Direct Detection of Red Blood Cell Fragments: A New Flow Cytometric Method to Evaluate Hemolysis in Blood Pumps

Pump induced hemolysis is presently evaluated by measuring plasma free hemoglobin (fHb). However, this method has disadvantages because quantification of fHb depends on hematocrit (HCT) and hemoglobin (Hb) levels. The aim of this work was to devise a hemoglobin independent method, capable of quantifying cell trauma directly by measuring the number of red blood cell (RBC) fragments. Whole blood flow cytometry was used to quantify circulating RBC fragments derived from a roller pump (Sarns, Inc. Model 2 M 6,002) and a centrifugal pump (Gyro C1E3, Kyocera Corp.). The pumps were tested in a mock circuit for 2 hr (5 L/min flow against 100 mm Hg pressure head). Red blood cell fragments were quantified by a phycoerythrin (PE) labeled glycophorin A antibody specific for erythrocytes. Red blood cell fragments were smaller than the intact RBC population and overlapped in size with the platelet population (based on forward- and side-light scattering measurements). For the roller pump, the values for RBC fragments increased from 1,090 ± 260/&mgr;l at 0 min to 14,880 ± 5,900/&mgr;l after 120 min. In contrast, using the centrifugal pump, there was little increase in RBC fragments (from 730 ± 270/&mgr;l at 0 min to 1,400 ± 840/&mgr;l after 120 min). Flow cytometry can be used for the rapid, sensitive, hemoglobin independent evaluation of pump induced RBC trauma.

[1]  D B Olsen,et al.  In vitro hematological testing of rotary blood pumps: remarks on standardization and data interpretation. , 2008, Artificial organs.

[2]  H. Thoma,et al.  In vitro investigation of thrombogenesis in rotary blood pumps. , 2008, Artificial organs.

[3]  G. Rau,et al.  Platelet Function and Hemolysis in Centrifugal Pumps: In Vitro Investigations , 1999, The International journal of artificial organs.

[4]  K. Sanger,et al.  Superiority of centrifugal pump over roller pump in paediatric cardiac surgery: prospective randomised trial. , 1998, European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery.

[5]  E. Fosse,et al.  Centrifugal pump and heparin coating improves cardiopulmonary bypass biocompatibility. , 1996, The Annals of thoracic surgery.

[6]  Y Ohara,et al.  Does hematocrit affect in vitro hemolysis test results? Preliminary study with Baylor/NASA prototype axial flow pump. , 1994, Artificial organs.

[7]  H. Harasaki,et al.  Hemolysis. A comparative study of four nonpulsatile pumps. , 1988, ASAIO transactions.

[8]  C. Högman,et al.  Red Cell Preservation in Protein‐Poor Media: III. Protection Against in vitro Hemolysis , 1981, Vox sanguinis.

[9]  J. Moake,et al.  Increased von Willebrand factor (vWf) binding to platelets associated with impaired vWf breakdown in thrombotic thrombocytopenic purpura , 1998, Journal of clinical apheresis.

[10]  Y Nosé,et al.  The need for standardizing the index of hemolysis. , 1994, Artificial organs.

[11]  R. Varco,et al.  SOME BIOLOGIC LIMITATIONS TO PROLONGED BLOOD PUMPING. , 1965, Transactions - American Society for Artificial Internal Organs.