Estimating SpO2 via Time-Efficient High-Resolution Harmonics Analysis and Maximum Likelihood Tracking

The accuracy of noninvasive oxygen saturation (SpO2), which is defined by the measurements based on photoplethysmographic (PPG) signals, is intensively affected by motion artifacts (MAs) and low perfusion. This study introduces a novel approach called ESPRIT-MLT to measure SpO2 when such interferences are present. In contrast to previous studies, the work focuses on the harmonic model of the PPG signal and the probability model of results from harmonic analysis. The optimized parametric ESPRIT method is applied to improve the accuracy of harmonic power estimation, and the maximum likelihood SpO2 tracking (MLT) technique is proposed to track the most probable uncontaminated harmonic of heart rate frequency. We construct an evaluation platform for testing the proposed method via generated signals and subject tests. Compared with the nonparametric periodogram method, the probability of correct harmonics being found is improved by 18.7% or 19.7%, when the signal is contaminated by motion artifacts or affected by low perfusion, respectively. In comparison with the reference methods, the proposed ESPRIT-MLT method exhibits a lower average root mean square error (RMSE) (1.17%) in the simulation using an MA-contaminated PPG signal, and a lower RMSE (2.70%) in the simulation using an extremely low (0.05%) perfusion index. A comprehensive subject test that consists of 4 activities and 20 subjects shows an average RMSE of 0.84% ($\pm$ 0.44%). Furthermore, the time-efficiency is optimized to be adaptable with wearable devices. Therefore, the proposed method has potential in enhancing the performance of clinical pulse oximetry and wearable SpO2 measurement devices for daily use.

[1]  Zhilin Zhang,et al.  Photoplethysmography-Based Heart Rate Monitoring in Physical Activities via Joint Sparse Spectrum Reconstruction , 2015, IEEE Transactions on Biomedical Engineering.

[2]  M. O'reilly Masimo Signal Extraction Technology Pulse Oximetry , 2011, Neonatology.

[3]  Sun K. Yoo,et al.  Motion artifact reduction in photoplethysmography using independent component analysis , 2006, IEEE Transactions on Biomedical Engineering.

[4]  K. Ashoka Reddy,et al.  Use of Fourier Series Analysis for Motion Artifact Reduction and Data Compression of Photoplethysmographic Signals , 2009, IEEE Transactions on Instrumentation and Measurement.

[5]  Tayard Desudchit,et al.  Comparative evaluation of adaptive filters in motion artifact cancellation for pulse oximetry , 2009, 2009 5th International Colloquium on Signal Processing & Its Applications.

[6]  Yuan-Ting Zhang,et al.  An Efficient Motion-Resistant Method for Wearable Pulse Oximeter , 2008, IEEE Transactions on Information Technology in Biomedicine.

[7]  E. Hari Krishna,et al.  Evaluation of wavelets for reduction of motion artifacts in photoplethysmographic signals , 2010, 10th International Conference on Information Science, Signal Processing and their Applications (ISSPA 2010).

[8]  Hartmut Gehring,et al.  The effects of motion artifact and low perfusion on the performance of a new generation of pulse oximeters in volunteers undergoing hypoxemia. , 2002, Respiratory care.

[9]  Guangyu Bin,et al.  Research on an anti-motion interference algorithm of blood oxygen saturation based on AC and DC analysis. , 2015, Technology and health care : official journal of the European Society for Engineering and Medicine.

[10]  Amal Jubran,et al.  Pulse oximetry , 1999 .

[11]  Y. Mendelson,et al.  A Comparative Evaluation of Adaptive Noise Cancellation Algorithms for Minimizing Motion Artifacts in a Forehead-Mounted Wearable Pulse Oximeter , 2007, 2007 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[12]  John G. Webster,et al.  Design of Pulse Oximeters , 1997 .

[13]  K. Ashoka Reddy,et al.  E2MD for reduction of motion artifacts from photoplethysmographic signals , 2014, 2014 International Conference on Electronics and Communication Systems (ICECS).

[14]  Adrian D. C. Chan,et al.  Effects of motion artifact on the blood oxygen saturation estimate in pulse oximetry , 2014, 2014 IEEE International Symposium on Medical Measurements and Applications (MeMeA).

[15]  Zhilin Zhang,et al.  TROIKA: A General Framework for Heart Rate Monitoring Using Wrist-Type Photoplethysmographic Signals During Intensive Physical Exercise , 2014, IEEE Transactions on Biomedical Engineering.

[16]  Mehrdad Nourani,et al.  A Motion-Tolerant Adaptive Algorithm for Wearable Photoplethysmographic Biosensors , 2014, IEEE Journal of Biomedical and Health Informatics.

[17]  T. Kailath,et al.  Estimation of Signal Parameters via Rotational Invariance Techniques - ESPRIT , 1986 .

[18]  Norbert Stuban,et al.  Non-invasive calibration method for pulse oximeters , 2008 .

[19]  Steve Warren,et al.  Two-Stage Approach for Detection and Reduction of Motion Artifacts in Photoplethysmographic Data , 2010, IEEE Transactions on Biomedical Engineering.

[20]  S J Barker,et al.  Effects of motion, ambient light, and hypoperfusion on pulse oximeter function. , 1998, Journal of clinical anesthesia.

[21]  E. Hari Krishna,et al.  Dual-tree complex wavelet transform for motion artifact reduction of PPG signals , 2012, 2012 IEEE International Symposium on Medical Measurements and Applications Proceedings.

[22]  Hyun Seok Choi,et al.  Motion Artifact Reduction in Blood Pressure Signals Using Adaptive Digital Filter with a Capacitive Sensor , 2007, 2007 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[23]  K. Sivani,et al.  HHT based signal decomposition for reduction of motion artifacts in photoplethysmographic signals , 2012, 2012 IEEE International Instrumentation and Measurement Technology Conference Proceedings.

[24]  Jianchu Yao,et al.  A Short Study to Assess the Potential of Independent Component Analysis for Motion Artifact Separation in Wearable Pulse Oximeter Signals , 2005, 2005 IEEE Engineering in Medicine and Biology 27th Annual Conference.

[25]  Thomas Kailath,et al.  ESPRIT-A subspace rotation approach to estimation of parameters of cisoids in noise , 1986, IEEE Trans. Acoust. Speech Signal Process..

[26]  Steven J. Barker,et al.  The effects of motion on the performance of pulse oximeters in volunteers (revised publication). , 1997 .

[27]  Sune Duun,et al.  Independent component analysis applied to pulse oximetry in the estimation of the arterial oxygen saturation (SpO2) - a comparative study , 2009, 2009 Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[28]  K. Reddy,et al.  Motion Artifact Reduction in Photoplethysmographic Signals using Singular Value Decomposition , 2007, 2007 IEEE Instrumentation & Measurement Technology Conference IMTC 2007.