihood estimation (MLE) and Bayesian estimation) for estimating afterload inertance based on observed pressure and flow waveforms. These methods are also applied to a real mock loop system. Results show that the system has an inertance comparable with the literature reference value of the entire systemic circulation, and that the expected variations over inlet average flow and pulse frequency are in general confirmed. Comparing the methods, the Bayesian approach results in higher and more stable estimations than the classical MLE.

[1]  F. Ruggeri,et al.  Nonlinear functional response parameter estimation in a stochastic predator-prey model. , 2011, Mathematical biosciences and engineering : MBE.

[2]  S. Sherwin,et al.  Pulse wave propagation in a model human arterial network: Assessment of 1-D visco-elastic simulations against in vitro measurements , 2011, Journal of biomechanics.

[3]  Gianfranco Beniamino Fiore,et al.  Response to the Letter to the Editor: A New Pulsatile Volumetric Device With Biomorphic Valves for the In Vitro Study of the Cardiovascular System by M.B. Munir et al. , 2011 .

[4]  Ettore Lanzarone,et al.  A new pulsatile volumetric device with biomorphic valves for the in vitro study of the cardiovascular system. , 2009, Artificial organs.

[5]  Roberto Fumero,et al.  Integrated Model of Endothelial NO Regulation and Systemic Circulation for the Comparison Between Pulsatile and Continuous Perfusion , 2009, IEEE Transactions on Biomedical Engineering.

[6]  Aron Andrade,et al.  Mock circulatory system for the evaluation of left ventricular assist devices, endoluminal prostheses, and vascular diseases. , 2008, Artificial organs.

[7]  S. Sherwin,et al.  Lumped parameter outflow models for 1-D blood flow simulations: Effect on pulse waves and parameter estimation , 2008 .

[8]  B. Yoffe,et al.  Experimental study of a novel suture-less aortic anastomotic device. , 2007, European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery.

[9]  G. Baselli,et al.  Model of arterial tree and peripheral control for the study of physiological and assisted circulation. , 2007 .

[10]  R. Figliola,et al.  In vitro study of flow regulation for pulmonary insufficiency. , 2007, Journal of biomechanical engineering.

[11]  Yi Wu,et al.  Construction of an Artificial Heart Pump Performance Test System , 2006, Cardiovascular engineering.

[12]  Jeffrey R. Gohean,et al.  Comparative in vitro study of bileaflet and tilting disk valve behavior in the pulmonary position. , 2006, Journal of biomechanical engineering.

[13]  Ajit P Yoganathan,et al.  The material properties of the native porcine mitral valve chordae tendineae: an in vitro investigation. , 2006, Journal of biomechanics.

[14]  Andrew Galbraith,et al.  A complete mock circulation loop for the evaluation of left, right, and biventricular assist devices. , 2005, Artificial organs.

[15]  A. Yoganathan,et al.  Mitral Valve Function and Chordal Force Distribution Using a Flexible Annulus Model: An In Vitro Study , 2005, Annals of Biomedical Engineering.

[16]  Eric Bertrand,et al.  In vitro atrial flow dynamics: normal conditions versus atrial fibrillation. , 2004, Journal of biomechanics.

[17]  George M Pantalos,et al.  Characterization of an Adult Mock Circulation for Testing Cardiac Support Devices , 2004, ASAIO journal.

[18]  S. James Press,et al.  Subjective and objective Bayesian statistics : principles, models, and applications , 2003 .

[19]  Martyn Plummer,et al.  JAGS: A program for analysis of Bayesian graphical models using Gibbs sampling , 2003 .

[20]  Patrick Segers,et al.  Mock loop testing of On-x prosthetic mitral valve with Doppler echocardiography. , 2002, Artificial organs.

[21]  A. Gallant,et al.  Numerical Techniques for Maximum Likelihood Estimation of Continuous-Time Diffusion Processes , 2002 .

[22]  P. Walker,et al.  The influence of ventricular input impedance on the hydrodynamic performance of bioprosthetic aortic roots in vitro. , 2001, The Journal of heart valve disease.

[23]  M K Sharp,et al.  Development of a hydraulic model of the human systemic circulation. , 1999, ASAIO journal.

[24]  T. Dohi,et al.  Nonsymmetric leaflet motion of St. Jude Medical mitral valves simulated with a computer-controlled hydraulic mock circulator. , 2008, Artificial organs.

[25]  Dani Gamerman,et al.  Markov Chain Monte Carlo: Stochastic Simulation for Bayesian Inference , 1997 .

[26]  K Sagawa,et al.  Translation of Otto Frank's paper "Die Grundform des Arteriellen Pulses" Zeitschrift für Biologie 37: 483-526 (1899). , 1990, Journal of molecular and cellular cardiology.

[27]  B. Øksendal Stochastic differential equations : an introduction with applications , 1987 .

[28]  R Fumero,et al.  Fluid dynamic models as a guide to determine prosthetic heart valve diameter. , 1986, Journal of biomechanics.

[29]  N. Westerhof,et al.  Input Impedance of the Pulmonary Arterial System in Normal Man: Effects of Respiration and Comparison to Systemic Impedance , 1984, Circulation research.

[30]  D M McQueen,et al.  In vitro hydrodynamic comparison of mitral valve prostheses at high flow rates. , 1978, The Journal of thoracic and cardiovascular surgery.

[31]  J F Cornhill,et al.  An aortic--left ventricular pulse duplicator used in testing prosthetic aortic heart valves. , 1977, The Journal of thoracic and cardiovascular surgery.

[32]  Swanson Wm,et al.  A simple cardiovascular system simulator: design and performance. , 1977 .

[33]  N. Westerhof,et al.  An artificial arterial system for pumping hearts. , 1971, Journal of applied physiology.

[34]  A Noordergraaf,et al.  Analog studies of the human systemic arterial tree. , 1969, Journal of biomechanics.

[35]  L. Katz,et al.  CLINICAL OBSERVATIONS ON THE DYNAMICS OF VENTRICULAR SYSTOLE: I. AURICULAR FIBRILLATION , 1923 .