Results of a combined monolithic crystal and an array of ASICs controlled SiPMs

Abstract In this work we present the energy and spatial resolutions we have obtained for a γ - ray detector based on a monolithic LYSO crystal coupled to an array of 256 SiPMs. Two crystal configurations of the same trapezoidal shape have been tried. In one approach all surfaces were black painted but the exit one facing the photosensor array which was polished. The other approach included a retroreflector (RR) layer coupled to the entrance face of the crystal powering the amount of transmitted light to the photosensors. Two coupling media between the scintillator and the SiPM array were used, namely direct coupling by means of optical grease and coupling through an array of light guides. Since the same operational voltage was supplied to the entire array, it was needed to equalize their gains before feeding their signals to the Data Acquisition system. Such a job was performed by means of 4 scalable Application Specific Circuits (ASICs). An energy resolution of about 24.4% has been achieved for the direct coupling with the RR layer together with a spatial resolution of approximately 2.9 mm at the detector center. With the light guides coupling the effects of image compression at the edges are significantly minimized, but worsening the energy resolution to about 33.1% with a spatial resolution nearing 4 mm at the detector center.

[1]  Sascha Moehrs,et al.  A detector head design for small-animal PET with silicon photomultipliers (SiPM) , 2006, Physics in medicine and biology.

[2]  F. Martínez,et al.  Innovative PET detector concept based on SiPMs and continuous crystals , 2012 .

[3]  J. Benlloch,et al.  Design of the PET-MR system for head imaging of the DREAM Project , 2013 .

[4]  A. Sebastia,et al.  Depth of /spl gamma/-ray interaction within continuous crystals from the width of its scintillation light-distribution , 2005, IEEE Transactions on Nuclear Science.

[5]  D. Schaart,et al.  Ultra precise timing with SiPM-based TOF PET scintillation detectors , 2009, 2009 IEEE Nuclear Science Symposium Conference Record (NSS/MIC).

[6]  Mohammad Reza Ay,et al.  COMPUTED TOMOGRAPHY BASED ATTENUATION CORRECTION IN PET/CT: PRINCIPLES, INSTRUMENTATION, PROTOCOLS, ARTIFACTS AND FUTURE TRENDS , 2007 .

[7]  Christoph W. Lerche,et al.  AMIC: An expandable front-end for gamma-ray detectors with light distribution analysis capabilities , 2010, 2010 17th IEEE-NPSS Real Time Conference.

[8]  Ciprian Catana,et al.  Simultaneous PET-MRI: a new approach for functional and morphological imaging , 2008, Nature Medicine.

[9]  C. Lerche Depth of Interaction Enhanced Gamma-Ray Imaging for Medical Applications , 2006 .

[10]  G. Landi Properties of the center of gravity as an algorithm for position measurements , 2002, 1908.04447.

[11]  M. J. Rodriguez-Alvarez,et al.  Minimization of Parallax Error in Dedicated Breast PET , 2013, IEEE Transactions on Nuclear Science.

[12]  Vicente Herrero-Bosch,et al.  Expandable programmable integrated front-end for scintillator based photodetectors , 2012, 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record (NSS/MIC).

[13]  V. Herrero-Bosch,et al.  Fast circuit topology for spatial signal distribution analysis , 2010, 2010 17th IEEE-NPSS Real Time Conference.

[14]  F. Beekman,et al.  A novel, SiPM-array-based, monolithic scintillator detector for PET , 2009, Physics in medicine and biology.

[15]  Julio Chaves,et al.  Introduction to Nonimaging Optics , 2008 .

[16]  A. Sebastia,et al.  Scanner calibration of a small animal PET camera based on continuous LSO crystals and flat panel PSPMTs , 2007 .

[17]  Manuel Desco,et al.  Performance evaluation of SiPM detectors for PET imaging in the presence of magnetic fields , 2008, 2008 IEEE Nuclear Science Symposium Conference Record.