Correlation of Impedance Minute Ventilation with Measured Minute Ventilation in a Rate Responsive Pacemaker

COLE, C.R., et al.: Correlation of Impedance Minute Ventilation with Measured Minute Ventilation in a Rate Responsive Pacemaker. Although rate responsive pacing based on impedance minute ventilation (IMV) is now standard, there is almost no data confirming the relationship between IMV from an implanted pacemaker and measured minute ventilation (VE) during exercise. Nineteen completely paced adults implanted with Medtronic Kappa 400 pacemakers underwent symptom‐limited maximal metabolic treadmill testing using a modified Minnesota Pacemaker Response Protocol. Minute ventilation (VE, L/min) was simultaneously measured using the flowmeter of a respiratory metabolic gas analysis system and the transthoracic impedance minute ventilation circuitry of the pacemaker. Correlation coefficients (r) were used to find the best fit line to describe the relationship between the two measurements. Mean (± SD) r values for the first, second, and third order polynomial equations and for log and exponential equations were: 0.92 ± 0.08, 0.94 ± 0.04, 0.95 ± 0.04, 0.91 ± 0.06, and 0.91 ± 0.07, respectively. None of the r values were statistically different from the first order equation. Transthoracic IMV as measured by the Medtronic Kappa 400 is closely correlated to measured minute ventilation and is represented well by a first order (linear) equation.

[1]  J. Bonnet,et al.  Evaluation of a Dual Sensor Rate Responsive Pacing System Based on a New Concept , 1998, Pacing and clinical electrophysiology : PACE.

[2]  A. Woodend,et al.  Correlation of the Heart Rate‐Minute Ventilation Relationship with Clinical Data: Relevance to Rate‐Adaptive Pacing , 1997, Pacing and clinical electrophysiology : PACE.

[3]  H. Rickli,et al.  Age and Sex Related Changes in Heart Rate to Ventilation Coupling: Implications for Rate Adaptive Pacemaker Algorithms , 1997, Pacing and clinical electrophysiology : PACE.

[4]  G. Kay,et al.  Rate-modulated cardiac pacing based on transthoracic impedance measurements of minute ventilation: correlation with exercise gas exchange. , 1989, Journal of the American College of Cardiology.

[5]  G S Butrous,et al.  Comparison of exercise performance of six rate-adaptive right ventricular cardiac pacemakers. , 1989, The American journal of cardiology.

[6]  D Hunt,et al.  Rate Responsive Pacing Using a Minute Ventilation Sensor , 1988, Pacing and clinical electrophysiology : PACE.

[7]  T. Simmons,et al.  Electrode Configurations for a Respiratory Impedance Measurement Suitable for Rate Responsive Pacing , 1986, Pacing and clinical electrophysiology : PACE.

[8]  B. Pederson,et al.  Continuous Ventricular Volume Assessment for Diagnosis and Pacemaker Control , 1984, Pacing and clinical electrophysiology : PACE.

[9]  P Rossi,et al.  Increasing Cardiac Rate by Tracking the Respiratory Rate , 1984, Pacing and clinical electrophysiology : PACE.

[10]  G. Rognoni,et al.  Respiration as a reliable physiological sensor for controlling cardiac pacing rate. , 1984, British heart journal.

[11]  P Rossi,et al.  Respiratory Rate as a Determinant of Optimal Pacing Rate , 1983, Pacing and clinical electrophysiology : PACE.

[12]  J. S. Janicki,et al.  Physiologic correlates of the heart rate response to upright isotonic exercise: relevance to rate-responsive pacemakers. , 1988, Journal of the American College of Cardiology.