An investigation into the temporal stability of CdTe-based photon counting detectors during spectral micro-CT acquisitions

Spectroscopic micro-computed tomography (CT) using photon counting detectors is a technology that promises to deliver material-specific images in pre-clinical research while suppressing noise related to dark current. Inherent to such applications is the need for a high spatial resolution, which can only be achieved with small focal spot sizes in the micrometre range. This limits the achievable x-ray fluxes and implies long acquisitions easily exceeding one hour, during which it is paramount to maintain a constant detector response. During the past years, the Medipix3RX detector has become a popular platform for research into low-flux and multi-energy CT, as it hosts a number of elaborate features that may serve as an inspiration for future systems when studied carefully. In this contribution, we illustrate our findings on how to reduce image artefacts in CT reconstructions using such a detector with a cadmium telluride sensor. We find that maintaining a constant temperature is a prerequisite to guarantee energy threshold stability. More importantly, we identify transient sensor leakage currents as a significant source to artefact formation. We demonstrate an improvement in image quality when allowing the chip to compensate for abnormally high leakge currents. The reduction of spectral resolution reported earlier for such settings is found to have no detrimental effect on material decomposition. Quite the contrary, the improved handling of transient leakage currents allows to even increase contrast-to-noise ratios in material decomposed images due to a reduced artefact level.

[1]  K. Zanio,et al.  Transient Currents in Semi‐Insulating CdTe Characteristic of Deep Traps , 1968 .

[2]  Uwe Oelfke,et al.  Investigating the feasibility of photon-counting K-edge imaging at high x-ray fluxes using nonlinearity corrections. , 2013, Medical physics.

[3]  R. Bellazzini,et al.  Chromatic X-ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC , 2013 .

[4]  Nigel G Anderson,et al.  Clinical applications of spectral molecular imaging: potential and challenges. , 2014, Contrast media & molecular imaging.

[5]  Uwe Oelfke,et al.  Imaging properties of small-pixel spectroscopic x-ray detectors based on cadmium telluride sensors , 2012, Physics in medicine and biology.

[6]  Josef Uher,et al.  Data acquisition and processing software package for Medipix2 , 2006 .

[7]  Anthony P. H. Butler,et al.  Energy Calibration of the Pixels of Spectral X-ray Detectors , 2015, IEEE Transactions on Medical Imaging.

[8]  M. Kachelriess,et al.  Performance of today's dual energy CT and future multi energy CT in virtual non-contrast imaging and in iodine quantification: A simulation study. , 2015, Medical physics.

[9]  Katsuyuki Taguchi,et al.  An analytical model of the effects of pulse pileup on the energy spectrum recorded by energy resolved photon counting x-ray detectors. , 2010 .

[10]  Thomas Koenig,et al.  Performance of a Medipix3RX Spectroscopic Pixel Detector With a High Resistivity Gallium Arsenide Sensor , 2015, IEEE Transactions on Medical Imaging.

[11]  C. Ponchut,et al.  Defect signature, instabilities and polarization in CdTe X-ray sensors with quasi-ohmic contacts , 2014 .

[12]  François Krummenacher,et al.  Pixel detectors with local intelligence: an IC designer point of view , 1991 .

[13]  I. Kawrakow,et al.  Investigation of variance reduction techniques for Monte Carlo photon dose calculation using XVMC , 2000, Physics in medicine and biology.

[14]  Polad M Shikhaliev,et al.  Photon counting multienergy x-ray imaging: effect of the characteristic x rays on detector performance. , 2009, Medical physics.

[15]  Vaclav Kraus,et al.  FITPix ? fast interface for Timepix pixel detectors , 2011 .

[16]  Mini Das,et al.  Energy calibration of photon counting detectors using x-ray tube potential as a reference for material decomposition applications , 2015, Medical Imaging.

[17]  M. Campbell,et al.  Count rate linearity and spectral response of the Medipix3RX chip coupled to a 300μm silicon sensor under high flux conditions , 2014 .

[18]  Uwe Oelfke,et al.  Saturation effects of CdTe photon counting detectors under high photon fluxes , 2013 .

[19]  D. Maneuski,et al.  Probing Defects in a Small Pixellated CdTe Sensor Using an Inclined Mono Energetic X-Ray Micro Beam , 2012, IEEE Transactions on Nuclear Science.

[20]  Michael Campbell,et al.  Charge Summing in Spectroscopic X-Ray Detectors With High-Z Sensors , 2013, IEEE Transactions on Nuclear Science.

[21]  Michael Campbell,et al.  The Medipix3RX: a high resolution, zero dead-time pixel detector readout chip allowing spectroscopic imaging , 2013 .

[22]  I. Kawrakow,et al.  Large efficiency improvements in BEAMnrc using directional bremsstrahlung splitting. , 2004, Medical physics.

[23]  A. Macovski,et al.  Energy-selective reconstructions in X-ray computerised tomography , 1976, Physics in medicine and biology.

[24]  J. Schlomka,et al.  Experimental feasibility of multi-energy photon-counting K-edge imaging in pre-clinical computed tomography , 2008, Physics in medicine and biology.

[25]  Michael Campbell,et al.  How spectroscopic x-ray imaging benefits from inter-pixel communication. , 2014, Physics in medicine and biology.

[26]  M. Campbell,et al.  Imaging properties of the Medipix2 system exploiting single and dual energy thresholds , 2004, IEEE Transactions on Nuclear Science.