Combining scatter reduction and correction to improve image quality in cone-beam computed tomography (CBCT).

PURPOSE The authors propose a combined scatter reduction and correction method to improve image quality in cone-beam computed tomography (CBCT). Although using a beam-block approach similar to previous techniques to measure the scatter, this method differs in that the authors utilize partially blocked projection data obtained during scatter measurement for CBCT image reconstruction. This study aims to evaluate the feasibility of the proposed approach. METHODS A 1D grid, composed of lead septa, was placed between the radiation source and the imaging object for scatter measurement. Image data were collected from the grid interspace regions while the scatter distribution was measured in the blocked regions under the grid. Scatter correction was performed by subtracting the measured scatter from the imaging data. Image information in the penumbral regions of the grid was derived. Three imaging modes were developed to reconstruct full CBCT images from partial projection data. The single-rotation half-fan mode uses interpolation to fill the missing data. The dual-rotation half-fan mode uses two rotations, with the grid offset by half a grid cycle, to acquire two complementary sets of projections, which are then merged to form complete projections for reconstruction. The single-rotation full-fan mode was designed for imaging a small object or a region of interest. Full-fan projection images were acquired over a 360 degrees scan angle with the grid shifting a distance during the scan. An enlarged Catphan phantom was used to evaluate potential improvement in image quality with the proposed technique. An anthropomorphic pelvis phantom was used to validate the feasibility of reconstructing a complete set of CBCT images from the partially blocked projections using three imaging modes. Rigid-body image registration was performed between the CBCT images from the single-rotation half-fan mode and the simulation CT and the results were compared to that for the CBCT images from dual-rotation mode and conventional CBCT images. RESULTS The proposed technique reduced the streak artifact index from 58% to 1% in comparison with the conventional CBCT. It also improved CT number linearity from 0.880 to 0.998 and the contrast-to-noise ratio (CNR) from 4.29 to 6.42. Complete sets of CBCT images with overall improved image quality were achieved for all three image modes. The longitudinal resolution was slightly compromised for the single-rotation half-fan mode. High resolution was retained for the dual-rotation half-fan and single-rotation full-fan modes in the longitudinal direction. The registration error for the CBCT images from the single-rotation half-fan mode was 0.8 +/- 0.3 mm in the longitudinal direction and negligible in the other directions. CONCLUSIONS The proposed method provides combined scatter correction and direct scatter reduction. Scatter correction may eliminate scatter artifacts, while direct scatter reduction may improve the CNR to compensate the CNR degradation due to scatter correction. Complete sets of CBCT images are reconstructed in all three imaging modes. The single-rotation mode can be used for rigid-body patient alignment despite degradation in longitudinal resolution. The dual-rotation mode may be used to improve CBCT image quality for soft tissue delineation in adaptive radiation therapy.

[1]  Christoph Hoeschen,et al.  The built-in capacity of CT D'Or's static ring for scatter correction , 2008, SPIE Medical Imaging.

[2]  O Majdani,et al.  Large scan field, high spatial resolution flat-panel detector based volumetric CT of the whole human skull base and for maxillofacial imaging. , 2007, Dento maxillo facial radiology.

[3]  Lei Zhu,et al.  Scatter Correction Method for X-Ray CT Using Primary Modulation: Theory and Preliminary Results , 2006, IEEE Transactions on Medical Imaging.

[4]  Fang-Fang Yin,et al.  Comparison of Similarity Measures for Rigid-body CT/Dual X-ray Image Registrations , 2007, Technology in cancer research & treatment.

[5]  Hans-Peter Howaldt,et al.  High resolution imaging of craniofacial bone specimens by flat-panel volumetric computed tomography. , 2008, Journal of cranio-maxillo-facial surgery : official publication of the European Association for Cranio-Maxillo-Facial Surgery.

[6]  W. Kalender,et al.  Combining deterministic and Monte Carlo calculations for fast estimation of scatter intensities in CT , 2006, Physics in medicine and biology.

[7]  K. P. Maher,et al.  Scatter correction in digital radiography using interpolated local sampling of aperture signals , 2001, Australasian Physics & Engineering Sciences in Medicine.

[8]  Lilach O Lerman,et al.  Assessment of myocardial microvascular function: new opportunities in fast computed tomography. , 2007, Trends in cardiovascular medicine.

[9]  Michael Grasruck,et al.  Musculoskeletal applications of flat-panel volume CT , 2008, Skeletal Radiology.

[10]  Weiguang Yao,et al.  An analytical approach to estimating the first order scatter in heterogeneous medium. II. A practical application. , 2009, Medical physics.

[11]  Lei Zhu,et al.  Scatter correction for cone-beam CT in radiation therapy. , 2009 .

[12]  Tien-Peng Wang,et al.  Cone Beam Breast CT with a Flat Panel Detector- Simulation, Implementation and Demonstration , 2005, 2005 IEEE Engineering in Medicine and Biology 27th Annual Conference.

[13]  Ruola Ning,et al.  X-ray scatter correction algorithm for cone beam CT imaging. , 2004, Medical physics.

[14]  Jim Malone,et al.  Computerized scatter correction in diagnostic radiology , 1997 .

[15]  J H Siewerdsen,et al.  Compensators for dose and scatter management in cone-beam computed tomography. , 2007, Medical physics.

[16]  S. Mori,et al.  Magnitude and effects of x-ray scatter in a 256-slice CT scanner. , 2006, Medical physics.

[17]  Jeroen J. Bax,et al.  Cardiac CT and CT coronary angiography: technology and application , 2008, Heart.

[18]  J A Sorenson,et al.  Performance characteristics of improved antiscatter grids. , 1980, Medical physics.

[19]  J. Wong,et al.  Flat-panel cone-beam computed tomography for image-guided radiation therapy. , 2002, International journal of radiation oncology, biology, physics.

[20]  Peter Munro,et al.  A quality assurance program for the on-board imager®. , 2006, Medical physics.

[21]  D. Jaffray,et al.  The influence of antiscatter grids on soft-tissue detectability in cone-beam computed tomography with flat-panel detectors. , 2004, Medical physics.

[22]  J. Dinten,et al.  A new method for x-ray scatter correction: first assessment on a cone-beam CT experimental setup , 2007, Physics in medicine and biology.

[23]  W. Kalender,et al.  Empirical cupping correction: a first-order raw data precorrection for cone-beam computed tomography. , 2006, Medical physics.

[24]  B. Wilson,et al.  Volume CT with a flat-panel detector on a mobile, isocentric C-arm: pre-clinical investigation in guidance of minimally invasive surgery. , 2005, Medical physics.

[25]  R. Figueroa,et al.  The use of Portable Intraoperative Computed Tomography Scanning for Real-Time Image Guidance: A Pilot Cadaver Study , 2008, American journal of rhinology.

[26]  S. Richard,et al.  A simple, direct method for x-ray scatter estimation and correction in digital radiography and cone-beam CT. , 2005, Medical physics.

[27]  P. Keall,et al.  Computational challenges for image-guided radiation therapy: framework and current research. , 2007, Seminars in radiation oncology.

[28]  Lei Zhu,et al.  Noise suppression in scatter correction for cone-beam CT. , 2009, Medical physics.

[29]  Shinichiro Mori,et al.  Four-dimensional measurement of lung tumor displacement using 256-multi-slice CT-scanner. , 2007, Lung cancer.

[30]  M. Persson,et al.  Total variation norm for three-dimensional iterative reconstruction in limited view angle tomography , 2001, Physics in medicine and biology.

[31]  X Pan Fast reconstruction with uniform noise properties in halfscan computed tomography. , 2000, Medical physics.

[32]  L. Xing,et al.  Iterative image reconstruction for CBCT using edge-preserving prior. , 2008, Medical physics.

[33]  D B Plewes,et al.  Quantitative digital subtraction angiography: two scanning techniques for correction of scattered radiation and veiling glare. Work in progress. , 1985, Radiology.

[34]  Lei Zhu,et al.  A Patient Set-up Protocol Based on Partially Blocked Cone-beam CT , 2010, Technology in cancer research & treatment.

[35]  Michael Poon,et al.  Technology Insight: cardiac CT angiography , 2006, Nature Clinical Practice Cardiovascular Medicine.

[36]  R. Kruger,et al.  Scatter estimation for a digital radiographic system using convolution filtering. , 1987, Medical physics.

[37]  David L Donoho,et al.  Compressed sensing , 2006, IEEE Transactions on Information Theory.

[38]  D. Jaffray,et al.  Optimization of x-ray imaging geometry (with specific application to flat-panel cone-beam computed tomography). , 2000, Medical physics.

[39]  Philippe Gailloud,et al.  CT fluoroscopy: novel application for the treatment of ventricular pathologies , 2007, Neuroradiology.

[40]  Benjamin Movsas,et al.  SU‐FF‐J‐01: Feasibility of a Pre‐Object‐Grid to Reduce Scatter and Improve Image Quality in Cone‐Beam Computed Tomography (CBCT) , 2009 .

[41]  U Neitzel,et al.  Grids or air gaps for scatter reduction in digital radiography: a model calculation. , 1992, Medical physics.

[42]  E. Sidky,et al.  Image reconstruction in circular cone-beam computed tomography by constrained, total-variation minimization , 2008, Physics in medicine and biology.

[43]  A. Macovski,et al.  Dual-energy x-ray projection imaging: two sampling schemes for the correction of scattered radiation. , 1988, Medical physics.