Surgeons’ Operation Skill-Based Control Strategy and Preliminary Evaluation for a Vascular Interventional Surgical Robot

BackgroundRemote vascular intervention surgery that employs surgical robot is an important research field. It is difficult to ensure that the surgical robot realize the surgeons’ operation skills precisely.ObjectiveTo solve this problem, surgical robot must be able to duplicate the surgeons’ changeable operation, and guarantee accuracy of synchronous motion between the surgeon’s unpredictable operation and robotic control of surgical tool.MethodIn this research, the surgeons’ operations were disassembled, eight basic actions were obtained. The surgical robot which can reproduce these basic actions is designed. Further, fuzzy control theory was applied to optimize the remote control and to guarantee the control precision of the synchronous operations.ResultUsing an experiment, the surgical procedures and operations of interventional surgical robot were obtained and evaluated. The robot can duplicate the surgeons’ operative skills when the control precision of the guidewire is guaranteed. And the axial displacement error of the surgical robot is limited to less than 2 mm.ConclusionThe new type of interventional surgical robot was designed and proved that the surgical robot can duplicate the operation skills of the surgeon. The control error of novel surgical robot reaches the accuracy required by the surgeon.

[1]  J. Brachmann,et al.  Atrial fibrillation ablation using a robotic catheter remote control system: initial human experience and long-term follow-up results. , 2008, Journal of the American College of Cardiology.

[2]  Kaspar Althoefer,et al.  Tension Sensing for a Linear Actuated Catheter Robot , 2015, ICIRA.

[3]  Shuxiang Guo,et al.  A virtual‐reality simulator and force sensation combined catheter operation training system and its preliminary evaluation , 2017, The international journal of medical robotics + computer assisted surgery : MRCAS.

[4]  Shuxiang Guo,et al.  A virtual reality‐based method of decreasing transmission time of visual feedback for a tele‐operative robotic catheter operating system , 2016, The international journal of medical robotics + computer assisted surgery : MRCAS.

[5]  Fumiya Iida,et al.  Improving Energy Efficiency of Hopping Locomotion by Using a Variable Stiffness Actuator , 2016, IEEE/ASME Transactions on Mechatronics.

[6]  Maria Drangova,et al.  Design and Evaluation of a Catheter Contact-Force Controller for Cardiac Ablation Therapy , 2016, IEEE Transactions on Biomedical Engineering.

[7]  Shuxiang Guo,et al.  A novel force feedback interventional surgery robotic system , 2015, 2015 IEEE International Conference on Mechatronics and Automation (ICMA).

[8]  Shuxiang Guo,et al.  Design and characteristics evaluation of a novel VR-based robot-assisted catheterization training system with force feedback for vascular interventional surgery , 2017 .

[9]  Shuxiang Guo,et al.  Safety Operation Consciousness Realization of a MR Fluids-Based Novel Haptic Interface for Teleoperated Catheter Minimally Invasive Neurosurgery , 2016, IEEE/ASME Transactions on Mechatronics.

[10]  Nicholas S. Peters,et al.  Experience of robotic catheter ablation in humans using a novel remotely steerable catheter sheath , 2008, Journal of Interventional Cardiac Electrophysiology.

[11]  Shuxiang Guo,et al.  Design and performance evaluation of a haptic interface based on MR fluids for endovascular tele-surgery , 2018 .

[12]  Shuxiang Guo,et al.  Study on motion following with feedback force disturbance in interventional surgical robot system , 2016, 2016 IEEE International Conference on Mechatronics and Automation.

[13]  Shuxiang Guo,et al.  Study on haptic feedback functions for an interventional surgical robot system , 2015, 2015 IEEE International Conference on Mechatronics and Automation (ICMA).

[14]  Jenny Dankelman,et al.  Current technology on minimally invasive surgery and Interventional Techniques , 2011, 2011 2nd International Conference on Instrumentation Control and Automation.

[15]  Shuxiang Guo,et al.  Design and characteristics evaluation of a novel teleoperated robotic catheterization system with force feedback for vascular interventional surgery , 2016, Biomedical microdevices.

[16]  Jinwoo Jung,et al.  Towards closed loop control of a continuum robotic manipulator for medical applications , 2011, 2011 IEEE International Conference on Robotics and Automation.

[17]  Robert D. Howe,et al.  Force control of flexible catheter robots for beating heart surgery , 2011, 2011 IEEE International Conference on Robotics and Automation.

[18]  Qiang Gao,et al.  Design and performance evaluation of a novel robotic catheter system for vascular interventional surgery , 2016 .

[19]  Zhuang Xiong,et al.  Fabrication of a silicon based vertical sensitive low-g inertial micro-switch for linear acceleration sensing , 2017 .

[20]  Rajnikant V. Patel,et al.  Robot-assisted Active Catheter Insertion: Algorithms and Experiments , 2009, Int. J. Robotics Res..

[21]  Shuxiang Guo,et al.  Design and experimental evaluation of a teleoperated haptic robot–assisted catheter operating system , 2016 .

[22]  Nan Xiao,et al.  A novel robotic catheter system with force and visual feedback for vascular interventional surgery , 2012, Int. J. Mechatronics Autom..

[23]  Shuxiang Guo,et al.  Development of a kind of robotic catheter manipulation system , 2011, 2011 IEEE International Conference on Robotics and Biomimetics.

[24]  Shuxiang Guo,et al.  Development of a novel robotic catheter manipulating system , 2012, 2012 International Conference on Manipulation, Manufacturing and Measurement on the Nanoscale (3M-NANO).

[25]  Shuxiang Guo,et al.  A robotic catheter system with real-time force feedback and monitor , 2012, Australasian Physical & Engineering Sciences in Medicine.

[26]  Fumihito Arai,et al.  In vitro patient-tailored anatomical model of cerebral artery for evaluating medical robots and systems for intravascular neurosurgery , 2005, 2005 IEEE/RSJ International Conference on Intelligent Robots and Systems.

[27]  Hugh Calkins,et al.  First experience with a novel robotic remote catheter system: Amigo™ mapping trial , 2013, Journal of interventional cardiac electrophysiology.

[28]  Shuxiang Guo,et al.  Performance evaluation of a strain-gauge force sensor for a haptic robot-assisted catheter operating system , 2017 .