AMBIENT DOSE EQUIVALENT RECEIVED AT 15 MV LINAC ENVIRONMENTS DUE TO PHOTONEUTRONS CONTAMINATION

Mapping of the photoneutron spectrum and flux performed at several positions over the treatment table of a 15 MV medical linac. Neutron activation technique has been applied to obtain the neutron spectra in the mixed γ -n field. Thick foils have been used due to the low neutron flux and the measurements were unfolded using the MINUIT minimizer. GEANT4 code simulations have been performed to reproduce the photoneutrons' spectrum and estimate the flux along the treatment table. The mixed radiation field around medical linear accelerator was evaluated and the variation of the ambient dose equivalent over the treatment table was estimated. Photoneutrons' ambient dose equivalent inside the irradiation field estimated as 0.56 ± 0.05 while the peripheral as 0.42 ± 0.04 mSv Gy-1. The effective dose inside the irradiation field is ranged between 0.004 and 0.049 while the peripheral up to 0.035 mSv Gy-1.

[1]  K. Theodorou,et al.  Thick-foils activation technique for neutron spectrum unfolding with the MINUIT routine-Comparison with GEANT4 simulations , 2018 .

[2]  M. Axiotis,et al.  SPECTRW: A software package for nuclear and atomic spectroscopy , 2016 .

[3]  H. Yücel,et al.  Measurement of Photo-Neutron Dose from an 18-MV Medical Linac Using a Foil Activation Method in View of Radiation Protection of Patients , 2016 .

[4]  S. Stoulos,et al.  Geant4 simulations on medical Linac operation at 18 MV: Experimental validation based on activation foils , 2016 .

[5]  A. Konefał,et al.  Measurements of neutron radiation and induced radioactivity for the new medical linear accelerator, the Varian TrueBeam , 2016 .

[6]  S. Stoulos,et al.  Analysis of improved neutron activation technique using thick foils for application on medical LINAC environment , 2016 .

[7]  M. Hult,et al.  Validation of a method for neutron dosimetry and spectrometry using neutron activation of metal discs. , 2010, Radiation protection dosimetry.

[8]  D. Followill,et al.  Secondary neutron spectra from modern Varian, Siemens, and Elekta linacs with multileaf collimators. , 2009, Medical physics.

[9]  N. M. Spyrou,et al.  Monte Carlo study of photoneutron production in the Varian Clinac 2100C linac , 2008 .

[10]  J. Castelo,et al.  Neutron measurements in a Varian 2,100C LINAC facility using a Bonner sphere system based on passive gold activation detectors. , 2007, Radiation protection dosimetry.

[11]  B. Koo,et al.  Evaluation of the photoneutron field produced in a medical linear accelerator. , 2007, Radiation protection dosimetry.

[12]  Gary D Fullerton,et al.  Calculation of effective dose from measurements of secondary neutron spectra and scattered photon dose from dynamic MLC IMRT for 6 MV, 15 MV, and 18 MV beam energies. , 2006, Medical physics.

[13]  N. Hertel,et al.  Effective quality factors for neutrons based on the revised ICRP/ICRU recommendations. , 2005, Radiation protection dosimetry.

[14]  G. Fullerton,et al.  Investigation of secondary neutron dose for 18 MV dynamic MLC IMRT delivery. , 2005, Medical physics.

[15]  D Huyskens,et al.  Peripheral neutron and gamma doses in radiotherapy with an 18 MV linear accelerator. , 2004, Radiation protection dosimetry.

[16]  K. Turek,et al.  Characterisation of neutron fields around high-energy x-ray radiotherapy machines. , 2004, Radiation protection dosimetry.

[17]  A. Dell'Acqua,et al.  Geant4 - A simulation toolkit , 2003 .

[18]  M. Pelliccioni,et al.  Overview of Fluence-to-Effective Dose and Fluence-to-Ambient Dose Equivalent Conversion Coefficients for High Energy Radiation Calculated Using the FLUKA Code , 2000 .

[19]  K. Kase,et al.  Neutron fluence and energy spectra around the Varian Clinac 2100C/2300C medical accelerator. , 1998, Health physics.

[20]  K. Kase,et al.  Calculations of the Giant-Dipole-Resonance Photoneutrons Using a Coupled EGS4-Morse Code , 1997 .

[21]  G. Borasi,et al.  Radiation transport in a radiotherapy room. , 1995, Health physics.

[22]  M Matzke,et al.  Unfolding of pulse height spectra : the HEPRO program system , 1994 .

[23]  C. Manfredotti,et al.  Evaluation of the Undesired Neutron Dose Equivalent to Critical Organs in Patients Treated by Linear Accelerator Gamma Ray Therapy , 1992 .

[24]  W. J. Meredith Radioactivity: Recommendations of the International Commission on Radiological Units and Measurements , 1963 .