Intensity-modulated radiation therapy using only jaws and a mask.

Intensity-modulated radiation therapy (IMRT) generally requires complex equipment for delivery. Just one study has investigated the use of 'jaws-only' IMRT with not discouraging conclusions. However, the monitor-unit efficiency is still considered to be too low compared with the use of a multileaf collimator (MLC). In this paper a new IMRT delivery technique is proposed which does not require the MLC and is only moderately more complex than the use of jaws alone. In this method a secondary collimator (mask) is employed together with the jaws. This mask may translate parallel to the jaw axes. Two types of mask have been investigated. One is a regular binary-attenuation pattern and the other is a random binary-attenuation pattern. Studies show that the monitor-unit efficiency of this 'jaws-plus-mask' technique, with a random binary mask, is more than double that of the jaws-only technique for typical two-dimensional intensity-modulated beams of size 10 x 10 bixels2 and with a peak value of 10 MU (or quantized into 10 fluence increments). For two-dimensional intensity-modulated beams of size 15 x 15 bixels2 with a peak value of 10 MU (or quantized into 10 fluence increments), the monitor-unit efficiency of the 'jaws-plus-mask' technique with a random binary mask is almost triple that of the jaws-only technique. Some further extensions to this concept are presented showing that some more practical mask arrangements are possible but with somewhat compromised monitor-unit efficiency. Some comments are provided on practicalities and on delivery times.

[1]  G J Budgell,et al.  Improved delivery efficiency for step and shoot intensity modulated radiotherapy using a fast-tuning magnetron. , 2001, Physics in medicine and biology.

[2]  P Cooper,et al.  High-resolution field shaping utilizing a masked multileaf collimator. , 2000, Physics in medicine and biology.

[3]  S. Webb Conformal intensity-modulated radiotherapy (IMRT) delivered by robotic linac--conformality versus efficiency of dose delivery. , 2000, Physics in medicine and biology.

[4]  W Schlegel,et al.  Dynamic X-ray compensation for conformal radiotherapy by means of multi-leaf collimation. , 1994, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[5]  N. S. Barnett,et al.  Private communication , 1969 .

[6]  Yi-min Hu,et al.  Intensity-modulation radiotherapy using independent collimators: an algorithm study. , 1999, Medical physics.

[7]  T. Bortfeld,et al.  Realization and verification of three-dimensional conformal radiotherapy with modulated fields. , 1994, International journal of radiation oncology, biology, physics.

[8]  C. Fiorino,et al.  Dynamic beam modulation by using a single computer-controlled absorber , 1995, Physics in medicine and biology.

[9]  W. Que Comparison of algorithms for multileaf collimator field segmentation. , 1999, Medical physics.

[10]  S. Webb Conformal intensity-modulated radiotherapy (IMRT) delivered by robotic linac--testing IMRT to the limit? , 1999, Physics in medicine and biology.

[11]  S Webb Concepts for shuttling multileaf collimators for intensity-modulated radiation therapy. , 2001, Physics in medicine and biology.