Design and performance testing of an axial-flow ventricular assist device developed at the Fu Wai Hospital in Beijing

Purpose Various ventricular assist devices (VADs) have been developed for clinical use in recent years. The aim of a multidisciplinary research team at the Fuwai Hospital of the Peking Union Medical College is to design and develop an axial flow left ventricular assist device (LVAD) for adults. Methods Using Computational Fluid Dynamics (CFD), the inflow characteristics of the axial flow pump were analyzed. After CFD analysis, the axial pump was fabricated using a 5-axis, computer numerical control (CNC) milling machine. Performances of the pump both in vitro and in vivo were tested. Results This VAD, which was developed after numerous CFD analyses for the flow characteristics of the pump, is 58.5 mm long, 30 mm wide and weighs 120 g. The pump can deliver 5 lpm for pressures of 100 mmHg over 8500 rpm. The NIH value was 0.01 g/100 L. The hemolysis, which was evaluated in an in vivo test, was a bit higher than the normal value, but remained within an acceptable range. Conclusions Performance of the pump in vitro and in vivo was considered sufficient for an LVAD. Further design improvements are being undertaken in terms of hemolysis and thrombosis to improve the biocompatibility of the pump.

[1]  Annetine Gelijns,et al.  Left ventricular assist devices as destination therapy: a new look at survival. , 2005, The Journal of thoracic and cardiovascular surgery.

[2]  Xinwei Song,et al.  TRANSIENT COMPUTATIONAL FLUID DYNAMICS STUDY OF AN AXIAL BLOOD PUMP , 2003 .

[3]  C Bludszuweit,et al.  Model for a general mechanical blood damage prediction. , 1995, Artificial organs.

[4]  Paul E. Allaire,et al.  Axial Flow Blood Pumps , 2003, ASAIO journal.

[5]  Y Nosé,et al.  The need for standardizing the index of hemolysis. , 1994, Artificial organs.

[6]  H. Reul,et al.  Estimation of Shear Stress-related Blood Damage in Heart Valve Prostheses - in Vitro Comparison of 25 Aortic Valves , 1990, The International journal of artificial organs.

[7]  William R Wagner,et al.  Design optimization of blood shearing instrument by computational fluid dynamics. , 2005, Artificial organs.

[8]  L. Sauvage,et al.  Extracorporeal circulation: the role of the pulse in maintenance of the systemic circulation during heart-lung by-pass. , 1955, Surgery.

[9]  M Y Jaffrin,et al.  Hemolysis during membrane plasma separation with pulsed flow filtration enhancement. , 1994, Journal of biomechanical engineering.

[10]  D. Goldstein,et al.  Worldwide Experience With the MicroMed DeBakey Ventricular Assist Device® as a Bridge to Transplantation , 2003, Circulation.

[11]  P E Allaire,et al.  Numerical Solution for Blood Flow in a Centrifugal Ventricular Assist Device , 1999, The International journal of artificial organs.

[12]  H Reul,et al.  Leakage flow at mechanical heart valve prostheses: improved washout or increased blood damage? , 1999, The Journal of heart valve disease.

[13]  R. Paul,et al.  Shear stress related blood damage in laminar couette flow. , 2003, Artificial organs.