Physical Characteristics Of The Ecat 953b/31: A New High Resolution Brain Positron Tomograph

A high spatial resolution brain positron tomograph, the ECAT 953B/31, is presented. The machine consists of two 76.5 cm diameter rings (patient port diameter: 36 cm) made of 24 8x8 BGO detector blocks each, each block being coupled to four 1 in(2) Hammamatsu phototubes. The machine has 15 9 cmx1 mm motorized removable septa, which allows reconstruction of 31 slices 3.375 mm apart. The transaxial resolution (FWHM) in the reconstructed image (with wobbling) is 4.5 mm at center, 5.2 mm (tangential) and 5.6 mm at 10 cm from center. With septa out of the field of view (FOV), the axial resolution degrades to 5.5 mm a center and 6.3 mm at 10 cm from center. At 1 muCi/cc with a 250 keV threshold and a 12 ns coincidence window, the sensitivity is 146100 true events, the ratio of randoms to trues is 0.10, the scatter fraction is 17% and the dead-time losses are 30%. With septa out of the FOV, the sensitivity is increased by a factor 3.6, while the scatter fraction reaches 0.9. Images obtained with (18 )F-DG and H(2)(15)O in human brains and (18 )F-DOPA in a baboon brain demonstrate that the ECAT 953B/31 will be suitable for high resolution and for low count rate brain studies.

[1]  M. Endo,et al.  Development of a high resolution PET , 1990 .

[2]  E. Hoffman,et al.  Measuring PET scanner sensitivity: relating countrates to image signal-to-noise ratios using noise equivalents counts , 1990 .

[3]  B. Mazoyer,et al.  Physical characteristics of TTV03, a new high spatial resolution time-of-flight positron tomograph , 1990 .

[4]  S. Holte,et al.  Evaluation of the Karolinska new positron camera system; the Scanditronix PC2048-15B , 1990 .

[5]  M. Casey,et al.  Design of a super fast three-dimensional projection system for positron emission tomography , 1990 .

[6]  Iwao Kanno,et al.  Design and evaluation of HEADTOME-IV, a whole-body positron emission tomograph , 1989 .

[7]  Thomas K. Lewellen,et al.  Performance measurements of the SP3000/UW time-of-flight positron emission tomograph , 1988 .

[8]  M. Gilardi,et al.  Physical performance of the latest generation of commercial positron scanner , 1988 .

[9]  E. Hoffman,et al.  Quantitation in positron emission computed tomography: 7. A technique to reduce noise in accidental coincidence measurements and coincidence efficiency calibration. , 1986, Journal of computer assisted tomography.

[10]  R. Nutt,et al.  A Multicrystal Two Dimensional BGO Detector System for Positron Emission Tomography , 1986, IEEE Transactions on Nuclear Science.

[11]  C. Bohm,et al.  Correction for Scattered Radiation in a Ring Detector Positron Camera by Integral Transformation of the Projections , 1983, Journal of computer assisted tomography.