Comparison of Alloys and Designs in a Hip Simulator Study of Metal on Metal Implants

Previous studies of metal on metal hip implants have shown that it is possible to obtain substantially lower volumetric wear than with metal on polyethylene pairings. To work toward design optimization, the gravimetric wear of serum lubricated, metal on metal implants was examined in a hip simulator apparatus. Seventeen implant specimens were fabricated from 3 alloys in various combinations of diameter and clearance and wear tested as many as 3 million cycles. Theory was developed to predict the thickness of elastohydrodynamic lubricant films, the linear wear, and wear zone geometry. To help interpret the data, implant diameter and clearance were combined into a single parameter called effective radius. Implants with larger effective radii (6-11 m) tended to experience lower wear, and theory suggested that this was a result of the correspondingly thicker elastohydrodynamic lubricant films providing an enhanced mixed film lubrication. As in most wear testing, the results were scattered but, overall, a low carbon, wrought alloy had lower wear than either a high carbon, cast alloy or a low carbon, commercial alloy. Because of the small number of specimens tested and possible variations in tribologic parameters such as calcium phosphate film formation and surface microgeometry, there was low statistical reliability of the results.

[1]  M. Müller The benefits of metal-on-metal total hip replacements. , 1995, Clinical Orthopaedics and Related Research.

[2]  H S Dobbs,et al.  Survivorship of total hip replacements. , 1980, The Journal of bone and joint surgery. British volume.

[3]  A Sarmiento,et al.  The origin of submicron polyethylene wear debris in total hip arthroplasty. , 1995, Clinical orthopaedics and related research.

[4]  J. Gorczyca,et al.  Low friction arthroplasty at 10 to 20 years. Consequences of plastic wear. , 1995, Clinical Orthopaedics and Related Research.

[5]  R M Rose,et al.  A comparative study of total hip replacement prostheses. , 1973, Journal of biomechanics.

[6]  V O Saikko,et al.  Wear of the polyethylene acetabular cup. Metallic and ceramic heads compared in a hip simulator. , 1993, Acta orthopaedica Scandinavica.

[7]  Michael Tanzer,et al.  Kinematics of the MATCO™ hip simulator and issues related to wear testing of metal-metal implants , 1997, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[8]  I C Clarke,et al.  Mechanism and clinical significance of wear debris-induced osteolysis. , 1992, Clinical orthopaedics and related research.

[9]  P S Walker,et al.  The tribology (friction, lubrication and wear) of all-metal artificial hip joints. 1971. , 1971, Clinical orthopaedics and related research.

[10]  P. S. Walker,et al.  Metal-on-metal lubrication in artificial human joints , 1972 .

[11]  P. Walker A comparison of normal and artificial human joints. , 1973, Acta orthopaedica Belgica.