Detection of intrafractional tumour position error in radiotherapy utilizing cone beam computed tomography.

PURPOSE To study the intrafraction tumour position error utilizing cone beam CT (CBCT) in patients receiving radiotherapy. METHODS AND MATERIALS Fifty-four patients were treated with Elekta Synergy S system, including 19 head and neck, 25 thoracic and 10 abdominal-pelvic tumours. All patients received CBCT after initial setup and some of them received CBCT after correction and after treatment. CBCT were registered to planning CT and errors of isocenter position on the left-to-right (LR), superior-inferior (SI) and anterior-posterior (AP) directions were analyzed. RESULTS After treatment the systematic (Sigma) and random uncertainty (sigma) increased, the increments of tumour Sigma were 0.1-0.3, 0.2-0.5 and 0.2-0.6 mm, respectively, while the increments of sigma were 0.1-0.3, 0.2-0.4 and 0.1-0.4 mm, respectively, for the head and neck, thoracic and abdominal-pelvic tumours. Based on 380 paired pre- and post-treatment CBCT, the intrafraction errors (mean+/-SD) in the LR, SI and AP directions were -0.1+/-0.9, -0.3+/-1.0 and -0.2+/-0.7 mm, respectively, for head and neck tumours, -0.1+/-1.2, -0.1+/-1.9 and 0.1+/-1.3 mm, respectively, for thoracic tumours, -0.1+/-1.1, 0.2+/-1.4 and -0.1+/-1.5 mm, respectively, for abdominal-pelvic tumours. Isotropic planning margins of 3.4, 6.1 and 5.4 mm were generated with linear addition of internal margin (IM) to CTV for the head and neck, thoracic and abdominal-pelvic tumours, respectively, while margins were only 2.4, 4.4 and 3.9 mm, respectively, if IM was added in quadrature. CONCLUSIONS Utilizing CBCT measurements before and after treatment to detect intrafraction tumour position errors was clinically feasible. The detected intrafraction errors could be applied to improve the accuracy of radiation delivery.

[1]  B. Heijmen,et al.  Geometrical uncertainties, radiotherapy planning margins, and the ICRU-62 report. , 2002, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[2]  Michael Flentje,et al.  Impact of target reproducibility on tumor dose in stereotactic radiotherapy of targets in the lung and liver. , 2003, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[3]  D. Yan,et al.  Computed tomography guided management of interfractional patient variation. , 2005, Seminars in radiation oncology.

[4]  Y. Nishimura,et al.  Analysis of interfractional set-up errors and intrafractional organ motions during IMRT for head and neck tumors to define an appropriate planning target volume (PTV)- and planning organs at risk volume (PRV)-margins. , 2006, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[5]  Christopher J Moore,et al.  A feasibility study for image guided radiotherapy using low dose, high speed, cone beam X-ray volumetric imaging. , 2005, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[6]  S. Korreman,et al.  IGRT of prostate cancer; is the margin reduction gained from daily IG time-dependent? , 2006, Acta oncologica.

[7]  M Goitein,et al.  Calculation of the uncertainty in the dose delivered during radiation therapy. , 1985, Medical physics.

[8]  D. Hallahan,et al.  A study on adaptive IMRT treatment planning using kV cone-beam CT. , 2007, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[9]  Ping Xia,et al.  Intra-fraction patient motion in head/neck cancer patients undergoing intensity-modulated radiation therapy (IMRT) , 2003 .

[10]  David Jaffray,et al.  Assessment of residual error in liver position using kV cone-beam computed tomography for liver cancer high-precision radiation therapy. , 2006, International journal of radiation oncology, biology, physics.

[11]  J. Crook,et al.  Prostate motion during standard radiotherapy as assessed by fiducial markers. , 1995, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[12]  Shinichi Shimizu,et al.  Registration accuracy and possible migration of internal fiducial gold marker implanted in prostate and liver treated with real-time tumor-tracking radiation therapy (RTRT). , 2002, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[13]  L. Muren,et al.  Organ motion, set-up variation and treatment margins in radical radiotherapy of urinary bladder cancer. , 2003, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[14]  G. Jiang,et al.  Application of active breathing control in 3-dimensional conformal radiation therapy for hepatocellular carcinoma: the feasibility and benefit. , 2008, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[15]  Matthias Guckenberger,et al.  Nonrigid Patient Setup Errors in the Head-and-Neck Region , 2007, Strahlentherapie und Onkologie.

[16]  Lei Dong,et al.  Intrafraction prostate motion during IMRT for prostate cancer. , 2001, International journal of radiation oncology, biology, physics.

[17]  G. Ezzell,et al.  Larynx motion associated with swallowing during radiation therapy. , 1994, International journal of radiation oncology, biology, physics.

[18]  K. Sheng,et al.  Is daily computed tomography image guidance necessary for nasal cavity and nasopharyngeal radiotherapy? An investigation based on helical tomotherapy , 2008, Journal of applied clinical medical physics.

[19]  P. Remeijer,et al.  Set-up verification using portal imaging; review of current clinical practice. , 2001, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[20]  David A Jaffray,et al.  Patient dose from kilovoltage cone beam computed tomography imaging in radiation therapy. , 2006, Medical physics.

[21]  A L Boyer,et al.  Dosimetric effects of patient displacement and collimator and gantry angle misalignment on intensity modulated radiation therapy. , 2000, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[22]  Lei Dong,et al.  Experience of ultrasound-based daily prostate localization. , 2003, International journal of radiation oncology, biology, physics.

[23]  H Alasti,et al.  Portal imaging for evaluation of daily on-line setup errors and off-line organ motion during conformal irradiation of carcinoma of the prostate. , 2001, International journal of radiation oncology, biology, physics.

[24]  M. Herk Errors and margins in radiotherapy. , 2004 .

[25]  T. Landberg,et al.  What margins should be added to the clinical target volume in radiotherapy treatment planning for lung cancer? , 1998, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[26]  Jatinder R Palta,et al.  Evaluation of intrafraction patient movement for CNS and head & neck IMRT. , 2004, Medical physics.

[27]  V. Khoo,et al.  X-ray volumetric imaging in image-guided radiotherapy: the new standard in on-treatment imaging. , 2006, International journal of radiation oncology, biology, physics.

[28]  Michael Unser,et al.  A pyramid approach to sub-pixel image fusion based on mutual information , 1996, Proceedings of 3rd IEEE International Conference on Image Processing.

[29]  H. Honda,et al.  Stereotactic radiotherapy for lung and liver tumors using a body cast system: setup accuracy and preliminary clinical outcome. , 2005, Radiation medicine.

[30]  Matthias Guckenberger,et al.  Intra-fractional uncertainties in cone-beam CT based image-guided radiotherapy (IGRT) of pulmonary tumors. , 2007, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[31]  Benjamin Movsas,et al.  Measurement of intrafractional prostate motion using magnetic resonance imaging. , 2002, International journal of radiation oncology, biology, physics.

[32]  Jan J W Lagendijk,et al.  Measurements and clinical consequences of prostate motion during a radiotherapy fraction. , 2002, International journal of radiation oncology, biology, physics.

[33]  U Oppitz,et al.  Stereotactic radiotherapy of extracranial targets: CT-simulation and accuracy of treatment in the stereotactic body frame. , 2000, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[34]  L. Feldkamp,et al.  Practical cone-beam algorithm , 1984 .

[35]  Katrien De Jaeger,et al.  Chapte rr 3 THEE EFFECT OF BREATHING AND SET-UP ERRORS ONN THE CUMULATIVE DOSE TO A LUNG TUMOUR , 2007 .

[36]  H Shirato,et al.  Use of an implanted marker and real-time tracking of the marker for the positioning of prostate and bladder cancers. , 2000, International journal of radiation oncology, biology, physics.

[37]  K. Lam,et al.  Uncertainties in CT-based radiation therapy treatment planning associated with patient breathing. , 1996, International journal of radiation oncology, biology, physics.

[38]  Stanley J. Rosenthal,et al.  A precision cranial immobilization system for conformal stereotactic fractionated radiation therapy. , 1995, International journal of radiation oncology, biology, physics.

[39]  Lena Specht,et al.  Intra- and interfraction breathing variations during curative radiotherapy for lung cancer. , 2007, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[40]  J Pouliot,et al.  Robustness and precision of an automatic marker detection algorithm for online prostate daily targeting using a standard V-EPID. , 2003, Medical physics.

[41]  Jan-Jakob Sonke,et al.  3D in vivo dose verification of entire hypo-fractionated IMRT treatments using an EPID and cone-beam CT. , 2008, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[42]  T. Marchant,et al.  Imaging doses from the Elekta Synergy X-ray cone beam CT system. , 2007, The British journal of radiology.

[43]  A L McKenzie,et al.  How should breathing motion be combined with other errors when drawing margins around clinical target volumes? , 2000, The British journal of radiology.

[44]  Michael G Herman,et al.  A methodology to determine margins by EPID measurements of patient setup variation and motion as applied to immobilization devices. , 2004, Medical physics.

[45]  Joos V Lebesque,et al.  Biologic and physical fractionation effects of random geometric errors. , 2003, International journal of radiation oncology, biology, physics.

[46]  K. Brock,et al.  Feasibility of a novel deformable image registration technique to facilitate classification, targeting, and monitoring of tumor and normal tissue. , 2006, International journal of radiation oncology, biology, physics.

[47]  John Wong,et al.  Assessment of residual error for online cone-beam CT-guided treatment of prostate cancer patients. , 2004, International journal of radiation oncology, biology, physics.

[48]  B G Fallone,et al.  A grey-level image alignment algorithm for registration of portal images and digitally reconstructed radiographs. , 1996, Medical physics.